Olympus Corporation of the Americas recalls
114 official recalls and safety alerts involving Olympus Corporation of the Americas since 2021-10, from FDA · Device. Newest first. Check the official notice for exact models and batch codes.
This list only includes recalls Recall Atlas has collected since 2021-10. A product not listed here may still be affected; check with Olympus Corporation of the Americas or the relevant agency.
- 2026-10-07
US
Olympus Corporation of the Americas recalls Olympus Tripp-Lite P580-025 25ft DisplayPort Cable, Model Number P580-025, an accessory used in conjunction wtih OLYSENSE…
While in the high-power operating mode, a video dropout/blackout may occur when using a particular accessory cable.
FDA · DeviceOlympus Corporation of the AmericasWorldwide distribution - US Nationwide and the country of Poland.Official notice →US
Olympus Corporation of the Americas recalls Olympus M3 Gold Autoclavable Foroblique Telescope 30. Model Number: M3-30A.
There are identified discrepancies between the sterilization parameters listed in the Instructions for Use and the sterilization parameters listed in the Japanese package insert.
FDA · DeviceOlympus Corporation of the AmericasInternational distribution to the country of Japan.Official notice →- 2026-09-30
US
Olympus Corporation of the Americas recalls Olympus BiCOAG Hemostasis Probe 7Fr Fixed Pin Connector. Model/Catalog Number: CD-B622LA.
Potential for detachment of ceramic tip due to manufacturing without the designated amount of adhesive.
FDA · DeviceOlympus Corporation of the AmericasWorldwide distribution - US Nationwide and the countries of Canada, Latin…Official notice →US
Olympus Corporation of the Americas recalls Olympus EZDilate Endoscopic Balloon Dilator Wired-guided 11, 12, 13mm. Model Number: BD-420X-1355.
Potential for the inflation chart label affixed to the shaft of the device to be incorrect and indicate a balloon diameter of 18, 19, or 20 mm,
FDA · DeviceOlympus Corporation of the AmericasInternational distribution in the country of Japan.Official notice →- 2026-09-02
US
Olympus Corporation of the Americas recalls POWERSEAL 5 MM (in diameter) 23 CM (long), Curved Jaw Sealer & Divider, Double Action. GEI / Electrosurgical, cutting &…
Potential labeling deficiency in which an outer shelf box is incorrectly labeled as a 37-cm device.
FDA · DeviceOlympus Corporation of the AmericasWorldwide - US Nationwide distribution in the states of Kansas and Florida.…Official notice →- 2026-07-29
US
Olympus Corporation of the Americas recalls Olympus High Flow Insufflation Unit - UHI-4. Model Number: UHI-4
Potential for unexpected instances where the front panel display turns off unexpectedly, and air delivery (insufflation) ceases.
FDA · DeviceOlympus Corporation of the AmericasWorldwide distribution - US Nationwide and the countries of Australia,…Official notice →US
Olympus Corporation of the Americas recalls Olympus 4K UHD LCD Monitor. Model Number: OEV321UH. An electrically powered visual display unit designed for displaying…
Potential for the HDMI and DisplayPort (DP) connectors can become faulty resulting in image loss or image abnormalities.
FDA · DeviceOlympus Corporation of the AmericasWorldwide - US Nationwide distribution in the states of OH, CA, NJ, FL and the…Official notice →- 2026-07-22
US
Olympus Corporation of the Americas recalls Olympus Ellik Evacuator Includes Glass Body, Male Connecting Tubing & Latex Bulb. Model Number: 194.
Some sterilization parameters included in the IFU may not consistently render the devices adequately sterile and therefore should no longer be used to sterilize the devices.
FDA · DeviceOlympus Corporation of the AmericasWorldwide distribution - US Nationwide and the countries of Canada, Japan, EU,…Official notice →- 2026-07-15
US
Olympus Corporation of the Americas recalls Olympus Single Use Distal Cover, Catalog Number MAJ-2315, Used with: EVIS EXERA III Duodenovideoscope TJF-Q190V- designed to…
Olympus has developed and released newly designed version of the Distal Cover (MAJ-2315) to replace previous design of the Single Use Distal Covers.
- 2026-04-29
US
Olympus Corporation of the Americas recalls OER-ELITE Endoscope Reprocessor Connecting Tubes. Model/Catalog Number: MAJ-2115.
XXX
FDA · DeviceOlympus Corporation of the AmericasWorldwide distribution - US Nationwide and the countries of Canada, China,…Official notice →US
Olympus Corporation of the Americas recalls OER-ELITE Endoscope Reprocessor Connecting Tubes. Model/Catalog Number: MAJ-2113.
Potential for the Version 2 reprocessor connecting tube lock levers may fail prematurely.
FDA · DeviceOlympus Corporation of the AmericasWorldwide distribution - US Nationwide and the countries of Canada, China,…Official notice →US
Olympus Corporation of the Americas recalls Olympus Thunderbeat II Shears w/Ultrasonic Mode, 5mm, 45cm. Model Number: TB2-0545FC.
Potential for detachment of a distal tip component of the device during use.
FDA · DeviceOlympus Corporation of the AmericasInternational distribution in the countries of Australia, Japan, Hong Kong,…Official notice →- 2026-04-08
US
Olympus Corporation of the Americas recalls Brand Name: Olympus SOLTIVE Premium SuperPulsed Laser System Product Name: SOLTIVE Premium Laser System TFL-PLS…
Following a complaint investigation, Olympus identified a potential defect in the 24V power supply module in some SOLTIVE laser units which may cause the system to become inoperable. Additionally, smoke or a burning…
FDA · DeviceOlympus Corporation of the AmericasWorldwide distribution - US Nationwide and the countries of Canada, Chile,…Official notice →- 2026-04-01
US
Olympus Corporation of the Americas recalls Brand Name: Olympus PKS Cutting Forceps, 5mm 33cm Product Name: PKS Cutting Forceps Model/Catalog Number: 3005PK Product…
Olympus identified that the Cutting Forceps contain components for which the supplier did not adequately validate the welding process. Defective welds can result in the Cutting Forceps' jaw breaking during clinical use.
