Beckman Coulter recalls

84 official recalls and safety alerts involving Beckman Coulter since 2021-10, from FDA · Device. Newest first. Check the official notice for exact models and batch codes.

This list only includes recalls Recall Atlas has collected since 2021-10. A product not listed here may still be affected; check with Beckman Coulter or the relevant agency.

  1. 2026-10-14
  2. US
    Drugs & medicalClass II

    Beckman Coulter Inc. recalls DxC 500 AU Clinical Chemistry Analyzers: 1. AU w/out ISE, Part No. C63519 2. AU w/ISE, Part No. C63520

    Beckman Coulter has identified a software issue affecting DxC 500 AU Clinical Analyzers when testing plasma samples. Following user configuration of a Clinical Chemistry test to use units different from the system…

    FDA · DeviceBeckman Coulter Inc.Worldwide - US Nationwide distribution.Official notice →
  3. US
    Drugs & medicalClass II

    Beckman Coulter Inc. recalls DxC 500i Clinical Analyzer Modules: 1. DxC 500 AU Module w/ISE, DxC 500i, Part No. C63522 2. DxC 500 AU Module, DxC 500i, Part No. C63521…

    Beckman Coulter has identified a software issue affecting DxC 500i Clinical Analyzers when testing plasma samples. Following user configuration of a Clinical Chemistry test to use units different from the system…

    FDA · DeviceBeckman Coulter Inc.Worldwide - US Nationwide distribution.Official notice →
  4. 2026-08-26
  5. US
    Drugs & medicalClass II

    Beckman Coulter Inc. recalls Beckman Coulter DxC 500 AU Clinical Analyzer Modules using software version 1.5 or lower: 1) AU w/out ISE, Part Number C63519; 2) AU…

    a software defect in which the Parameter Package (PP) import logic overwrites all predefined calculated test rules, including those previously modified by the user, upon import of a new Parameter Package version. As a…

    FDA · DeviceBeckman Coulter Inc.Worldwide distribution - US Nationwide.Official notice →
  6. 2026-08-12
  7. US
    Drugs & medicalClass II

    Beckman Coulter Inc. recalls REF # D23354, MicroScan LabPro V5.03 Panel Update 11 DVD Kit (US) on the following instruments: - B1018-280 MicroScan autoSCAN-4…

    Analyzers contain incorrect algorithms which may affect antimicrobial susceptibility test results, leading to false susceptibility classifications."

    FDA · DeviceBeckman Coulter Inc.US Nationwide distribution including in the states of AK, AL, AR, CA, CO, DE,…Official notice →
  8. 2026-04-29
  9. US
    Drugs & medicalClass II

    Beckman Coulter, Inc. recalls Access Total T4 Calibrator, Catalog No. 33805

    Beckman Coulter has identified that Access Total T4 Calibrator lots (PN 33805) exhibit a negative bias on DxI 600/800 instruments, with slopes outside product specifications. A false low patient result could cause a…

    FDA · DeviceBeckman Coulter, Inc.Worldwide distribution - US Nationwide and the countries of Argentina,…Official notice →
  10. 2026-04-15
  11. US
    Drugs & medicalClass II

    Beckman Coulter, Inc. recalls MicroScan Neg Multidrug Resistant MIC 1 REF C32368 UDI-DI code: 15099590720568 Intended Use- For use with MicroScan Dried Gram Negative…

    Due to manufacturing issue (incorrectly processing) with in vitro multidrug resistant test.

    FDA · DeviceBeckman Coulter, Inc.International distribution to the country of Poland.Official notice →
  12. US
    Drugs & medicalClass II

    Beckman Coulter, Inc. recalls MicroScan Neg MIC 3J REF C54814

    Due to a manufacturing issue - missing a drug in well 4/4 but contain double the amount of antibiotics

    FDA · DeviceBeckman Coulter, Inc.International distribution in the country of Japan.Official notice →
  13. 2026-04-01
  14. US
    Drugs & medicalClass II

    Beckman Coulter, Inc. recalls MicroScan Neg MIC 56 REF C42464 UDI-DI code: 15099590731212 MicroScan panels are designed for use in determining antimicrobial agent…

    Due to the likely presence of contamination in well(s).

