Beckman Coulter, Inc. recalls Beckman Coulter Access Free T3, Triiodothyronine, Free, Catalog #A13422, each box contains 2 reagent packs.
The reagent lot produces elevated results (approximately >30%) in some (15-20%)patient samples that do not correlate with the results of other thyroid markers nor Free T3 results generated on other platforms. This…
Read the official notice →- Date
- 2023-04-26
- Country
- United States
- Brand
- Beckman Coulter
- Product
- Beckman Coulter Access Free T3, Triiodothyronine, Free, Catalog #A13422, each box contains 2 reagent packs.
- Classification
- Class II
- Units
- Worldwide distribution - US Nationwide, including Puerto Rico and the…
- Category
- Drugs & medical
- Source
- FDA · Device
Copied automatically from the US Food and Drug Administration. Details may be shortened or out of date; the official notice is the authoritative source. Disclaimer
What to do
- Check the product name, strength and lot number against the official notice.
- Do not stop a prescribed medicine without talking to your doctor or pharmacist first.
- Ask the pharmacy or supplier about a replacement or refund.
General guidance from Recall Atlas. The official notice from the US Food and Drug Administration has the exact models, batch codes and remedy, and it takes priority.
More Beckman Coulter recalls
- 2023-07-05
US
Beckman Coulter, Inc. recalls Access Hybritech p2PSA reagent, REF B03704 and REF A49752.
Reagent lot numbers include an insufficient concentration of blocking reagent.
FDA · DeviceBeckman Coulter, Inc.Distribution was made to CA, CO, CT, FL, GA, IL, IN, MA, MI, NC, NE, NJ, NY,…Official notice →- 2023-05-24
US
Beckman Coulter Inc. recalls Beckman Coulter Bicarbonate, REF: OSR6137, 4x25 mL
There is a potential for a decrease in calibration ODs for Bicarbonate reagent which could result in calibration failures and/or quality control failures.
FDA · DeviceBeckman Coulter Inc.Worldwide Distribution: US (nationwide): AK, AL, AR, AZ, CA, CO, CT, DC, DE,…Official notice →US
Beckman Coulter, Inc. recalls DxU 840m Iris Urine Microscopy Analyzer, Catalog Number C76947, in-vitro diagnostic device, urine analysis
Beckman Coulter has become aware of an intermittent issue in which the optional flag "Previous Sample Had Sperm" was enabled but not displayed so that a carryover event could have been investigated.
- 2023-03-22
US
Beckman Coulter, Inc. recalls Access Immunoassay Systems (1) Access 2 Immunoassay Analyzer (81600N); (2) Access 2 Immunoassay Analyzer, Refurbished (386220); (3) Access…
An increased incidence of damage to the nut band from a new vendor (Boamax) has led to an increased volume of complaints from consignees regarding wash arm motion failures, system check failures, and/or assay…
FDA · DeviceBeckman Coulter, Inc.Worldwide and Nationwide Distribution. There was government/military…Official notice →- 2023-01-11
US
Beckman Coulter, Inc. recalls BECKMAN COULTER IRISpec CA/CB/CC control, REF 900-7211, For in vitro diagnostic use only.
Beckman Coulter has become aware of an increase in customer complaints generated for intermittent IRISpec CB glucose control false positive results when used in conjunction with iChemVelocity Urine Chemistry Strips PN…
FDA · DeviceBeckman Coulter, Inc.US and Canada, Brazil, Korea, Mexico, Puerto Rico, TaiwanOfficial notice →- 2022-11-30
US
Beckman Coulter, Inc. recalls MicroScan autoSCAN-4 Instrument, Catalog Number B1018-280
Due to MicroScan autoSCAN-4 Instruments being manufactured with an incorrect diffuser plate.
FDA · DeviceBeckman Coulter, Inc.Worldwide - U.S. Nationwide distribution in the states of AK, FL, IA, ID, MI…Official notice →
Other drugs & medical recalls in the United States
- 2023-05-03
US
Wellness BioSciences recalls CBD Metered Dose Inhaler (CBD) 5 mg Dose, 1 Metered Dose Inhaler, 100 Metered Sprays, Wellness BioSciences Rx
Marketed Without an Approved NDA/ANDA
US
Methapharm Inc recalls Pantoprazole sodium for Injection 40 mg*/vial, Single-dose Vial, Rx only, For Intravenous Infusion Only, Mfd. for Methapharm, Inc. Coral…
CGMP Deviations; The impacted product and lot number was inadvertently placed into saleable inventory, which does not comply with approved procedures.
- 2023-04-26
US
Baxter Healthcare Corporation recalls ExactaMed Oral Dispenser - 20 mL Clear Pharmacy Pack, REF H9387120; Oral medication syringe
ExataMed 20 mL oral dispenser component of the Clear Pharmacy Pack may have particulate matter.
US
Olympus Corporation of the Americas recalls Single Use Rotatable Clip Fixing Device Model HX 201LR 135 - Rotatable, 165cm long, 2.8 scope channel, standard clip Model…
There have been complaints that the clip did not come out of the tube sheath during the procedure.
US
PHILIPS MEDICAL SYSTEMS NEDERLAND B.V. recalls SmartPerfusion, release R1.0, R1.1, R1.1.1, R1.1.5, R1.1.5.1 and R1.1.6
There are technical issues related to signal generation and processing, which can lead to inaccurate presentations.
FDA · DevicePHILIPS MEDICAL SYSTEMS NEDERLAND B.V.Worldwide - US Nationwide distribution.Official notice →US
Becton Dickinson & Co. recalls BD Kiestra InoqulA+ with BD Kiestra" InoqulA+" and BeA Software version 5.1 and 5.1.1 as part of the Icefall A platform. An in vitro…
Upon installation of BD Kiestra" InoqulA+" and BeA version 5.1 and 5.1.1 as part of the Icefall A platform, plate information was not visible in Synapsys" after processing has occurred. This could possibly lead to a…
FDA · DeviceBecton Dickinson & Co.Worldwide - US Nationwide distribution in the states of IL, NY and the…Official notice →