FDA · DeviceOlympus Corporation of the AmericasWorldwide - US Nationwide and the countries of Canada, Mexico, EU, Australia,…Official notice →US
Olympus Corporation of the Americas recalls Olympus Endoscope Suction Valve - MAJ-1443. Model Number: MAJ-1443.
The MAJ-1443 and MAJ-1444 are no longer reprocessing compatible with the OER-Pro and OER-Elite automated endoscope reprocessors.
- 2026-03-11
US
Olympus Corporation of the Americas recalls Brand Name: Olympus ShockPulse-SE Lithotripsy System with Generator Product Name: ShockPulse-SE Lithotripsy System -…
Mis-wired component-the improperly wired component can result in additional noise on the power supply output. Noise on the ultrasonic input circuit is not consistent with electromagnetic compatibility (EMC) standards…
FDA · DeviceOlympus Corporation of the AmericasInternational distribution in the country of Canada, Germany, Singapore,…Official notice →US
Olympus Corporation of the Americas recalls Olympus Single use Biopsy Valve. Model/Catalog Number: MAJ-210. 20 units per box. The MAJ-210 has been designed to be…
Potential for rubber fragment detachment during use.
US
Olympus Corporation of the Americas recalls Brand Name: ShockPulse-SE Lithotripsy System Product Name: ShockPulse-SE Lithotripsy System - Generator Model/Catalog…
Ongoing investigations identified additional instances of the device Generator remaining in a blinking phase waiting to recognize the transducer. Damage to the transducer plug and/or generator receptacle may cause…
FDA · DeviceOlympus Corporation of the AmericasWorldwide distribution - US Nationwide and the countries of Canada, Mexico,…Official notice →- 2026-03-04
US
Olympus Corporation of the Americas recalls Olympus Sheath, 10 Fr. Model No. A37004A. for endoscopic diagnosis and treatment in urological applications.
Complaints of the ceramic tip of the resection sheath breaking have been received.
US
Olympus Corporation of the Americas recalls Olympus Cystoscope Outer Sheath; Model/Catalog number: WA22810A;
Product being removed due to incompatibility when used in conjunction with a GreenLight Laser for BPH therapy which could damage the tip of the device.
- 2026-02-25
US
Olympus Corporation of the Americas recalls Olympus Thunderbeat 5 mm, 45 cm, Front-Actuated Grip Type S
Firm is initiating a removal due to continued reports of adverse events.
FDA · DeviceOlympus Corporation of the AmericasWorldwide distribution - US Nationwide and the countries of Brazil, Canada,…Official notice →US
Olympus Corporation of the Americas recalls Olympus High Flow Insufflation Unit. Model Number: UHI-3. Used for insufflation of the abdominal cavity, and automatic…
Issue with software algorithm which may lead to overpressure events.
- 2026-02-11
US
Olympus Corporation of the Americas recalls Product Name: Single Use 3-Lumen Sphincterotome V Model/Catalog Number: KD-V411M-1520 Product Description: These instruments…
Devices which did not undergo thermoforming could deform and lose performance.
FDA · DeviceOlympus Corporation of the AmericasWorldwide - US Nationwide distribution including in the states of AL, AR, AZ,…Official notice →US
Olympus Corporation of the Americas recalls Brand Name: Single Use Electrosurgical Knife Product Name: Olympus Single Use Electrosurgical Knife KD-640L Model/Catalog…
A deterioration of the cutting knife, including overheating and burning, can contribute to tip breakage during use.
- 2025-12-03
US
Olympus Corporation of the Americas recalls Brand Name: EVIS EXERA III Duodenovideoscope Product Name: Olympus TJF-Q190V Model/Catalog Number: TJF-Q190V Product…
Continued reports of positive cultures and infections have identified updates to the reprocessing materials to minimize potential deviations in TJF duodenoscope reprocessing. Olympus is informing users of these…
US
Olympus Corporation of the Americas recalls Brand Name: Single Use Ligating Device Product Name: Olympus HX-400U-30 Model/Catalog Number: HX-400U-30 Used with an…
Potential that the ligation loop was unable to release or detach as expected during use, causing the loop to become unintentionally fixed around patient anatomy.
- 2025-11-26
US
Olympus Corporation of the Americas recalls Brand Name: EZDilate Nephrostomy Balloon Catheter and Inflation Device Kit Product Name: Olympus Nephro-EZDilate Nephrostomy…
Device kits contained an incorrectly sized 4mm x 4cm urinary tract balloon catheter (BURS0404) instead of the correct 8mm x 15cm nephrostomy tract balloon catheter (BPCN0815). All other components in the kits were…
FDA · DeviceOlympus Corporation of the AmericasUS Domestic: AK, CA, IA, KY, WA, WI; OUS International: Canada and Germany.Official notice →- 2025-11-19
US
Olympus Corporation of the Americas recalls Brand Name: Olympus Single Use Single Use Electrosurgical Snare SD-400 Product Name: Single Use Single Use Electrosurgical…
Sterility compromised from small holes that may occur during packaging of the device pouches into the shelf carton and/or during transportation of the shelf carton.
