Olympus Corporation of the Americas recalls ShockPulse-SE Lithotripsy Generator Model/Serial: SPL-S
Mis-wired component-the improperly wired component can result in additional noise on the power supply output. Noise on the ultrasonic input circuit is not consistent with electromagnetic compatibility (EMC) standards…
Read the official notice →- Date
- 2022-08-17
- Country
- United States
- Brand
- Olympus Corporation of the Americas
- Product
- ShockPulse-SE Lithotripsy Generator Model/Serial: SPL-S
- Classification
- Class II
- Units
- US Nationwide distribution in the states of AL, NM.
- Category
- Drugs & medical
- Source
- FDA · Device
Copied automatically from the US Food and Drug Administration. Details may be shortened or out of date; the official notice is the authoritative source. Disclaimer
What to do
- Check the product name, strength and lot number against the official notice.
- Do not stop a prescribed medicine without talking to your doctor or pharmacist first.
- Ask the pharmacy or supplier about a replacement or refund.
General guidance from Recall Atlas. The official notice from the US Food and Drug Administration has the exact models, batch codes and remedy, and it takes priority.
More Olympus Corporation of the Americas recalls
- 2022-12-14
US
Olympus Corporation of the Americas recalls HF Resection Electrodes, Model Numbers WA22602S, WA22603S, WA22621S and WA22657S
An incompatible HF cable may be packaged with the HF Resection Electrodes. This cable cannot be connected to the electrosurgical generator and may lead to delay or cancellation of the surgical procedure.
FDA · DeviceOlympus Corporation of the AmericasUS Nationwide domestic distribution.Official notice →- 2022-10-19
US
Olympus Corporation of the Americas recalls Berkeley VC-10 Vacuum Curettage System, Model No. VC-10. For transcervical aspiration of the uterine cavity.
Update to the Instruction for Use (IFU): Olympus is adding a new warning against the use of accessories which are not listed in the IFU. There have been reported adverse events involving the use of collection canisters…
FDA · DeviceOlympus Corporation of the AmericasDomestic distribution nationwide. Foreign distribution to Canada and Japan.Official notice →- 2022-10-05
US
Olympus Corporation of the Americas recalls Olympus POWERSEAL 5MM, 37CM, CURVED JAW SEALER & DIVIDER, DOUBLE-ACTION Model Number: PS-0537CJDA. An electrosurgical device…
An out of specification component preventing the blue retention ring from keeping the rotation knob in place. This may result in a loose rotation collar and retention ring and cause delay of treatment
FDA · DeviceOlympus Corporation of the AmericasNationwide OUS: AUSTRALIA, CANADA , HONG KONG, CHINA, KOREAOfficial notice →- 2022-07-06
US
Olympus Corporation of the Americas recalls SOLTIVE SuperPulsed Laser Fibers, Model# Labeled TFL-FBX150BS (Carton) - 150 Micron Ball Tip Single Use Fiber Model #…
Device model and size for affected lots of Laser Fibers may be mismatched between the product's carton, immediate pouch packaging, and/or the device.
FDA · DeviceOlympus Corporation of the AmericasNationwide distribution in the US. Foreign distribution worldwide.Official notice →- 2022-06-29
US
Olympus Corporation of the Americas recalls EVIS EXERA II Gastrointestinal Videoscope-indicated for use within the upper digestive tract (including the esophagus,…
Part (A-rubber) intended for a different model of gastrointestinal videoscope may have been used when replacing part on GIF-HQ190 during the last repair at Olympus. An Incorrect A-rubber fits tighter on the GIF-HQ190…
FDA · DeviceOlympus Corporation of the AmericasUS Nationwide distribution in the states of AL, FL, NC, NY.Official notice →US
Olympus Corporation of the Americas recalls Colonovideoscope Model CF-H180AL
A (Camera unit) CCD intended for a different model of colonoscope may have been installed in the CF-H180AL, viewing angle would be narrower than a normal CF-H180AL and the use of EndoTherapy devices may stick out…
FDA · DeviceOlympus Corporation of the AmericasUS Nationwide distribution in the state of KY.Official notice →
Other drugs & medical recalls in the United States
- 2022-08-24
US
HERON THERAPEUTICS, INC. recalls Zynrelef (bupivacaine and meloxicam), 400 mg bupivacaine and 12 mg meloxicam single dose application, packaged in a kit, Rx only,…
Defective Delivery System: An incorrect 10 mL (12 mL) Luer (slip) syringe packaged in one lot of Zynrelef 400 mg/12 mg kit
- 2022-08-17
US
Aesculap Implant Systems LLC recalls Aeos¿ Robotic Digital Microscope with software versions 2.7 or earlier Article Number: PV010
Application error problems: Rebalance scope message: 1.Rebalancing process 2.Usage of Robot Recovery: 3. Autofocus not available:
FDA · DeviceAesculap Implant Systems LLCUS Distribution to states of: AR, IN, NE and NCOfficial notice →US
Roche Diagnostics Operations, Inc. recalls cobas e801 Immunoassay Analyzer
Software issue resulting in signals and sample test results of Pre-Wash tests being impacted. The following assays are considered impacted by the issue: Anti-HAV2, Anti-HBc IgM, IGF-1, Myogloin, Rubella IgG, and Toxo…
FDA · DeviceRoche Diagnostics Operations, Inc.US Nationwide distribution including in the states of AK, AL, AR, AZ, CA, CO,…Official notice →US
Spacelabs Healthcare, Inc. recalls Xhibit Telemetry Receiver, Model: 96280
Software related alarm escalation defect occurs after approximately 25 days of continuous use. Technical alarm escalation for the following alarms, signal loss, all leads off, low battery, SpO2 sensor off, and signal…
FDA · DeviceSpacelabs Healthcare, Inc.Worldwide distribution - US Nationwide distribution in the states of NY, AR,…Official notice →US
GE Healthcare, LLC recalls Service manuals for the CT, PET, or NM table as follows: GT1700/GT1700V/VT1700, GT2000/VT2000, GT2000x/VT2000x, High Capacity Table, Lite…
As a result of unclear guidance in the service manual, the table can drop during a specific service activity if the wrong screws are removed by the service personnel without the appropriate table supports in place…
FDA · DeviceGE Healthcare, LLCUS Nationwide distribution including PR. There was also government/military…Official notice →US
Medical Modeling, Inc. recalls 3D Systems VSP Reconstruction, Model: VSPR-301 Product: Scapula Marking Guide (Anatomically Shaped), Rx Only, Non-Sterile
Scapula Marking Guide (anatomically Shaped) was incorrectly designed.
FDA · DeviceMedical Modeling, Inc.International distribution in the country of Germany.Official notice →