    FDA · DeviceBeckman Coulter, Inc.U.S. Nationwide distribution in the states of CA, CO, IN, MA, MD, MN, MO, NC,…Official notice →
  15. 2026-02-11
  16. US
    Drugs & medicalClass II

    Beckman Coulter, Inc. recalls CellMek SPS Sample Preparation System, REF: C44603, with software

    Sample preparation system designed to automate staining, lysing, cell washing has a software error that causes samples to be dispensed close to or touching tube bottoms, which could dislodge fluid line, result in…

    FDA · DeviceBeckman Coulter, Inc.Worldwide - US Nationwide distribution in the states of MO, WA, MA, IL, AL,…Official notice →
  17. US
    Drugs & medicalClass II

    Beckman Coulter, Inc. recalls COULTER DxH Diluent, REF 628017 COULTER DxH ECO Diluent REF C67250 For use as an isotonic buffered diluent in conjunction with a…

    Due to specific lots of diluents contributing to elevated platelet Daily Checks background counts cycle with use with analyzers.

    FDA · DeviceBeckman Coulter, Inc.Worldwide - U.S. Nationwide distribution including in the states of AL, AR,…Official notice →
  18. 2026-01-14
  19. US
    Drugs & medicalClass II

    Beckman Coulter, Inc. recalls Beckman Coulter UniCel Dxl 600, Part numbers: A71460, A71461, A30260; Immunoassay Method, Troponin Subunit

    Beckman Coulter has determined that, due to a potential communication issue, the analyzers listed above may not successfully restart following a standard shutdown, even after multiple attempts. If this issue occurs,…

    FDA · DeviceBeckman Coulter, Inc.Worldwide Distribution: US (nationwide) and OUS (foreign) to countries of:…Official notice →
  20. 2025-12-17
  21. US
    Drugs & medicalClass II

    Beckman Coulter, Inc. recalls Access 2 Reaction Vessels, individual, disposable, plastic tubes used to process patient, quality control and calibration samples on the…

    It has been determined that certain lots of Access 2 Reaction Vessels may contain manufacturing deformities. The bottom (round end) of the Reaction Vessel may have a small protrusion or hair like deformity. This could…

    FDA · DeviceBeckman Coulter, Inc.US Nationwide. Global Distribution.Official notice →
  22. 2025-11-26
  23. US
    Drugs & medicalClass II

    Beckman Coulter Inc. recalls DxFLEX Flow Cytometer Catalog Number C78500 UDI-DI code: 15099590754013 The DxFLEX Flow Cytometer is intended for use as an in vitro…

    Due to a software error code that crashes during acquisition on patient sample when running customer defined acquisition protocol. This may lead to delayed patient results.

    FDA · DeviceBeckman Coulter Inc.Worldwide - U.S. Nationwide distribution in the states of CA, CT, FL, IL, NY,…Official notice →
  24. 2025-11-12
  25. US
    Drugs & medicalClass II

    Beckman Coulter Inc. recalls SYNCHRON Systems Direct LDL Cholesterol (LDLD) Reagent, REF: 969706

    Beckman Coulter has identified that there is an increased probability Synchron Systems Direct LDL Cholesterol (LDLD REF 969706) reagent lot M404166 may generate false high results. The issue was discovered internally…

    FDA · DeviceBeckman Coulter Inc.Worldwide - US Nationwide distribution in the states of AL, AR, CA, FL, GA,…Official notice →
  26. US
    Drugs & medicalClass II

    Beckman Coulter, Inc. recalls Beckman Coulter UniCel DxI 600, Part Numbers: A71460, A30260; Photometric Chemistry Analyzer

    An internal investigation has determined that due to defective sub-components within the gantry assembly, a z-axis CAD movement motion error may occur and cause an analyzer to enter a Not Ready or Paused state, leading…

    FDA · DeviceBeckman Coulter, Inc.WorldwideOfficial notice →
  27. 2025-11-05
  28. US
    Drugs & medicalClass II

    Beckman Coulter, Inc. recalls CellMek SPS Sample Preparation System, REF: C44603. Software versions 2.2.122.0 and prior, and Instructions for Use (IFU) revisions AD and…

    Automated pipetting, diluting and specimen processing workstations for flow cytometric analysis device instructions for use list incorrect dead volumes: Cassette Type/List volume/correct volume, B/750/1582 microliters,…