FDA · DeviceOlympus Corporation of the AmericasWorldwide distribution - US Nationwide and the countries of GU, and MP.Official notice →US
Olympus Corporation of the Americas recalls Brand Name: Lithotriptor, Ultrasonic Product Name: ShockPulse Lithotripsy Transducer Model/Catalog Number: SPL-T Product…
The device may either fail to start up, or the transducer may start briefly and then stop, accompanied by an error light on the generator. In addition, the body of the transducer handpiece may gradually increase in…
FDA · DeviceOlympus Corporation of the AmericasWorldwide distribution - US Nationwide and the countries of Canada, India,…Official notice →- 2025-10-22
US
Olympus Corporation of the Americas recalls EVIS EXERA BRONCHOVIDEOSCOPE OLYMPUS BF TYPE XT160
Additional IFU updates to provide further clarification on safe and effective use of bronchoscopes when used in conjunction with laser, argon plasma coagulation, and high-frequency therapy equipment.
- 2025-09-17
US
Olympus Corporation of the Americas recalls ViziShot 2FLEX. Single Use Aspiration Needle. Model Number: NA-U403SX-4019. The ViziShot 2 FLEX tissue biopsy device is made…
Potential for undetected, deformed a-traumatic tips.
FDA · DeviceOlympus Corporation of the AmericasWorldwide distribution - US Nationwide and the countries of Australia,…Official notice →- 2025-08-27
US
Olympus Corporation of the Americas recalls Electrosurgical, Cutting and Coagulation and Accessories. Electrosurgical Generator ESG-410 Surgical Energy Platform (SEP).…
Potential for the ESG-410 Electrosurgical Generator to display an error message: "E0662 System Error", followed by either automatic single reboots of the device, or continuous reboot 'loops'.
- 2025-07-30
US
Olympus Corporation of the Americas recalls Ellik Evacuator. Model Number: 194. The Ellik Evacuator and adapter accessories are compatible with Gyrus ACMI Elite and…
Discrepancies between sterilization instructions in the Japanese language product package inserts and those published in the manufacturer product Instructions For Use (IFU).
FDA · DeviceOlympus Corporation of the AmericasInternational distribution to the country of Japan.Official notice →- 2025-06-18
US
Olympus Corporation of the Americas recalls Brand Name: Powered Laser Surgical Instrument Product Name: Soltive SuperPulsed Laser Fibers - Single Use Model/Catalog…
Potential for thermal incident including an unexpected occurrence of smoke, sparks, burning, or unsteady flames during clinical use. The fracturing of the fiber is the cause of, or contributes to, these thermal…
FDA · DeviceOlympus Corporation of the AmericasWorldwide distribution - US Nationwide and the countries of AU, CA, CL, DE,…Official notice →- 2025-06-11
US
Olympus Corporation of the Americas recalls Olympus EVIS EXERA III Duodenovideoscope. Model Number: TJF-Q190V
Recent post-market surveillance data suggests a possible association of higher microbial contamination levels from TJF duodenoscopes when manual cleaning was delayed beyond one hour and a presoak was performed,…
FDA · DeviceOlympus Corporation of the AmericasWorldwide Distribution: US (Nationwide) and OUS (International) countries to:…Official notice →- 2025-05-07
US
Olympus Corporation of the Americas recalls Brand Name: Powered Laser Surgical Instrument Product Name: SOLTIVE Pro SuperPulsed Laser System Model/Catalog Numbers:…
Preset treatment parameters are not consistently being used in accordance with the IFU.
FDA · DeviceOlympus Corporation of the AmericasWorldwide distribution - US Nationwide and the countries of AU, CA, SG, DE,…Official notice →- 2025-02-19
US
Olympus Corporation of the Americas recalls Olympus Single Use Guide Sheath Kits with the following product descriptions and model numbers: 1. Single Use Guide Sheath…
Potential for the radiopaque tip of the guide sheath component to disassociate and fall off into the patient.
- 2025-01-29
US
Olympus Corporation of the Americas recalls Brand Name: Olympus Thunderbeat Front-Actuated Grip Type S Hand Instrument Product Name: THUNDERBEAT 5 mm,35 cm,…
The probe tips of the hand instruments are being damaged or breaking including pad damage and detachment.
FDA · DeviceOlympus Corporation of the AmericasUS Nationwide distribution and International: Algeria, Armenia, Australia,…Official notice →- 2025-01-22
US
Olympus Corporation of the Americas recalls LithoCrushV Single Use Mechanical Lithotriptor V. Model Number: BML-V442QR-30. Single Use Mechanical Lithotriptor V used…
Potential for distal tip tearing of the Mechanical Lithotriptor V.
US
Olympus Corporation of the Americas recalls Olympus MAJ-891 Forceps/Irrigation Plug (Isolated Type). Model/Catalog Number: MAJ-891. This instrument is designed to be…
Potential for inadvertent retention of biomaterial when the reprocessing procedure is not performed as instructed, so as not completely removing and disassembling the device.
- 2024-12-25
US
Olympus Corporation of the Americas recalls Duravent Silicone Ventilation Tube. Model Number: 240075.
Potential that Duravent Silicone Ventilation Tubes may contain incorrect product, specifically, 1.32 mm Tiny T Tubes (model: 70240076) instead of the intended Duravent Silicone Ventilation Tubes.
FDA · DeviceOlympus Corporation of the AmericasAL, AZ, CA, IL, KY, MD, MN, MO, NC, ND, NY, OK, PA, SC, TN, TX, WA, WI.…Official notice →- 2024-12-11
US
Olympus Corporation of the Americas recalls EVIS EXERA III Gastrointestinal Videoscope GIF-1TH190. Model Number: GIF-1TH190
High-level disinfection ("HLD") with Acecide-C disinfectant in the Olympus OER-Pro Automated Endoscope Reprocessor ("AER") has been removed as a compatible reprocessing method from the Instructions for Use when…
FDA · DeviceOlympus Corporation of the AmericasUS Nationwide distribution including in the states of AK, AL, AR, AZ, CA, CO,…Official notice →- 2024-11-27
US
Olympus Corporation of the Americas recalls Brand Name: EVIS EXERA III Gastrointestinal Videoscope GIF-1TH190 Product Name: Olympus GIF-1TH190 Model/Catalog Number:…
To better define the drying procedures described in the IFU to improve the instructions for easier understanding of each drying step after human errors and concerns were identified when completing a Human Factors…
FDA · DeviceOlympus Corporation of the AmericasDomestic: US Nationwide Distribution.Official notice →- 2024-11-20
US
Olympus Corporation of the Americas recalls Brand Name: UroPass Ureteral Access Sheaths Product Name: Ureteral Access Sheaths Model/Catalog Number: 61024BX, 61038BX,…
Exposure of the UroPass product to Ultraviolet ( UV ) Radiation can cause brittleness of the device dilator tip, which may lead to broken dilator tips in the package or in patients during surgical procedures.
