Olympus Corporation of the Americas recalls Single Use Rotatable Clip Fixing Device Model HX 201LR 135 - Rotatable, 165cm long, 2.8 scope channel, standard clip Model…
There have been complaints that the clip did not come out of the tube sheath during the procedure.
Read the official notice →- Date
- 2023-04-26
- Country
- United States
- Brand
- Olympus Corporation of the Americas
- Product
- Single Use Rotatable Clip Fixing Device Model HX 201LR 135 - Rotatable, 165cm long, 2.8 scope channel, standard clip Model HX 201UR 135 -…
- Classification
- Class II
- Units
- US Nationwide Distribution
- Category
- Drugs & medical
- Source
- FDA · Device
Copied automatically from the US Food and Drug Administration. Details may be shortened or out of date; the official notice is the authoritative source. Disclaimer
What to do
- Check the product name, strength and lot number against the official notice.
- Do not stop a prescribed medicine without talking to your doctor or pharmacist first.
- Ask the pharmacy or supplier about a replacement or refund.
General guidance from Recall Atlas. The official notice from the US Food and Drug Administration has the exact models, batch codes and remedy, and it takes priority.
More Olympus Corporation of the Americas recalls
- 2023-06-28
US
Olympus Corporation of the Americas recalls Single Use Repositionable Clip QUICKCLIP PRO 165 CM 10/BX- Intended for hemostatic metal clip for the GI Tract Catalog…
Complaints on clip deployment occurring during clinical procedures, leading to hazardous situations where clips fail to achieve their expected function or deploy incorrectly, potentially causing harms to patients. The…
- 2023-05-31
US
Olympus Corporation of the Americas recalls ViziShot 2 Single Use Aspiration Needle, 21G- To be used with ultrasound endoscopes for ultrasound guided fine needle…
Product non-conformance pertaining to the angle of the needle does not meet manufacturing specifications and remained unbent, resulting to slight resistance experienced during needle advancement and retraction.
US
Olympus Corporation of the Americas recalls Olympus Single Use Distal Cover Model Number MAJ-2315 used with EVIS EXERA III Duodenovideoscope TJF-Q190V; Item code:…
The Single Use Distal Cover MAJ-2315 may cause mucosal injury and/or become detached from the endoscope during use. Consequences of a detached cover could include the risk of aspiration, inhalation, or obstruction that…
- 2023-02-22
US
Olympus Corporation of the Americas recalls EVIS EXERA III Colonovideoscope-For endoscopy and endoscopic surgery within the lower digestive tract (including the anus,…
A single CF-Q180AL colonovideoscope was utilized in a veterinary endoscopy procedure in advance of being assigned to a medical facility as a service loaner in error, potential for microbial contamination
FDA · DeviceOlympus Corporation of the AmericasUS Nationwide distribution in the states of CA, FL, GA, NY, OR, VA.Official notice →- 2023-02-15
US
Olympus Corporation of the Americas recalls SPiN Thoracic Navigation System Model: SYS-4000 (SYS-4000D)
A grainy image is observed when using a SPiN Vision Single-Use Flexible Bronchoscope (models INS-7100 or INS-7130) with the SYS-4000 system due to an incorrect DVI cable supplied may result in prolonged…
FDA · DeviceOlympus Corporation of the AmericasFL, GA, IN, MD, MS, NY, OH, PA, RI, SC, VAOfficial notice →- 2022-12-14
US
Olympus Corporation of the Americas recalls HF Resection Electrodes, Model Numbers WA22602S, WA22603S, WA22621S and WA22657S
An incompatible HF cable may be packaged with the HF Resection Electrodes. This cable cannot be connected to the electrosurgical generator and may lead to delay or cancellation of the surgical procedure.
FDA · DeviceOlympus Corporation of the AmericasUS Nationwide domestic distribution.Official notice →
Other drugs & medical recalls in the United States
- 2023-05-03
US
Methapharm Inc recalls Pantoprazole sodium for Injection 40 mg*/vial, Single-dose Vial, Rx only, For Intravenous Infusion Only, Mfd. for Methapharm, Inc. Coral…
CGMP Deviations; The impacted product and lot number was inadvertently placed into saleable inventory, which does not comply with approved procedures.
- 2023-04-26
US
Baxter Healthcare Corporation recalls ExactaMed Oral Dispenser - 20 mL Clear Pharmacy Pack, REF H9387120; Oral medication syringe
ExataMed 20 mL oral dispenser component of the Clear Pharmacy Pack may have particulate matter.
US
Beckman Coulter, Inc. recalls Beckman Coulter Access Free T3, Triiodothyronine, Free, Catalog #A13422, each box contains 2 reagent packs.
The reagent lot produces elevated results (approximately >30%) in some (15-20%)patient samples that do not correlate with the results of other thyroid markers nor Free T3 results generated on other platforms. This…
FDA · DeviceBeckman Coulter, Inc.Worldwide distribution - US Nationwide, including Puerto Rico and the…Official notice →US
PHILIPS MEDICAL SYSTEMS NEDERLAND B.V. recalls SmartPerfusion, release R1.0, R1.1, R1.1.1, R1.1.5, R1.1.5.1 and R1.1.6
There are technical issues related to signal generation and processing, which can lead to inaccurate presentations.
FDA · DevicePHILIPS MEDICAL SYSTEMS NEDERLAND B.V.Worldwide - US Nationwide distribution.Official notice →US
Becton Dickinson & Co. recalls BD Kiestra InoqulA+ with BD Kiestra" InoqulA+" and BeA Software version 5.1 and 5.1.1 as part of the Icefall A platform. An in vitro…
Upon installation of BD Kiestra" InoqulA+" and BeA version 5.1 and 5.1.1 as part of the Icefall A platform, plate information was not visible in Synapsys" after processing has occurred. This could possibly lead to a…
FDA · DeviceBecton Dickinson & Co.Worldwide - US Nationwide distribution in the states of IL, NY and the…Official notice →US
NeuroLogica Corporation recalls Samsung GM85 Digital X-ray Imaging System-A Digital Diagnostic Mobile X-ray System
Issue related to the operation of the arm latch within the column that supports the arm, potential bodily harm due to a fall of the arm