    FDA · DeviceBeckman Coulter, Inc.Worldwide - US Nationwide distribution in the states of AL, AR, FL, IL, IN,…Official notice →
  29. 2025-10-29
  30. US
    Drugs & medicalClass II

    Beckman Coulter, Inc. recalls DxI 9000 Access Immunoassay Analyzer, Part Number C11137

    Beckman Coulter has confirmed a software anomaly on DxI 9000 Access Immunoassay Analyzer and the DxC 500i Clinical Analyzer that may prevent correct application of user-specified manual dilution factors when ordering…

    FDA · DeviceBeckman Coulter, Inc.Worldwide - US Nationwide distribution.Official notice →
  31. 2025-10-15
  32. US
    Drugs & medicalClass II

    Beckman Coulter Inc. recalls SYNCHRON Systems Phosphorus (PHOSm) Reagent REF: 467868 For In Vitro Diagnostic Use, Rx Only.

    Specific SYNCHRON Systems Modular Chemistry Phosphorus (PHOSm REF 467868) reagent lots may fail calibration before the lot expiration date which causes a delay in obtaining patient results.

    FDA · DeviceBeckman Coulter Inc.US: AL KY LA MA MD MS NC NE NJ NY OK PA TX WA WV OUS: Canada Mexico SingaporeOfficial notice →
  33. 2025-10-08
  34. US
    Drugs & medicalClass II

    Beckman Coulter Inc. recalls DxC 500 AU Module w/ISE, DxC 500i, With ISE, REF: C63522 and DxC 500 AU Module, DxC 500i, Without ISE, REF: C6352; part of the DxC 500i…

    On a clinical analyzer, when ordering a Clinical Chemistry (CC) combination test simultaneously with an Immunoassay (IA) test, if the IA test is processed between the CC tests, CC sample status will remain In Progress…

    FDA · DeviceBeckman Coulter Inc.Internationally distribution to the countries of Poland, India, Dominican…Official notice →
  35. 2025-09-24
  36. US
    Drugs & medicalClass II

    Beckman Coulter Inc. recalls UniCel DxH 800 COULTER Cellular Analysis System, configured with the Led HGB Photometer REFs:B24802, 629029, B63322

    Hematology analyzers configured with HGB photometers may result in erroneously high hemoglobin (HGB) results on samples with elevated white blood cell (WBC) counts.

    FDA · DeviceBeckman Coulter Inc.Worldwide - US Nationwide distribution including in the states of AL, AZ, CA,…Official notice →
  37. 2025-08-20
  38. US
    Drugs & medicalClass II

    Beckman Coulter, Inc. recalls Access Thyroglobulin assay, a simultaneous one-step immunoenzymatic assay, Catalog Number 33860

    Access Thyroglobulin reagent lot 439163 may generate erroneously high patient results. Falsely increased Access Thyroglobulin results may lead a physician to pursue unnecessary diagnostic imaging studies and/or…

    FDA · DeviceBeckman Coulter, Inc.Worldwide - US Nationwide and the countries of Brazil, Canada, Georgia, Italy,…Official notice →
  39. 2025-08-13
  40. US
    Drugs & medicalClass II

    Beckman Coulter, Inc. recalls DxI 800 Access Immunoassay Analyzer W/Dual Gantry, Part Number A71457

    Beckman Coulter has identified that the UniCel DxI 600/800 systems with linear slide PnP gantries running software versions 5.7 and up faced incorrect PnP X-motor current settings where the PnP X-motor current…

    FDA · DeviceBeckman Coulter, Inc.Worldwide distribution - US Nationwide.Official notice →
  41. US
    Drugs & medicalClass II

    Beckman Coulter Inc. recalls UIBC (Unsaturated Iron Binding Capacity), REF: OSR61205,

    Beckman Coulter identified that the Unsaturated Iron Binding Capacity (UIBC) assay is not meeting their labeled hemoglobin interference claims of no significant interference (i.e., less than 10%) up to 200 mg/dL…

    FDA · DeviceBeckman Coulter Inc.US: AK, AL, AR, CO, CT, DC, DE, ID, UT OUS: Algeria, Argentina, Australia,…Official notice →
  42. 2025-07-30
  43. US
    Drugs & medicalClass II

    Beckman Coulter Inc. recalls Estrone RIA, REF: DSL8700. Estrone RIA is an IVD used for the quantitative measurement of estrone in human serum and plasma.