FDA · DeviceOlympus Corporation of the AmericasWorldwide distribution - US Nationwide and the countries of Australia, Canada,…Official notice →US
Olympus Corporation of the Americas recalls Brand Name: Olympus Product Name: Cystoscope Outer Sheath Model/Catalog Number: WA22810A Software Version: N/A Product…
Olympus is removing the statement of compatibility with a GreenLight Laser for BPH therapy from the Instruction for Use (IFU) due to the potential for a damaged tip during use of a laser probe.
FDA · DeviceOlympus Corporation of the AmericasUS Nationwide distribution including in the states of AK, AL, AR, AZ, CA, CO,…Official notice →- 2024-11-06
US
Olympus Corporation of the Americas recalls Olympus Electrosurgical Generator, Model Number ESG-410 (SEP) Version WA91327U
There was an increase in complaints indicating an occurrence of system error E1226 "Could not write to Audio Buffer" during procedures using the ESG-410 (SEP).
US
Olympus Corporation of the Americas recalls Quest Spectrum 3 Fluorescent Imaging System Model/Catalog Number: ART-SYS-0029, ART-SYS-0030 The Quest Spectrum3 is a…
Does not meet the requirements for electromagnetic compatibility (EMC) of the standard IEC 60601-1-2. Electromagnetic interference occurring during a procedure may cause the device to malfunction requiring the…
US
Olympus Corporation of the Americas recalls Olympus Endoscopic Flushing Pump, Model/Catalog Number: OFP-2 (K10001141)
During product testing, Olympus identified an intermittent loss of function of the OFP-2 Flushing Pump caused by an internal component connection failure.
- 2024-10-30
US
Olympus Corporation of the Americas recalls EZDilate Wire Guided Balloon- Indicated for use in adult and adolescent (> 12 years) populations to endoscopically dilate…
Mislabeled Glo Cath Label (attached to the device) states: 11 mm, 12 mm, 13 mm however the balloon is 8.5mm, 9.5mm and 10.5 mm and may result in prolonged surgery. All other labeling is correct for the balloon,…
FDA · DeviceOlympus Corporation of the AmericasWorldwide Distribution: US distribution to states of: CA, FL, IL, NJ, NY, PA,…Official notice →US
Olympus Corporation of the Americas recalls Olympus Gastrointestinal Videoscope, Model/Catalog Number: GIF-1100
It was discovered during device performance testing that the CCD imaging sensors were programmed with the incorrect color correction data and therefore, specifications are not met.
US
Olympus Corporation of the Americas recalls Soltive Premium SuperPulsed Laser - Intended for incision, excision, resection, ablation, coagulation, hemostasis, and…
The term Bladder Stone was incorrectly translated in both Spanish and Portuguese to Kidney Stone(Clculo renal) on the systems Graphical User Interface (GUI), a potential to deliver an incorrect or unintended amount of…
FDA · DeviceOlympus Corporation of the AmericasWorldwide distribution - US Nationwide and the countries of AU, CA, CL, DE,…Official notice →- 2024-10-23
US
Olympus Corporation of the Americas recalls EVIS EXERA III Video System Center Model Number CV-190. Intended to be used with OLYMPUS camera heads, endoscopes, light…
Certain serial numbers of the CV-190 do not start up properly because parts that deviated from the specification were assembled into the power supply unit.
- 2024-09-18
US
Olympus Corporation of the Americas recalls Olympus Triangle Tip Electrosurgical Knives- to be used with Olympus endoscopes and electrosurgical units to cut tissue using…
Due to increase in complaints for the KD-640L and KD-645L Triangle Tip Electrosurgical Knives breaking off during use -issuing reminder to users to utilize the electrosurgical knife in accordance with the Instructions…
- 2024-08-14
US
Olympus Corporation of the Americas recalls The Everest Bipolar Cutting Forceps are intended to be passed through a 5-mm or 10-mm cannula, depending on the model/outer…
Fractures and breakages in packaging trays and Tyvek covers, which may result in a sterility breach
FDA · DeviceOlympus Corporation of the AmericasWorldwide distribution - US Nationwide and the countries of CA, DE, AU, SG, KR.Official notice →US
Olympus Corporation of the Americas recalls Soltive Premium SuperPulsed Laser System, Model TFL-PLS
Difficulties in pairing the wireless footswitch with the Soltive Laser, potentially delaying the surgical procedure.
FDA · DeviceOlympus Corporation of the AmericasWorldwide distribution - US Nationwide and the countries of Australia, Canada,…Official notice →- 2024-07-31
US
Olympus Corporation of the Americas recalls ESG-410 Electrosurgical Generator- In conjunction with electrosurgical accessories intended for cutting and coagulation of…
Electrosurgical Generators ESG-410 does not power on upon use, may result in delays in initiating treatment or prolonged surgery
- 2024-07-10
US
Olympus Corporation of the Americas recalls EasySuite 4K Integrated Operating Room System Model/Catalog Number: EasySuite 4K- The EasySuite 4K system includes a touch…
A subset of the ES4K systems do not have complete records for earth leakage testing. The earth leakage test is a series of measurements intended to confirm that the amount of electrical current flowing to the earth is…
FDA · DeviceOlympus Corporation of the AmericasWorldwide distribution - US Nationwide and the countries of Germany, Hong Kong.Official notice →- 2024-06-12
US
Olympus Corporation of the Americas recalls OES 4000 Hysteroscope, Model/Catalog Number: A4674A
The required leakage test was not performed following of a re-work for a non-conformity. This could potentially result in liquid ingress into the eyepiece and cause a foggy image.
