Drugs & medicalClass II

Roche Molecular Systems, Inc. recalls cobas SARS-CoV-2 & Influenza A/B Qualitative nucleic acid test for use on the cobas 5800/6800/8800 Systems; Material Number…

The firm received customer complaints regarding false negative Influenza A (Flu A) results and late Flu A Target Ct values. A false negative influenza A result may lead to additional testing, psychological distress,…

Read the official notice →
Date
2023-02-15
Country
United States
Brand
Roche Molecular Systems
Product
cobas SARS-CoV-2 & Influenza A/B Qualitative nucleic acid test for use on the cobas 5800/6800/8800 Systems; Material Number 09446125190
Classification
Class II
Units
Domestic distribution nationwide. Foreign distribution worldwide.
Category
Drugs & medical
Source
FDA · Device

Copied automatically from the US Food and Drug Administration. Details may be shortened or out of date; the official notice is the authoritative source. Disclaimer

What to do

  1. Check the product name, strength and lot number against the official notice.
  2. Do not stop a prescribed medicine without talking to your doctor or pharmacist first.
  3. Ask the pharmacy or supplier about a replacement or refund.

General guidance from Recall Atlas. The official notice from the US Food and Drug Administration has the exact models, batch codes and remedy, and it takes priority.

More Roche Molecular Systems recalls

  1. 2022-10-26
  2. US
    Drugs & medicalClass II

    Roche Molecular Systems, Inc. recalls cobas 5800 instrument, Material No. 08707464001. Used for automated Polymerase Chain Reaction (PCR) based Nucleic Acid Testing…

    Reported false positive and invalid results on the affected devices due to anomalous baselines. This may lead to erroneous or delayed diagnoses.

    FDA · DeviceRoche Molecular Systems, Inc.International distribution in the countries of Austria, Belgium, Czech…Official notice →
  3. 2022-04-06
  4. US
    Drugs & medicalClass II

    Roche Molecular Systems, Inc. recalls cobas LiatSystem, respiratory virus panel nucleic acid assay system, Material Number 07341920190

    Roche has identified a small number of cobas Liat analyzer units that have consistently high noise levels (background signal) originating from the amber detector (used for Influenza B detection). This noise increases…

    FDA · DeviceRoche Molecular Systems, Inc.Worldwide distribution - US Nationwide distribution in the states of CO, IL,…Official notice →
  5. 2021-10-20
  6. US
    Drugs & medicalClass II

    Roche Molecular Systems, Inc. recalls cobas EGFR Mutation Test v2-Somatic gene mutation detection system Catalog Number :07248563190

    False Mutation Detected results for the exon 20 insertion (Ex20Ins) mutation when using the cobas EGFR Mutation Test v2, may lead to inappropriate treatment

    FDA · DeviceRoche Molecular Systems, Inc.Worldwide distribution - US Nationwide distribution and the countries of…Official notice →
All 4 Roche Molecular Systems recalls →

Other drugs & medical recalls in the United States

  1. 2023-02-22
  2. US
    Drugs & medicalClass II

    McGuff Compounding Pharmacy Services, Inc. recalls Thiamine Hydrochloride 200 mg/2mL (100mg/mL), Riboflavin 2 mg/2mL (1mg/mL) Injection, FOR INTRAMUSCULAR OR SLOW…

    Presence of Particulate Matter: Presence of foreign substances were reported in vials at the pharmacy.

    FDA · DrugMcGuff Compounding Pharmacy Services, Inc.Nationwide in the USAOfficial notice →
  3. 2023-02-15
  4. US
    Drugs & medicalClass II

    Bard Peripheral Vascular Inc recalls The Covera Vascular Covered Stent is a flexible, self-expanding endoprothesis comprised of ePTFE encapsulating a nitinol stent…

    An increase in the reported complaint rate for inability to deploy was identified and related to poor slide block bonding.

    FDA · DeviceBard Peripheral Vascular IncWorldwide - U.S. Nationwide distribution including in the states of AK, AL,…Official notice →
  5. US
    Drugs & medicalClass II

    Trackx Technology Inc recalls TrackX Insight Base Modular Instrument Tracker, 5L, REF #604-3TX, Sterile; and TrackX Insight Base Modular Instrument Tracker, 6.5L, Model…

    The device is breaking and falling off the Nuvasive T-Handle with Hudson Adapter on certain Nuvasive XLIF instruments during surgery when malleted.

    FDA · DeviceTrackx Technology IncUS Nationwide distribution in the states of AZ, FL, LA, NC, PA, VA, WA, WI.Official notice →
  6. US
    Drugs & medicalClass II

    Olympus Corporation of the Americas recalls SPiN Thoracic Navigation System Model: SYS-4000 (SYS-4000D)

    A grainy image is observed when using a SPiN Vision Single-Use Flexible Bronchoscope (models INS-7100 or INS-7130) with the SYS-4000 system due to an incorrect DVI cable supplied may result in prolonged…

    FDA · DeviceOlympus Corporation of the AmericasFL, GA, IN, MD, MS, NY, OH, PA, RI, SC, VAOfficial notice →
  7. US
    Drugs & medicalClass II

    Wishbone Medical, Inc. recalls WishBone Medical Smart Correction Dual Joint Express Struts, external skeleton fixation device Type Part Number XXS 10-702-080095-0 XS…

    Firm identified that user may be unable to adjust the length of the Smart Correction Dual Joint Express Struts due to the collar release pin backing out. This may lead to a delay in correction.

    FDA · DeviceWishbone Medical, Inc.US Nationwide distribution in the state of South Carolina.Official notice →
  8. US
    Drugs & medicalClass II

    TriMed Inc. recalls RipCord Syndesmosis Button, REF: STA001K

    Overly aggressive tension on the pull suture, while passing the medial button through a bone/soft tissue tunnel across the fibula and tibia, can exceed its strength causing it to break.

    FDA · DeviceTriMed Inc.US Nationwide distribution in the states of IL, PA, TX, GA, TN, AL, MO, FL,…Official notice →
See all drugs & medical recalls →