    A specific lot of bovine serum albumin (BSA) used for manufacturing of coated tubes (CT) that are included in the Estrone RIA kit lots caused different affinity of manufactured coated tubes to patient samples, causing…

    FDA · DeviceBeckman Coulter Inc.US: North Carolina OUS: Argentina, Austria, Belgium, Canada, France, Germany,…Official notice →
  44. 2025-07-23
  45. US
    Drugs & medicalClass II

    Beckman Coulter, Inc. recalls Access Vitamin B12 Reagent, Catalog Number 33000, Lot Number 439850

    Beckman Coulter has received complaints of erroneous Vitamin B12 results when using Access Vitamin B12 Reagent lot 439850. The affected reagent lot may cause erroneously low or high Vitamin B12 results, leading to a…

    FDA · DeviceBeckman Coulter, Inc.International distribution to the country of Andorra, Bulgaria, Denmark,…Official notice →
  46. 2025-07-02
  47. US
    Drugs & medicalClass II

    Beckman Coulter Inc. recalls REMISOL Advance UDI-DI code: 13700962601874 All Software Versions: B92487; B92488; C24317; C28652; C37500; D04164; C44703; C57017; C69412;…

    Due to a software issue, the software may incorrectly process the date of birth (DoB) received from the Laboratory Information System (LIS) and when the DoB is set to 01/01/1900 or left empty, REMISOL Advance may…

    FDA · DeviceBeckman Coulter Inc.Worldwide - U.S. Nationwide distribution including in the states of AK, AL,AR,…Official notice →
  48. US
    Drugs & medicalClass II

    Beckman Coulter, Inc. recalls DxI 9000 Access Immunoassay Analyzer C11137

    Analyzer has calibration issues where curves switched from passed to failed due to system errors during aspiration. This anomaly causes erroneous results and delays if unnoticed. RLU values may be misinterpreted as…

    FDA · DeviceBeckman Coulter, Inc.Worldwide - US Nationwide distribution in the states of AR, CA, FL, GA, HI,…Official notice →
  49. 2025-06-18
  50. US
    Drugs & medicalClass II

    Beckman Coulter, Inc. recalls DxI 9000 Access Immunoassay Analyzer, Catalog Number C11137, running software versions 1.20

    A new motor/encoder and cable was introduced into instruments manufactured starting with serial number 300141 incorrectly. This incorrect installation is introducing error codes for the Dxl 9000 which prevents the…

    FDA · DeviceBeckman Coulter, Inc.Worldwide distribution: US (Nationwide) and OUS Global (Foreign) to countries…Official notice →
  51. 2025-06-11
  52. US
    Drugs & medicalClass II

    Beckman Coulter, Inc. recalls DxI 9000 Access Immunoassay Analyzer, Catalog Number C11137, running software version 1.20

    A new software update (v1.20) is introducing error codes for the Dxl 9000 which prevents the assay from being performed and a delay in results from being generated, and may require field service to resolve the delay.

    FDA · DeviceBeckman Coulter, Inc.Worldwide Distribution: US (Nationwide) and OUS (Global) to countries of:…Official notice →
  53. US
    Drugs & medicalClass II

    Beckman Coulter, Inc. recalls Access Testosterone assay, Catalog Number 33560

    Inadequate preventative maintenance (PM) performed on the scales of the RPL5 fill line, led to splashing (cross-contamination) between the wells of testosterone reagent packs lot 538081, which could potentially result…

    FDA · DeviceBeckman Coulter, Inc.US State: FL. UAE, VietnamOfficial notice →
  54. 2025-05-21
  55. US
    Drugs & medicalClass II

    Beckman Coulter Inc. recalls DxC 500 AU Clinical Chemistry Analyzer, Catalog Numbers/UDI codes: C63519 / 14987666545058 C63520 / 14987666545065. The Beckman Coulter…

    Beckman Coulter is recalling their DxC 500 AU Clinical Chemistry Analyzer because a software error causes the analyzer to not run a requested calibration order in the following scenario: when a reagent blank or…

    FDA · DeviceBeckman Coulter Inc.U.S: AL, CA, GA, ID, IN, LA, MA, MI, MN, MT, NC, NV, PA, PR, TN , TX, and WI…Official notice →
  56. 2025-05-07
  57. US
    Drugs & medicalClass II

    Beckman Coulter Inc. recalls DxC 500 AU Clinical Chemistry Analyzer, Catalog Numbers/UDI codes: C63519 / 14987666545058 C63520 / 14987666545065 Software versions:…

    The reason for the recall is incorrect sample barcode identification and processing errors that result in the reporting of erroneous and delayed results.