FDA · DeviceOlympus Corporation of the AmericasUS Nationwide distribution in the state of NY.Official notice →- 2024-05-29
US
Olympus Corporation of the Americas recalls InstaClear Lens Sheath. The InstaClear lens cleaner system and versatile sheath cleans and clears the lens of a rigid…
Potential for distal end of the sheath to break off during a procedure.
FDA · DeviceOlympus Corporation of the AmericasDomestic: Nationwide Distribution; Foreign: Australia, Canada, Germany, Japan.Official notice →- 2024-05-08
US
Olympus Corporation of the Americas recalls POWERSEAL 5 mm, 44 cm, Curved Jaw Sealer and Divider, Double Action is indicated for use in general surgery and such…
Upon initial activation of the seal function, the POWERSEAL non-conformance results in an immediate Incomplete Seal Cycle tone with the accompanying message on the generator screen, and no energy would be delivered to…
FDA · DeviceOlympus Corporation of the AmericasWorldwide distribution - US Nationwide and the countries of AU, CA, HK, KR.Official notice →- 2024-04-17
US
Olympus Corporation of the Americas recalls Hemostasis Probe, Model: CD-B622LA
The BICOAG Hemostasis Probe with a coaxial plug was mispackaged into the package and labeled as a BICOAG Hemostasis Probe with a fixed pin connector.
- 2024-04-03
US
Olympus Corporation of the Americas recalls SOLTIVE Pro SuperPulsed Laser (TFL-SLS )
A non-compliant power cord may have been supplied in regions that do not utilize the Type B or Type E/F plug configurations.
FDA · DeviceOlympus Corporation of the AmericasDomestic: OUS Only; International distribution in the states of Australia,…Official notice →US
Olympus Corporation of the Americas recalls Insufflation unit, UHI-4.
Olympus has become aware of an increased trend of both repairs and customer complaints reporting "UHI-4 stopping air supply due to the front panel LED turning off".
US
Olympus Corporation of the Americas recalls InstaClear Lens Cleaner-indicated for use during routine diagnostic procedures and during endoscopic sinus surgery. Model…
Japanese IFU version had slightly different language and illustrations from the English IFU, which could contribute to the tip and plate falling off. In order to maintain consistency globally, the Japanese IFU will be…
FDA · DeviceOlympus Corporation of the AmericasInternational distribution to the country of Japan.Official notice →- 2024-03-06
US
Olympus Corporation of the Americas recalls Colonoscope, Model Number PCF-H190DL.
Specific repaired colonovideoscopes were assembled without an adhesive which protects them from certain physical conditions such as high levels of vibration, fluctuation in temperature or severe shock loads.
- 2024-02-28
US
Olympus Corporation of the Americas recalls ESG PK CUTTING FORCEPS, 5MM, 33CM
The jaw may fracture prior to the procedure during the inspection instructed per the IFU, or during the procedure.
FDA · DeviceOlympus Corporation of the AmericasWorldwide distribution - US Nationwide and the countries of Australia, Canada,…Official notice →- 2024-02-21
US
Olympus Corporation of the Americas recalls Single Use Distal Cover MAJ-2315 Used with: EVIS EXERA III Duodenovideoscope TJF-Q190V- designed to be attached to Olympus…
Single Use Distal Cover MAJ-2315 may unexpectedly detach from the duodenoscope during use result include the risk of mucosal injury, tissue damage, bleeding, perforation, or obstruction due to retained distal cover in…
- 2024-02-14
US
Olympus Corporation of the Americas recalls Water Filter MAJ-2318. This MAJ-2318 water filter has been designed to be used with Olympus endoscope reprocessor for prevent…
Water filter may have been assembled incorrectly, resulting in unfiltered tap water being used for reprocessing. This may expose a patient to a potentially contaminated device, presenting a risk of infection or sepsis.
FDA · DeviceOlympus Corporation of the AmericasDomestic US distribution nationwide.Official notice →US
Olympus Corporation of the Americas recalls Olympus/Gyrus Paparrela-Type Vent Tube 1.02MM, Silicone, Sterile- to ventilate the middle ear subsequent to otitis media…
The expiration date is missing from the primary (inner) packaging. The secondary (outer) box label correctly states the expiration date of February 12, 2033, for this lot
FDA · DeviceOlympus Corporation of the AmericasWorldwide distribution - US Nationwide and the countries of Australia.Official notice →- 2024-02-07
US
Olympus Corporation of the Americas recalls EZDilate Wire Guided Balloon 6-7-8. indicated for endoscopic dilation of strictures in the alimentary tract in adults and…
Product balloon issues: inflating, deflating and/or retrieving the devices; reports of bursting and leaking and events involving foreign body in patient and prolonged procedure.
