Drugs & medicalClass II

Olympus Corporation of the Americas recalls HF Resection Electrodes, Model Numbers WA22602S, WA22603S, WA22621S and WA22657S

An incompatible HF cable may be packaged with the HF Resection Electrodes. This cable cannot be connected to the electrosurgical generator and may lead to delay or cancellation of the surgical procedure.

Read the official notice →
Date
2022-12-14
Country
United States
Brand
Olympus Corporation of the Americas
Product
HF Resection Electrodes, Model Numbers WA22602S, WA22603S, WA22621S and WA22657S
Classification
Class II
Units
US Nationwide domestic distribution.
Category
Drugs & medical
Source
FDA · Device

Copied automatically from the US Food and Drug Administration. Details may be shortened or out of date; the official notice is the authoritative source. Disclaimer

What to do

  1. Check the product name, strength and lot number against the official notice.
  2. Do not stop a prescribed medicine without talking to your doctor or pharmacist first.
  3. Ask the pharmacy or supplier about a replacement or refund.

General guidance from Recall Atlas. The official notice from the US Food and Drug Administration has the exact models, batch codes and remedy, and it takes priority.

More Olympus Corporation of the Americas recalls

  1. 2023-04-26
  2. US
    Drugs & medicalClass II

    Olympus Corporation of the Americas recalls Single Use Rotatable Clip Fixing Device Model HX 201LR 135 - Rotatable, 165cm long, 2.8 scope channel, standard clip Model…

    There have been complaints that the clip did not come out of the tube sheath during the procedure.

    FDA · DeviceOlympus Corporation of the AmericasUS Nationwide DistributionOfficial notice →
  3. 2023-02-22
  4. US
    Drugs & medicalClass II

    Olympus Corporation of the Americas recalls EVIS EXERA III Colonovideoscope-For endoscopy and endoscopic surgery within the lower digestive tract (including the anus,…

    A single CF-Q180AL colonovideoscope was utilized in a veterinary endoscopy procedure in advance of being assigned to a medical facility as a service loaner in error, potential for microbial contamination

    FDA · DeviceOlympus Corporation of the AmericasUS Nationwide distribution in the states of CA, FL, GA, NY, OR, VA.Official notice →
  5. 2023-02-15
  6. US
    Drugs & medicalClass II

    Olympus Corporation of the Americas recalls SPiN Thoracic Navigation System Model: SYS-4000 (SYS-4000D)

    A grainy image is observed when using a SPiN Vision Single-Use Flexible Bronchoscope (models INS-7100 or INS-7130) with the SYS-4000 system due to an incorrect DVI cable supplied may result in prolonged…

    FDA · DeviceOlympus Corporation of the AmericasFL, GA, IN, MD, MS, NY, OH, PA, RI, SC, VAOfficial notice →
  7. 2022-10-19
  8. US
    Drugs & medicalClass II

    Olympus Corporation of the Americas recalls Berkeley VC-10 Vacuum Curettage System, Model No. VC-10. For transcervical aspiration of the uterine cavity.

    Update to the Instruction for Use (IFU): Olympus is adding a new warning against the use of accessories which are not listed in the IFU. There have been reported adverse events involving the use of collection canisters…

    FDA · DeviceOlympus Corporation of the AmericasDomestic distribution nationwide. Foreign distribution to Canada and Japan.Official notice →
  9. 2022-10-05
  10. US
    Drugs & medicalClass II

    Olympus Corporation of the Americas recalls Olympus POWERSEAL 5MM, 37CM, CURVED JAW SEALER & DIVIDER, DOUBLE-ACTION Model Number: PS-0537CJDA. An electrosurgical device…

    An out of specification component preventing the blue retention ring from keeping the rotation knob in place. This may result in a loose rotation collar and retention ring and cause delay of treatment

    FDA · DeviceOlympus Corporation of the AmericasNationwide OUS: AUSTRALIA, CANADA , HONG KONG, CHINA, KOREAOfficial notice →
  11. 2022-08-17
  12. US
    Drugs & medicalClass II

    Olympus Corporation of the Americas recalls ShockPulse-SE Lithotripsy Generator Model/Serial: SPL-S

    Mis-wired component-the improperly wired component can result in additional noise on the power supply output. Noise on the ultrasonic input circuit is not consistent with electromagnetic compatibility (EMC) standards…

    FDA · DeviceOlympus Corporation of the AmericasUS Nationwide distribution in the states of AL, NM.Official notice →
All 114 Olympus Corporation of the Americas recalls →

Other drugs & medical recalls in the United States

  1. 2022-12-14
  2. US
    Drugs & medicalClass II

    Philips North America Llc recalls Philips SmartPath to Ingenia Elition X MR System, Model Numbers 782118, 782132, 782144

    Gradient Coil may act as a heat source with a potential to produce smoke and/or fire.

    FDA · DevicePhilips North America LlcUS Nationwide.Official notice →
  3. US
    Drugs & medicalClass II

    Intersect ENT, Inc. recalls Propel Mini Mometasone Furoate Sinus Implant, with Straight Delivery System, REF 60044

    Sinus Implant product label displays incorrect product information including: "Straight Delivery System", with incorrect reference number and UDI-DI.

    FDA · DeviceIntersect ENT, Inc.US Nationwide distribution including in the states of NC, IL, KS, OR, WI, TX,…Official notice →
  4. US
    Drugs & medicalClass II

    Qvella Corporation recalls The FAST System and the FAST PBC Prep Cartridge together are a closed, rapid and automated sample preparation system for isolating…

    Device exceeds the limitations of a Class I 510(k)-exempt device so the user manual, instructions for use, and promotional materials are being corrected to remove all references to downstream applications.

    FDA · DeviceQvella CorporationUS: CA, NY, AL, UTOfficial notice →
  5. US
    Drugs & medicalClass II

    Insulet Corporation recalls Omnipod 5 Automated Insulin Delivery System Product catalog numbers: PT-000409: ASM, Omnipod 5, PDM, Programmed PT-000428: Adaptor,…

    The firm has become aware of an issue with the Omnipod 5 Controller where the Controller charging port and charging cable are melting, deforming, or discoloring due to heat generated by increased resistance in the…

    FDA · DeviceInsulet CorporationDomestic distribution nationwide.Official notice →
  6. US
    Drugs & medicalClass II

    Getinge Usa Sales Inc recalls Flow-c Anesthesia System (Product Code 6887700) Flow-e Anesthesia System (Product Code 6887900)

    The power backup battery, under certain conditions, may exhibit reduced battery run-time while in use, and there is a potential that the battery may fail to provide the expected run time in the event of a mains power…

    FDA · DeviceGetinge Usa Sales IncUS Nationwide distribution in the states of CO, FL, IN, MA, MO, and VA.Official notice →
  7. US
    Drugs & medicalClass II

    Integrity Implants Inc. recalls LineSider Spinal System pedicle screws, 5.5/6.0mm, Non-mod Standard 7.1mm - Dual Lead, Non-cannulated. Model Number LS-N6SG6545.…

    Potential for intraoperative screw head separation at weld location.

    FDA · DeviceIntegrity Implants Inc.US distribution to California, Indiana, New York, Oklahoma, Washington DCOfficial notice →