    FDA · DeviceBeckman Coulter Inc.Worldwide - U.S. Nationwide distribution including in the states of AL, CA,…Official notice →
  58. 2025-04-30
  59. US
    Drugs & medicalClass II

    Beckman Coulter, Inc. recalls Beckman Coulter DxI 9000 Access Immunoassay Analyzer

    When DxI 9000 Access Immunoassay Analyzer is connected to a host system (laboratory information system (LIS) or middleware) and has accumulated canceled QC test results that were not sent by LIS, the analyzer may lose…

    FDA · DeviceBeckman Coulter, Inc.Worldwide distribution.Official notice →
  60. 2025-04-16
  61. US
    Drugs & medicalClass II

    Beckman Coulter, Inc. recalls DxI 9000 Access Immunoassay Analyzer, catalog number C11137 in vitro diagnostic device used for the quantitative, semi-quantitative, or…

    Ground screws on Dxl 9000 PCB boards may cause electric shorts, leading to smoke and potentially delaying patient results. Smoke also poses a hazard to equipment and users.

    FDA · DeviceBeckman Coulter, Inc.OUS (Foreign) distribution to countries of: Australia, Croatia, Czechia,…Official notice →
  62. 2025-04-09
  63. US
    Drugs & medicalClass II

    Beckman Coulter Inc. recalls DxC 500 AU Clinical Chemistry Analyzer, Catalog Numbers/UDI codes: C63519 / 14987666545058 C63520 / 14987666545065 The Beckman Coulter DxC…

    Beckman Coulter has determined that device software versions V1.3, V1.4, V1.4.1, V1.4.2, V1.4.3. have a defect due of which, when all constituent tests of a calculated result are rerun, either manually ordered by the…

    FDA · DeviceBeckman Coulter Inc.Worldwide Distribution: U.S. (nationwide) to states of: AL, CA, CO, GA, ID,…Official notice →
  64. 2025-03-26
  65. US
    Drugs & medicalClass II

    Beckman Coulter, Inc. recalls DxI 9000 Access Immunoassay Analyzer, Catalog Number C11137.

    the DxI 9000 encoder wheel was misaligned on the motor shaft, leading to increased slippage errors and causing the instrument to stop, posing a high safety risk due to potential delays in patient results.

    FDA · DeviceBeckman Coulter, Inc.US Nationwide. Global Distribution.Official notice →
  66. US
    Drugs & medicalClass II

    Beckman Coulter, Inc. recalls DxI 9000 Access Immunoassay Analyzer, Catalog Number C11137, with Component PCB, TSI/RSI firmware C29278

    Beckman Coulter identified that an issue for DxI 9000 tip stocker interface (TSI) and reagent storage interface (RSI) boards with the firmware version 1.1.0.0 can cause sub-system communication failures. This failure…

    FDA · DeviceBeckman Coulter, Inc.Worldwide - US Nationwide distribution in the states of FL, TX, WA.Official notice →
  67. US
    Drugs & medicalClass II

    Beckman Coulter, Inc. recalls Access 2 Immunoassay Analyzer, Catalog Number 81600N

    Beckman Coulter identified that an issue for Access 2 Sample Motors (B48016) due to improper alignment, causing the motors to rub against the sample wheelbase plate and index sensor. The issue was traced to a worn tool…

    FDA · DeviceBeckman Coulter, Inc.Worldwide - US Nationwide distribution.Official notice →
  68. 2025-03-19
  69. US
    Drugs & medicalClass II

    Beckman Coulter, Inc. recalls DxI 9000 Access Immunoassay Analyzer, Catalog Number C11137