FDA · DeviceOlympus Corporation of the AmericasNationwide. Foreign: AU BR BR CA DE HK IN JP SGOfficial notice →- 2024-01-24
US
Olympus Corporation of the Americas recalls SOLTIVE SuperPulsed Laser Fibers Model Numbers TFL-FBX150S, TFL-FBX200S, TFL-FBX365S, TFL-FBX550S, TFL-FBX940S,…
Olympus identified inconsistencies in the Instructions for Use regarding cleaning the fiber tip and fiber cleaving and stripping instructions. Following the current instructions presents the potential for the fiber to…
FDA · DeviceOlympus Corporation of the AmericasWorldwide distribution - US Nationwide and the countries of Canada, Europe,…Official notice →US
Olympus Corporation of the Americas recalls SOLTIVE SuperPulsed Laser Fibers, Model No. TFL-FBX200BS
Affected products may potentially have a breached sterile pouch seal. A breached sterile pouch seal may lead to minor procedure delays (if noticed prior to use) or patient infections (if not noticed prior to use in…
FDA · DeviceOlympus Corporation of the AmericasDomestic distribution to the following states: AK AR AZ CA FL GA IA ID IN KS…Official notice →US
Olympus Corporation of the Americas recalls Soltive SuperPulsed Laser System TFL Fiber Stripper, Model Numbers TFL-AFS150, TFL-AFS200, TFL-AFS365, TFL-AFS550, TFL-AFS940
The strippers and cleavers are provided non-sterile and labelled as reusable/autoclavable. Olympus does not have validated cleaning and sterilization instructions. Use of a non-sterile fiber stripper or cleaver on a…
- 2024-01-17
US
Olympus Corporation of the Americas recalls Olympus ENDOEYE HD II Rigid Videoscope Catalog Numbers: WA50040A, WA50042A -used with a video system center, light source,…
Reports of a pink or green coloration of the image. If the issue is encountered during a procedure, device exchange could potentially result in prolonged surgery; potential visual impacts mucosal injury or bleeding may…
FDA · DeviceOlympus Corporation of the AmericasWorldwide - US Nationwide distribution.Official notice →US
Olympus Corporation of the Americas recalls OLYMPUS ENDOEYE FLEX DEFLECTABLE VIDEOSCOPE (5.4 mm diameter) Model: LTF-S190-5 Used with a video system center, light…
Instructions for Use-User Deviations related to the reprocessing of the videoscope detailed in the Reprocessing Manual Instructions. The steps are required for the proper reprocessing of these videoscopes
- 2023-12-27
US
Olympus Corporation of the Americas recalls ACMI Roller Bar Electrodes bearing incorrect pouch label of ACMI USA Elite System and USA Series Right Angle Cutting Loop…
Incorrect pouch labels, the contents within the mislabeled pouch were Roller Bar electrodes.
FDA · DeviceOlympus Corporation of the AmericasUS: AL, FL, GA, KS, LA, MD, MI, NC, NJ, NY, TX, WI & WV and Canada.Official notice →- 2023-12-20
US
Olympus Corporation of the Americas recalls TRACHEAL INTUBATION FIBERSCOPE OLYMPUS LF-GP, Model Number: LF-GP
Fiberscopes do not have adequate data to support that the suction flow rate meets the device specification.
FDA · DeviceOlympus Corporation of the AmericasWorldwide - US Nationwide distribution in the states of AK, AL, AZ, CA, GA,…Official notice →- 2023-11-15
US
Olympus Corporation of the Americas recalls EVIS EXERA III Gastrointestinal Videoscope
Sterilization failures when devices are sterilized per the Reprocessing Manual due to lack of drying time for the endoscope channel.
US
Olympus Corporation of the Americas recalls EVIS EXERA III BRONCHOVIDEOSCOPE OLYMPUS Model Number BF-P190
There is a risk of endobronchial combustion if high-frequency cauterization is performed while supplying oxygen [and/or] the electrode section of the electrosurgical accessory is too close to the distal end of the…
- 2023-10-25
US
Olympus Corporation of the Americas recalls Olympus High-Flow Insufflation Unit, Model UHI-4. For laparoscopic surgery.
There have been reports of patients suffering arrhythmias, reported as short cardiac arrests, during surgical procedures where UHI-4s were used. These events may have been due to an over insufflation of the abdominal…
- 2023-10-18
US
Olympus Corporation of the Americas recalls THUNDERBEAT, 5MM, 35CM, Front-actuated Grip Type S. Model: TB-0535FCS Indicated for open, laparoscopic (including…
Thunderbeats blue seal button may remain in the engaged position after the button is released and not immediately return to a neutral position may result in prolonged surgery
- 2023-10-11
US
Olympus Corporation of the Americas recalls Veran : Endobronchial Always on Track Models: INS-0392 AOTT 21ga Needle, 12mm L, 1.8mm OD, 5/BX INS-0382 AOTT 19ga Needle,…
Non-conformances in Veran devices (ig4 Image Guided System, SPiN Thoracic Navigation System, SPiN Planning Laptop Workstation, and all associated navigation instruments) manufacturing to design requirements and…
FDA · DeviceOlympus Corporation of the AmericasWorldwide distribution - US Nationwide and the countries of CN, DE, HK, IT,…Official notice →- 2023-09-06
US
Olympus Corporation of the Americas recalls Olympus Bronchovideoscope, Models BF-3C40 & BF-N20.
Complaint of endoscope model becoming lodged in the endotracheal tube connector due to diameter being too large.
US
Olympus Corporation of the Americas recalls Visera Hysterovideoscope Olympus HYF Type V
IFU labeling is being updated to revise EtO gas mixture and elimination of reusable cleaning brushes.
FDA · DeviceOlympus Corporation of the AmericasUS Nationwide distribution including in the states of AL, AZ, CA, CO, CT, FL,…Official notice →US
Olympus Corporation of the Americas recalls EVIS EXERA III DUODENOVIDEOSCOPE OLYMPUS TJF-Q190V
Recent reports of patient infection.
