Drugs & medicalClass II

Roche Diagnostics Operations, Inc. recalls The Elecsys Anti-TSHR immunoassay is a three-step competitive immunoassay with streptavidin-coated microparticles and…

Issues identified: 1) Results vary by Instrument Platform; 2) Calibrator Lot-to-Lot Variability. Issues may lead to delayed or incorrect diagnosis, as well as delay of treatment or incorrect treatment.

Read the official notice →
Date
2026-01-14
Country
United States
Brand
Roche Diagnostics Operations
Product
The Elecsys Anti-TSHR immunoassay is a three-step competitive immunoassay with streptavidin-coated microparticles and…
Classification
Class II
Units
US Nationwide distribution in the states of AZ, CA, FL, GA, IL, IN, ME, MI,…
Category
Drugs & medical
Source
FDA · Device

Copied automatically from the US Food and Drug Administration. Details may be shortened or out of date; the official notice is the authoritative source. Disclaimer

What to do

  1. Check the product name, strength and lot number against the official notice.
  2. Do not stop a prescribed medicine without talking to your doctor or pharmacist first.
  3. Ask the pharmacy or supplier about a replacement or refund.

General guidance from Recall Atlas. The official notice from the US Food and Drug Administration has the exact models, batch codes and remedy, and it takes priority.

More Roche Diagnostics Operations recalls

  1. 2026-03-11
  2. US
    Drugs & medicalClass II

    Roche Diagnostics Operations, Inc. recalls cobas pro integrated solutions with cobas c 503 analytical units: cobas pro sample supply unit, material number 08464502001…

    Software defect, which allows the system to accept erroneous, non-monotonous calibrations for Spline-type assays. Cobas pro integrated solutions is an automated analyzer, intended for running qualitative,…

    FDA · DeviceRoche Diagnostics Operations, Inc.US Nationwide distribution.Official notice →
  3. 2024-11-27
  4. US
    Drugs & medicalClass II

    Roche Diagnostics Operations, Inc. recalls cobas HCYS, Homocysteine Enzymatic Assay, Material Number REF 06542921190, for use on cobas c 701/702, in vitro test for total…

    Abnormal calibration signals and control recovery issues on the Homocysteine Enzymatic (HCYS) Assay can produce erroneous patient results (positive or negative bias)

    FDA · DeviceRoche Diagnostics Operations, Inc.US: NJ, TX, NY, FL, IN, TX, OK, SC, IA,Official notice →
  5. 2024-10-02
  6. US
    Drugs & medicalClass II

    Roche Diagnostics Operations, Inc. recalls Creatine Kinase (CK) used on cobas c 311/501/502 and COBAS INTEGRA 400 plus-In vitro test for the quantitative determination…

    Creatine Kinase (CK) Assay Abnormal calibrations on the cobas¿ c 311/501/502 and COBAS INTEGRA 400 plus Analyzers may lead to erroneous results and misdiagnosis.

    FDA · DeviceRoche Diagnostics Operations, Inc.US Nationwide distribution.Official notice →
  7. 2023-11-29
  8. US
    Drugs & medicalClass II

    Roche Diagnostics Operations, Inc. recalls Tina-quant D-Dimer Test System, Fibrinogen/Fibrin Degradation Products Assay, Material Numbers 03001245322 - (cobas c 311,…

    Elevated results were detected with Li-heparin plasma samples when compared to citrated plasma samples.

    FDA · DeviceRoche Diagnostics Operations, Inc.US Nationwide distribution.Official notice →
  9. 2023-01-18
  10. US
    Drugs & medicalClass II

    Roche Diagnostics Operations, Inc. recalls cobas infinity central lab, Material Number 07154003001

    A complaint investigation revealed that an incorrect behavior relating to the rejection of orders capability of the Host Connectivity Agent (HCA) could occur where the order received from the Laboratory Information…

    FDA · DeviceRoche Diagnostics Operations, Inc.Domestic distribution nationwide. Foreign distribution pending.Official notice →
  11. 2022-08-17
  12. US
    Drugs & medicalClass II

    Roche Diagnostics Operations, Inc. recalls cobas e801 Immunoassay Analyzer

    Software issue resulting in signals and sample test results of Pre-Wash tests being impacted. The following assays are considered impacted by the issue: Anti-HAV2, Anti-HBc IgM, IGF-1, Myogloin, Rubella IgG, and Toxo…

    FDA · DeviceRoche Diagnostics Operations, Inc.US Nationwide distribution including in the states of AK, AL, AR, AZ, CA, CO,…Official notice →
All 8 Roche Diagnostics Operations recalls →

Other drugs & medical recalls in the United States

  1. 2026-01-14
  2. US
    Drugs & medicalClass II

    Beckman Coulter, Inc. recalls Beckman Coulter UniCel Dxl 600, Part numbers: A71460, A71461, A30260; Immunoassay Method, Troponin Subunit

    Beckman Coulter has determined that, due to a potential communication issue, the analyzers listed above may not successfully restart following a standard shutdown, even after multiple attempts. If this issue occurs,…

    FDA · DeviceBeckman Coulter, Inc.Worldwide Distribution: US (nationwide) and OUS (foreign) to countries of:…Official notice →
  3. US
    Drugs & medicalClass II

    Meridian Bioscience Inc recalls Revogene C. difficile, in vitro diagnostic assay intended for use with the Revogene instrument, Catalog Number 410300

    The affected lots show a decline in performance over time, which may lead to false-negative results.

    FDA · DeviceMeridian Bioscience IncUS Nationwide distribution and the OUS country of Italy.Official notice →
  4. US
    Drugs & medicalClass II

    Applied Medical Resources Corp recalls Kii Low Profile, Bladed Dual Pack , Qty. 1 5x55mm REF: CTB23

    There is the potential for extended length of the blade obturator after assembly into the cannula and seal housing.

    FDA · DeviceApplied Medical Resources CorpUS: MI, CA, ME, OUS: France Great Britain, GermanyOfficial notice →
  5. US
    Drugs & medicalClass II

    Philips Healthcare (Suzhou) Co., Ltd. recalls Philips Incisive CT

    Potential for incomplete scan due to unstable connection inside of floating sensor.

    FDA · DevicePhilips Healthcare (Suzhou) Co., Ltd.U.S. and U.S. territoriesOfficial notice →
  6. US
    Drugs & medicalClass I

    AVID Medical, Inc. recalls Halyard ORGAN RECOVERY OR PACK, Kit Code: LLOF1000-17.

    Devices are not suitable for organ transplant.

    FDA · DeviceAVID Medical, Inc.US Nationwide distribution in the states of FL, GA and Puerto Rico.Official notice →
  7. US
    Drugs & medicalClass II

    Siemens Medical Solutions USA, Inc recalls LUMINOS Lotus Max, Model 11574100 and Luminos dRF Max, Model 10762471

    A potential issue associated with LUMINOS Lotus Max and Luminos dRF Max systems with software version VF11 with bucky wall stand for mobile detectors and without ceiling stand .When switching quickly between organ…

    FDA · DeviceSiemens Medical Solutions USA, IncU.S. and OUSOfficial notice →
See all drugs & medical recalls →