    The DxI 9000 Access Immunoassay Analyzer wash wheel mixer may not spin freely within the idler pulley housing. When the problem occurs, the analyzer may generate a system event for a Wash Wheel Dispense Arm Spin Mixer…

    FDA · DeviceBeckman Coulter, Inc.US Nationwide. Global Distribution.Official notice →
  70. 2025-03-12
  71. US
    Drugs & medicalClass II

    Beckman Coulter, Inc. recalls Access Erythropoietin (EPO), Catalog Number - A16364 a glycoprotein (~30,400 Daltons) produced primarily by the kidney, is the principal…

    Affected lot (439363) exhibited a negative dose drop of -22% with native patient samples compared to alternate reagent lots. Patient samples tested with the affected lot may demonstrate repeatable falsely decreased…

    FDA · DeviceBeckman Coulter, Inc.Domestic distribution to AL AR AZ CA CT FL ID IL IN KS MA MI NC NE NJ NY OH OR…Official notice →
  72. 2025-02-12
  73. US
    Drugs & medicalClass II

    Beckman Coulter, Inc. recalls Beckman Coulter DxI 9000 Access Immunoassay Analyzer, catalog number C11137; in-vitro diagnostic instrument

    A software error caused an increase in tray gripper motion errors that causes intermittent hard stops and a delay in patient results.

    FDA · DeviceBeckman Coulter, Inc.Worldwide distribution - US Nationwide and the countries of Australia,…Official notice →
  74. US
    Drugs & medicalClass II

    Beckman Coulter Inc. recalls Access PCT Reagent Pack, Catalog Number C53987, when used on Access 2 and UniCel Dxl Immunoassay Systems

    Identified reagent lots are experiencing a high rate of calibration failures with failure code of "Max Iterate" on Access 2 and UniCel Dxl Immunoassay Systems. Reporting of results may be delayed due to the time…

    FDA · DeviceBeckman Coulter Inc.Worldwide distribution - US Nationwide and the countries of Austria, Bahrain,…Official notice →
  75. 2025-01-15
  76. US
    Drugs & medicalClass II

    Beckman Coulter, Inc. recalls Access Intact PTH assay, a paramagnetic particle, chemiluminescent immunoassay for the quantitative determination of intact parathyroid…

    Beckman Coulter has identified that some Access PTH reagent packs from lot 339071 may be over- or under-filled in well 0. The dispensed material in well 0 could be above or below specification, risking sealing issues,…

    FDA · DeviceBeckman Coulter, Inc.Worldwide distribution - US Nationwide and the country of Canada.Official notice →
  77. 2025-01-01
  78. US
    Drugs & medicalClass II

    Beckman Coulter, Inc. recalls Access hsTnI Reagent, REF B52699, Immunoassay Method, Troponin Subunit. For the quantitative determination of cardiac troponin I (cTnI)…

    Beckman Coulter, Inc. is recalling their Access hsTnI assay, a medical device by correction. The reason for the recall is: high troponin antigen (>55,000 pg/mL) can contaminate an analyzers probe and cause inter-assay…

    FDA · DeviceBeckman Coulter, Inc.WorldwideOfficial notice →
  79. 2024-12-18
  80. US
    Drugs & medicalClass II

    Beckman Coulter Inc. recalls Access Intact PTH assay on the DxI 9000 Access Immunoassay Analyzer, Catalog number A16972, for the quantitative determination of intact…

    The reason for the recall is that EDTA and lithium heparin plasma samples that are stored in the DxI 9000 Access Immunoassay Analyzer sample wheel for a prolonged period before performing the intact PTH (parathyroid…

    FDA · DeviceBeckman Coulter Inc.Foreign distribution to Albania, Australia, Austria, Bahrain, Belgium, Brazil,…Official notice →
  81. 2024-11-13
  82. US
    Drugs & medicalClass II

    Beckman Coulter, Inc. recalls The Access Cortisol assay is a paramagnetic particle, chemiluminescent immunoassay for the quantitative determination of cortisol levels…

    The Access Cortisol assay protocol file (APF) applies an incorrect calibration curve acceptance parameter at the S1 calibrator level that can lead to a high rate of calibration failures which can vary by reagent lot…

    FDA · DeviceBeckman Coulter, Inc.Worldwide distribution.Official notice →
  83. 2024-11-06
  84. US
    Drugs & medicalClass II

    Beckman Coulter, Inc. recalls Access Thyroglobulin Antibody II, REF A32898, containing Instructions for Use, IFU REF B14994, B63978, and A38168. The Access…

    IFUs for the device incorrectly states the biotin interference was remediated. The issue could result in potentially false-low results.

