Roche Diagnostics Operations, Inc. recalls cobas e801 Immunoassay Analyzer
Software issue resulting in signals and sample test results of Pre-Wash tests being impacted. The following assays are considered impacted by the issue: Anti-HAV2, Anti-HBc IgM, IGF-1, Myogloin, Rubella IgG, and Toxo…
Read the official notice →- Date
- 2022-08-17
- Country
- United States
- Brand
- Roche Diagnostics Operations
- Product
- cobas e801 Immunoassay Analyzer
- Classification
- Class II
- Units
- US Nationwide distribution including in the states of AK, AL, AR, AZ, CA, CO,…
- Category
- Drugs & medical
- Source
- FDA · Device
Copied automatically from the US Food and Drug Administration. Details may be shortened or out of date; the official notice is the authoritative source. Disclaimer
What to do
- Check the product name, strength and lot number against the official notice.
- Do not stop a prescribed medicine without talking to your doctor or pharmacist first.
- Ask the pharmacy or supplier about a replacement or refund.
General guidance from Recall Atlas. The official notice from the US Food and Drug Administration has the exact models, batch codes and remedy, and it takes priority.
More Roche Diagnostics Operations recalls
- 2026-01-14
US
Roche Diagnostics Operations, Inc. recalls The Elecsys Anti-TSHR immunoassay is a three-step competitive immunoassay with streptavidin-coated microparticles and…
Issues identified: 1) Results vary by Instrument Platform; 2) Calibrator Lot-to-Lot Variability. Issues may lead to delayed or incorrect diagnosis, as well as delay of treatment or incorrect treatment.
FDA · DeviceRoche Diagnostics Operations, Inc.US Nationwide distribution in the states of AZ, CA, FL, GA, IL, IN, ME, MI,…Official notice →- 2024-11-27
US
Roche Diagnostics Operations, Inc. recalls cobas HCYS, Homocysteine Enzymatic Assay, Material Number REF 06542921190, for use on cobas c 701/702, in vitro test for total…
Abnormal calibration signals and control recovery issues on the Homocysteine Enzymatic (HCYS) Assay can produce erroneous patient results (positive or negative bias)
FDA · DeviceRoche Diagnostics Operations, Inc.US: NJ, TX, NY, FL, IN, TX, OK, SC, IA,Official notice →- 2024-10-02
US
Roche Diagnostics Operations, Inc. recalls Creatine Kinase (CK) used on cobas c 311/501/502 and COBAS INTEGRA 400 plus-In vitro test for the quantitative determination…
Creatine Kinase (CK) Assay Abnormal calibrations on the cobas¿ c 311/501/502 and COBAS INTEGRA 400 plus Analyzers may lead to erroneous results and misdiagnosis.
- 2023-11-29
US
Roche Diagnostics Operations, Inc. recalls Tina-quant D-Dimer Test System, Fibrinogen/Fibrin Degradation Products Assay, Material Numbers 03001245322 - (cobas c 311,…
Elevated results were detected with Li-heparin plasma samples when compared to citrated plasma samples.
- 2023-01-18
US
Roche Diagnostics Operations, Inc. recalls cobas infinity central lab, Material Number 07154003001
A complaint investigation revealed that an incorrect behavior relating to the rejection of orders capability of the Host Connectivity Agent (HCA) could occur where the order received from the Laboratory Information…
FDA · DeviceRoche Diagnostics Operations, Inc.Domestic distribution nationwide. Foreign distribution pending.Official notice →- 2021-11-24
US
Roche Diagnostics Operations, Inc. recalls Cobas u 601 urinalysis test system, Catalog number 06334601001 The cobas U 601 urinalysis test system is comprised of the…
A potential risk for false negative nitrite results exists when endogenous creatinine levels are 15,000 mg/L and above. No interference of Nitrite results was observed at creatinine levels up to 9,000 mg/L.
Other drugs & medical recalls in the United States
- 2022-08-24
US
MEDEK LLC recalls M Hand Sanitizer Ethyl Alcohol Antiseptic 80%v/v, Topical Solution 128 oz/3,785 mL, Made in Mexico by: Grupo Plast-Y-Kosas S.A. de C.V.…
Chemical Contamination and Subpotent Drug: FDA analysis found product to contain methanol and below label claim for ethanol.
US
HERON THERAPEUTICS, INC. recalls Zynrelef (bupivacaine and meloxicam), 400 mg bupivacaine and 12 mg meloxicam single dose application, packaged in a kit, Rx only,…
Defective Delivery System: An incorrect 10 mL (12 mL) Luer (slip) syringe packaged in one lot of Zynrelef 400 mg/12 mg kit
- 2022-08-17
US
Aesculap Implant Systems LLC recalls Aeos¿ Robotic Digital Microscope with software versions 2.7 or earlier Article Number: PV010
Application error problems: Rebalance scope message: 1.Rebalancing process 2.Usage of Robot Recovery: 3. Autofocus not available:
FDA · DeviceAesculap Implant Systems LLCUS Distribution to states of: AR, IN, NE and NCOfficial notice →US
Olympus Corporation of the Americas recalls ShockPulse-SE Lithotripsy Generator Model/Serial: SPL-S
Mis-wired component-the improperly wired component can result in additional noise on the power supply output. Noise on the ultrasonic input circuit is not consistent with electromagnetic compatibility (EMC) standards…
FDA · DeviceOlympus Corporation of the AmericasUS Nationwide distribution in the states of AL, NM.Official notice →US
Spacelabs Healthcare, Inc. recalls Xhibit Telemetry Receiver, Model: 96280
Software related alarm escalation defect occurs after approximately 25 days of continuous use. Technical alarm escalation for the following alarms, signal loss, all leads off, low battery, SpO2 sensor off, and signal…
FDA · DeviceSpacelabs Healthcare, Inc.Worldwide distribution - US Nationwide distribution in the states of NY, AR,…Official notice →US
GE Healthcare, LLC recalls Service manuals for the CT, PET, or NM table as follows: GT1700/GT1700V/VT1700, GT2000/VT2000, GT2000x/VT2000x, High Capacity Table, Lite…
As a result of unclear guidance in the service manual, the table can drop during a specific service activity if the wrong screws are removed by the service personnel without the appropriate table supports in place…
FDA · DeviceGE Healthcare, LLCUS Nationwide distribution including PR. There was also government/military…Official notice →