- 2023-08-30
US
Olympus Corporation of the Americas recalls The Air/Water Valve is provided/used with the following Endoscopes: ULTRASONIC GASTROFIBERSCOPE, ULTRASONIC…
The air/water valve MAJ-1444 used with OER-Pro and OER-Elite may become damaged and result in loss of the one-way valve functionality by repeated automated endoscope reprocessing in Olympus OER machines, causing body…
- 2023-08-23
US
Olympus Corporation of the Americas recalls HF Resection Electrodes - HF resection electrodes are part of a resectoscope system for endoscopic diagnosis and treatment…
Additional Caution statements in IFU: (1) Due to complaints where the loop wire at the distal end of the HF resection electrode broke after getting in contact with metal objects, such as other endoscopic equipment,…
US
Olympus Corporation of the Americas recalls UroPass Ureteral Access Sheaths, 5 pieces/box
Dilator tips may break in the package and in patients during surgical procedures.
FDA · DeviceOlympus Corporation of the AmericasWorldwide distribution - United States Nationwide and the countries of South…Official notice →- 2023-08-16
US
Olympus Corporation of the Americas recalls Tracheal Intubation Fiberscope OLYMPUS LF-P- To be used with an Olympus Light Source, documentation equipment, display…
Revised OER-Pro, OER-Elite, OER-Mini endoscope reprocessor labeling- the LF-V and LF-P are no longer compatible endoscopes for reprocessing in the OERs. The addendum highlights the following changes: 1. Removal of LF-V…
- 2023-08-09
US
Olympus Corporation of the Americas recalls Always-On Tip Tracked Instruments (SPiN Drive instruments) Triple Needle Brush, 12mm L, 1.8mm OD-Intended for use in clinical…
Electromagnetic (EM) sensor tracking malfunction impacts the use of navigation functionality during endobronchial procedures and prompts the system to report a coil break warning message. This could result in a…
FDA · DeviceOlympus Corporation of the AmericasWorldwide Distribution: US (Nationwide) and OUS (foreign) to countries of: CN,…Official notice →- 2023-07-26
US
Olympus Corporation of the Americas recalls Olympus BRONCHOFIBERSCOPE BF Type, Model Numbers BF-PE2, BF-TE2
There have been complaints of endobronchial combustion during therapeutic laser procedures with the bronchoscope.
- 2023-06-28
US
Olympus Corporation of the Americas recalls Single Use Repositionable Clip QUICKCLIP PRO 165 CM 10/BX- Intended for hemostatic metal clip for the GI Tract Catalog…
Complaints on clip deployment occurring during clinical procedures, leading to hazardous situations where clips fail to achieve their expected function or deploy incorrectly, potentially causing harms to patients. The…
- 2023-05-31
US
Olympus Corporation of the Americas recalls ViziShot 2 Single Use Aspiration Needle, 21G- To be used with ultrasound endoscopes for ultrasound guided fine needle…
Product non-conformance pertaining to the angle of the needle does not meet manufacturing specifications and remained unbent, resulting to slight resistance experienced during needle advancement and retraction.
US
Olympus Corporation of the Americas recalls Olympus Single Use Distal Cover Model Number MAJ-2315 used with EVIS EXERA III Duodenovideoscope TJF-Q190V; Item code:…
The Single Use Distal Cover MAJ-2315 may cause mucosal injury and/or become detached from the endoscope during use. Consequences of a detached cover could include the risk of aspiration, inhalation, or obstruction that…
- 2023-04-26
US
Olympus Corporation of the Americas recalls Single Use Rotatable Clip Fixing Device Model HX 201LR 135 - Rotatable, 165cm long, 2.8 scope channel, standard clip Model…
There have been complaints that the clip did not come out of the tube sheath during the procedure.
- 2023-02-22
US
Olympus Corporation of the Americas recalls EVIS EXERA III Colonovideoscope-For endoscopy and endoscopic surgery within the lower digestive tract (including the anus,…
A single CF-Q180AL colonovideoscope was utilized in a veterinary endoscopy procedure in advance of being assigned to a medical facility as a service loaner in error, potential for microbial contamination
FDA · DeviceOlympus Corporation of the AmericasUS Nationwide distribution in the states of CA, FL, GA, NY, OR, VA.Official notice →- 2023-02-15
US
Olympus Corporation of the Americas recalls SPiN Thoracic Navigation System Model: SYS-4000 (SYS-4000D)
A grainy image is observed when using a SPiN Vision Single-Use Flexible Bronchoscope (models INS-7100 or INS-7130) with the SYS-4000 system due to an incorrect DVI cable supplied may result in prolonged…
FDA · DeviceOlympus Corporation of the AmericasFL, GA, IN, MD, MS, NY, OH, PA, RI, SC, VAOfficial notice →- 2022-12-14
US
Olympus Corporation of the Americas recalls HF Resection Electrodes, Model Numbers WA22602S, WA22603S, WA22621S and WA22657S
An incompatible HF cable may be packaged with the HF Resection Electrodes. This cable cannot be connected to the electrosurgical generator and may lead to delay or cancellation of the surgical procedure.
FDA · DeviceOlympus Corporation of the AmericasUS Nationwide domestic distribution.Official notice →- 2022-10-19
US
Olympus Corporation of the Americas recalls Berkeley VC-10 Vacuum Curettage System, Model No. VC-10. For transcervical aspiration of the uterine cavity.