    FDA · DeviceBeckman Coulter, Inc.Distribution was made nationwide to the U.S., including Puerto Rico. There was…Official notice →
  85. 2024-10-23
  86. US
    Drugs & medicalClass II

    Beckman Coulter Inc. recalls IMMAGE Immunochemistry Systems IGM Immunoglobulin M Reagent REF 447610 IGM reagent, when used in conjunction with IMMAGE Immunochemistry…

    Firm identified two lots of reagents that are not meeting the sensitivity statement in the IFU

    FDA · DeviceBeckman Coulter Inc.U.S.: AZ, CA, CO, FL, GA, HI, IL, MA, MI, MO, ND, NY, OH, PA, and TX O.U.S.:…Official notice →
  87. US
    Drugs & medicalClass II

    Beckman Coulter, Inc. recalls The Access TSH (3rd IS) assay is a paramagnetic particle, chemiluminescent immunoassay for the quantitative etermdination of human…

    Certain lots of Access TSH reagent packs may produce erroneously low TSH (thyroid-stimulating hormone) results.

    FDA · DeviceBeckman Coulter, Inc.Worldwide distribution - US Nationwide and the countries of Austria, Belgium,…Official notice →
  88. US
    Drugs & medicalClass III

    Beckman Coulter, Inc. recalls Beckman Coulter IRISPEC CA/CB/CC Urine Chemistry Controls, Catalog Number REF 800-7211 and 800-7702, in vitro diagnostic device.

    Beckman Coulter has become aware of an increase in customer complaints for intermittent IRISpec CB negative glucose control failure, resulting in false positive control results when used in conjunction with…

    FDA · DeviceBeckman Coulter, Inc.worldwideOfficial notice →
  89. 2024-10-16
  90. US
    Drugs & medicalClass II

    Beckman Coulter, Inc. recalls Beckman Coulter Dxl 9000 Access Immunoassay Analyzer, REF #C11137. An in vitro diagnostic device used for the quantitative,…

    When the DxI 9000 Access Immunoassay Analyzer with system software version 1.16.2 and prior is configured to automatically request a calibration order, the instrument can process the order with expired calibrator…

    FDA · DeviceBeckman Coulter, Inc.Distribution was made to FL, GA, IN, KS, NH, NJ, OK, PA, and TX. There was…Official notice →
  91. 2024-09-25
  92. US
    Drugs & medicalClass II

    Beckman Coulter, Inc. recalls Beckman Coulter DxI 9000 Access Immunoassay Analyzer, Model Number C21510, RSM Pneumatic Tubing Kit item #2, REF C11137; in vitro…

    Beckman Coulter is recalling the DxI 9000 Access Immunoassay Analyzer due to undersized Pneumatics Tubing in the reagent storage module causing air leaks, compressor failures or sample processing errors. Four…

    FDA · DeviceBeckman Coulter, Inc.Worldwide distribution - US Nationwide and the countries of Australia,…Official notice →
  93. 2024-09-11
  94. US
    Drugs & medicalClass II

    Beckman Coulter, Inc. recalls DxI 9000 Access Immunoassay Analyzer, REF C11137, vitro diagnostic device used for the quantitative, semi-quantitative, or qualitative…

    Beckman Coulter manufacturing discovered cracks around a threaded insert on the bottom of the Internal Wash Buffer Tank that is used within the Wash Buffer Supply Drawer on the DxI 9000.This can result in leaking,…

    FDA · DeviceBeckman Coulter, Inc.Worldwide - US Nationwide distribution in the state of New Hampshire and the…Official notice →
  95. 2024-09-04
  96. US
    Drugs & medicalClass II

    Beckman Coulter, Inc. recalls DxI 9000 Access Immunoassay Analyzer, Catalog #C11137.

