Olympus Corporation of the Americas recalls SPiN Thoracic Navigation System Model: SYS-4000 (SYS-4000D)
A grainy image is observed when using a SPiN Vision Single-Use Flexible Bronchoscope (models INS-7100 or INS-7130) with the SYS-4000 system due to an incorrect DVI cable supplied may result in prolonged…
Read the official notice →- Date
- 2023-02-15
- Country
- United States
- Brand
- Olympus Corporation of the Americas
- Product
- SPiN Thoracic Navigation System Model: SYS-4000 (SYS-4000D)
- Classification
- Class II
- Units
- FL, GA, IN, MD, MS, NY, OH, PA, RI, SC, VA
- Category
- Drugs & medical
- Source
- FDA · Device
Copied automatically from the US Food and Drug Administration. Details may be shortened or out of date; the official notice is the authoritative source. Disclaimer
What to do
- Check the product name, strength and lot number against the official notice.
- Do not stop a prescribed medicine without talking to your doctor or pharmacist first.
- Ask the pharmacy or supplier about a replacement or refund.
General guidance from Recall Atlas. The official notice from the US Food and Drug Administration has the exact models, batch codes and remedy, and it takes priority.
More Olympus Corporation of the Americas recalls
- 2023-05-31
US
Olympus Corporation of the Americas recalls Olympus Single Use Distal Cover Model Number MAJ-2315 used with EVIS EXERA III Duodenovideoscope TJF-Q190V; Item code:…
The Single Use Distal Cover MAJ-2315 may cause mucosal injury and/or become detached from the endoscope during use. Consequences of a detached cover could include the risk of aspiration, inhalation, or obstruction that…
- 2023-04-26
US
Olympus Corporation of the Americas recalls Single Use Rotatable Clip Fixing Device Model HX 201LR 135 - Rotatable, 165cm long, 2.8 scope channel, standard clip Model…
There have been complaints that the clip did not come out of the tube sheath during the procedure.
- 2023-02-22
US
Olympus Corporation of the Americas recalls EVIS EXERA III Colonovideoscope-For endoscopy and endoscopic surgery within the lower digestive tract (including the anus,…
A single CF-Q180AL colonovideoscope was utilized in a veterinary endoscopy procedure in advance of being assigned to a medical facility as a service loaner in error, potential for microbial contamination
FDA · DeviceOlympus Corporation of the AmericasUS Nationwide distribution in the states of CA, FL, GA, NY, OR, VA.Official notice →- 2022-12-14
US
Olympus Corporation of the Americas recalls HF Resection Electrodes, Model Numbers WA22602S, WA22603S, WA22621S and WA22657S
An incompatible HF cable may be packaged with the HF Resection Electrodes. This cable cannot be connected to the electrosurgical generator and may lead to delay or cancellation of the surgical procedure.
FDA · DeviceOlympus Corporation of the AmericasUS Nationwide domestic distribution.Official notice →- 2022-10-19
US
Olympus Corporation of the Americas recalls Berkeley VC-10 Vacuum Curettage System, Model No. VC-10. For transcervical aspiration of the uterine cavity.
Update to the Instruction for Use (IFU): Olympus is adding a new warning against the use of accessories which are not listed in the IFU. There have been reported adverse events involving the use of collection canisters…
FDA · DeviceOlympus Corporation of the AmericasDomestic distribution nationwide. Foreign distribution to Canada and Japan.Official notice →- 2022-10-05
US
Olympus Corporation of the Americas recalls Olympus POWERSEAL 5MM, 37CM, CURVED JAW SEALER & DIVIDER, DOUBLE-ACTION Model Number: PS-0537CJDA. An electrosurgical device…
An out of specification component preventing the blue retention ring from keeping the rotation knob in place. This may result in a loose rotation collar and retention ring and cause delay of treatment
FDA · DeviceOlympus Corporation of the AmericasNationwide OUS: AUSTRALIA, CANADA , HONG KONG, CHINA, KOREAOfficial notice →
Other drugs & medical recalls in the United States
- 2023-02-15
US
Bard Peripheral Vascular Inc recalls The Covera Vascular Covered Stent is a flexible, self-expanding endoprothesis comprised of ePTFE encapsulating a nitinol stent…
An increase in the reported complaint rate for inability to deploy was identified and related to poor slide block bonding.
FDA · DeviceBard Peripheral Vascular IncWorldwide - U.S. Nationwide distribution including in the states of AK, AL,…Official notice →US
Trackx Technology Inc recalls TrackX Insight Base Modular Instrument Tracker, 5L, REF #604-3TX, Sterile; and TrackX Insight Base Modular Instrument Tracker, 6.5L, Model…
The device is breaking and falling off the Nuvasive T-Handle with Hudson Adapter on certain Nuvasive XLIF instruments during surgery when malleted.
FDA · DeviceTrackx Technology IncUS Nationwide distribution in the states of AZ, FL, LA, NC, PA, VA, WA, WI.Official notice →US
Roche Molecular Systems, Inc. recalls cobas SARS-CoV-2 & Influenza A/B Qualitative nucleic acid test for use on the cobas 5800/6800/8800 Systems; Material Number…
The firm received customer complaints regarding false negative Influenza A (Flu A) results and late Flu A Target Ct values. A false negative influenza A result may lead to additional testing, psychological distress,…
FDA · DeviceRoche Molecular Systems, Inc.Domestic distribution nationwide. Foreign distribution worldwide.Official notice →US
Wishbone Medical, Inc. recalls WishBone Medical Smart Correction Dual Joint Express Struts, external skeleton fixation device Type Part Number XXS 10-702-080095-0 XS…
Firm identified that user may be unable to adjust the length of the Smart Correction Dual Joint Express Struts due to the collar release pin backing out. This may lead to a delay in correction.
FDA · DeviceWishbone Medical, Inc.US Nationwide distribution in the state of South Carolina.Official notice →US
TriMed Inc. recalls RipCord Syndesmosis Button, REF: STA001K
Overly aggressive tension on the pull suture, while passing the medial button through a bone/soft tissue tunnel across the fibula and tibia, can exceed its strength causing it to break.
FDA · DeviceTriMed Inc.US Nationwide distribution in the states of IL, PA, TX, GA, TN, AL, MO, FL,…Official notice →US
Quest Medical, Inc. recalls Precision Delivery Set with Filter, REF: PD06F; and Precision Delivery Set with Trifurcated Luer, REF: PDT05CV2
Delivery Sets may have tubing that could become detached that could result in patient infection.