Update to the Instruction for Use (IFU): Olympus is adding a new warning against the use of accessories which are not listed in the IFU. There have been reported adverse events involving the use of collection canisters…
FDA · DeviceOlympus Corporation of the AmericasDomestic distribution nationwide. Foreign distribution to Canada and Japan.Official notice →- 2022-10-05
US
Olympus Corporation of the Americas recalls Olympus POWERSEAL 5MM, 37CM, CURVED JAW SEALER & DIVIDER, DOUBLE-ACTION Model Number: PS-0537CJDA. An electrosurgical device…
An out of specification component preventing the blue retention ring from keeping the rotation knob in place. This may result in a loose rotation collar and retention ring and cause delay of treatment
FDA · DeviceOlympus Corporation of the AmericasNationwide OUS: AUSTRALIA, CANADA , HONG KONG, CHINA, KOREAOfficial notice →- 2022-08-17
US
Olympus Corporation of the Americas recalls ShockPulse-SE Lithotripsy Generator Model/Serial: SPL-S
Mis-wired component-the improperly wired component can result in additional noise on the power supply output. Noise on the ultrasonic input circuit is not consistent with electromagnetic compatibility (EMC) standards…
FDA · DeviceOlympus Corporation of the AmericasUS Nationwide distribution in the states of AL, NM.Official notice →- 2022-07-06
US
Olympus Corporation of the Americas recalls SOLTIVE SuperPulsed Laser Fibers, Model# Labeled TFL-FBX150BS (Carton) - 150 Micron Ball Tip Single Use Fiber Model #…
Device model and size for affected lots of Laser Fibers may be mismatched between the product's carton, immediate pouch packaging, and/or the device.
FDA · DeviceOlympus Corporation of the AmericasNationwide distribution in the US. Foreign distribution worldwide.Official notice →- 2022-06-29
US
Olympus Corporation of the Americas recalls EVIS EXERA II Gastrointestinal Videoscope-indicated for use within the upper digestive tract (including the esophagus,…
Part (A-rubber) intended for a different model of gastrointestinal videoscope may have been used when replacing part on GIF-HQ190 during the last repair at Olympus. An Incorrect A-rubber fits tighter on the GIF-HQ190…
FDA · DeviceOlympus Corporation of the AmericasUS Nationwide distribution in the states of AL, FL, NC, NY.Official notice →US
Olympus Corporation of the Americas recalls Colonovideoscope Model CF-H180AL
A (Camera unit) CCD intended for a different model of colonoscope may have been installed in the CF-H180AL, viewing angle would be narrower than a normal CF-H180AL and the use of EndoTherapy devices may stick out…
FDA · DeviceOlympus Corporation of the AmericasUS Nationwide distribution in the state of KY.Official notice →- 2022-06-22
US
Olympus Corporation of the Americas recalls Olympus URETERO-RENO FIBERSCOPE Model: URF-P6R
Incorrect adhesive used to secure the diopter ring and nameplate ring. Improperly connected diopter ring may result in the loss of adjustment function leading to a blurred field of vision and unintentional contact with…
- 2022-06-01
US
Olympus Corporation of the Americas recalls Olympus DISPOSABLE GRASPING FORCEPS, FG-51D
Forceps do not comply with Olympus standards for the amount of force required to open and close the forceps
- 2022-05-25
US
Olympus Corporation of the Americas recalls Olympus Uretero-reno videoscope, Model No. URF-V2
The bending section of the URF-V2 may crack, resulting in an abnormal shape or abnormal angulation of the bending section.
FDA · DeviceOlympus Corporation of the AmericasUS Nationwide distribution in the state of AZ.Official notice →- 2022-05-18
US
Olympus Corporation of the Americas recalls BF-3C160: EVIS EXERA Bronchovideoscope
Bronchofiberscope Updated instructions on the use of certain chemicals to be used for reprocessing, conditions for ETO gas sterilization, rinsing steps after high-level disinfection, and a recommendation on using…
- 2022-05-11
US
Olympus Corporation of the Americas recalls Ultrasonic Gastrovideoscope, GF-UC140P-AL5. Intended for endoscopic real-time ultrasound imaging.
GF-UC140P-AL5 is no longer a compatible endoscope for reprocessing in the OER-Pro and presents an infection control risk to patients
- 2022-02-23
US
Olympus Corporation of the Americas recalls OER-Elite Endoscope Reprocessor-for use in cleaning and high-level disinfection of heat sensitive Olympus flexible…
The detergent and alcohol tank tubing may have been incorrectly assembled in the OER-Elite. If the tubing is incorrectly assembled, alcohol will be dispensed instead of detergent for the cleaning cycle and detergent…
- 2022-02-09
US
Olympus Corporation of the Americas recalls Spiration Valve System (SVS) Valve in Cartridge -A one-way endobronchial valves indicated for adult patients with shortness…
Mislabeled Patient Charge Label indicating model number SVS-V7-00 (7mm) instead of SVS-V9-00 (9mm)
FDA · DeviceOlympus Corporation of the AmericasUS Nationwide distribution in the states of AZ, CA, MA, NJ, PA, SC, TX, WI.Official notice →- 2021-12-22
US
Olympus Corporation of the Americas recalls EVIS EXERA II DUODENOVIDEOSCOPE OLYMPUS TJF-Q180V Model Number: TJF-Q180V
Use of a TJF-Q180V with adhesive deterioration or other damages may pose a risk of endoscope contamination due to ineffective reprocessing or fluid invasion; contaminated endoscope can present an infection risk to…
- 2021-11-03
US
Olympus Corporation of the Americas recalls ASC PneumoLiner device, Part No. WA90500US (US market only) The Pneumoliner device is intended for use as a multiple…
Due to manufacturing error, the Pneumoliner Bag Distal Tab that exits the Introducer shaft is in the wrong orientation. This results in the user deploying the bag upside down, which will make tissue encapsulation and…
FDA · DeviceOlympus Corporation of the AmericasUS Nationwide distribution to states of: AL, CA, CT, FL, IL, MD, MI, MN, MO,…Official notice →- 2021-10-06
US
Olympus Corporation of the Americas recalls BF-XP60 OES Bronchofiberscope, Model No. BF-XP60
The firm is issuing validated, new reprocessing instructions according to the FDA 2015 guidance, "Reprocessing Medical Devices in Health Care Settings: Validation Methods and Labeling."
FDA · DeviceOlympus Corporation of the AmericasDomestic distribution nationwide. Product also distributed globally.Official notice →