    When a DxI 9000 Access Immunoassay Analyzer is connected to a Laboratory Automation System (LAS) and is running with the system software versions 1.17.0 and below, the analyzer does not send a message to the LAS when…

    FDA · DeviceBeckman Coulter, Inc.Worldwide - US Nationwide distribution in the states of CA, FL, GA, IN, KS,…Official notice →
  97. 2024-08-21
  98. US
    Drugs & medicalClass II

    Beckman Coulter, Inc. recalls Discrete photometric chemistry analyzer for clinical use.

    Beckman Coulter, Inc. has identified a rise in reagent carousel motion errors in Access 2 Immunoassay System Analyzers that appear to be caused by misaligned index wheels housed within the reagent carousel module.…

    FDA · DeviceBeckman Coulter, Inc.Worldwide - US Nationwide distribution in the states of NE, MT, LA, TX, NY,…Official notice →
  99. US
    Drugs & medicalClass II

    Beckman Coulter, Inc. recalls Beckman Coulter DxI 9000 Access Immunoassay Analyzer, Catalog #C11137.

    The reason for this recall is Beckman Coulter has determined that when a DxI 9000 Access Immunoassay Analyzer is connected to a Laboratory Automation System (LAS); is running with the software versions 1.17 and below;…

    FDA · DeviceBeckman Coulter, Inc.Worldwide distribution - US Nationwide distribution in the states of CA, FL,…Official notice →
  100. 2024-07-31
  101. US
    Drugs & medicalClass II

    Beckman Coulter Inc. recalls Beckman Coulter, REF: OSR61173, IGM 4x14 mL R1, 4x11 mL R2

    LIH Influence check settings for Lipemia are incorrect which may result patient samples being reported as false low results or cause a high result to report as normal

    FDA · DeviceBeckman Coulter Inc.Worldwide - US Nationwide distribution including in the states of AL, AR, AZ,…Official notice →
  102. US
    Drugs & medicalClass II

    Beckman Coulter, Inc. recalls Dxl 9000 Access Immunoassay Analyzer, Catalog #C11137.

    The analyzers were built with nonconforming USB cables made with brass contacts instead of specified gold contacts. Brass contacts may corrode over time, leading to loss of internal instrument communication and delayed…

    FDA · DeviceBeckman Coulter, Inc.Worldwide - US Nationwide distribution in the states of FL, IN, KS, NH, PA,…Official notice →
  103. 2024-07-10
  104. US
    Drugs & medicalClass II

    Beckman Coulter, Inc. recalls DxI 9000 Access Immunoassay Analyzer, C11137; in-vitro diagnostic

    Beckman Coulter is recalling their DxI 9000 Access Immunoassay Analyzers, an in vitro diagnostic device used for the quantitative, semi-quantitative, or qualitative determination of various analyte concentrations found…

    FDA · DeviceBeckman Coulter, Inc.Worldwide distribution.Official notice →
  105. US
    Drugs & medicalClass II

    Beckman Coulter, Inc. recalls Beckman Coulter DxI 9000 Access Immunoassay Analyzer is a discrete photometric chemistry analyzer for clinical use. It is an in-vitro…

    The reason for the recall is the pneumatic tubing used in the analytical pick-and-place (PnP) assembly cable may fully or partially disconnect from the flex cable assembly fittings during normal operation. The DxI 9000…

    FDA · DeviceBeckman Coulter, Inc.Worldwide - US Nationwide distribution in the state of Indiana and the…Official notice →
  106. 2024-04-03
  107. US
    Drugs & medicalClass II

    Beckman Coulter, Inc. recalls Access Substrate, REF 81906, For use with the Access Immunoassay Systems

    On 07 February 2024 Beckman Coulter made the decision to conduct a field action on Access Substrate (81906) lots: 234600, 337984, 338040, 338214, 338229, 338289, 338322, 338333, 338451, 338587, and 338538. Beckman…

    FDA · DeviceBeckman Coulter, Inc.Worldwide Distribution: US (nationwide) and OUS (foreign) to countries of:…Official notice →
  108. 2024-03-27
  109. US
    Drugs & medicalClass II

    Beckman Coulter, Inc. recalls Access Total T4, REF 33800, IVD, thyroxine reagent assay.

    Four lots of Access Total T4 reagent lots demonstrated imprecision outside of the 10% specification which may result in erroneous or delayed results.

    FDA · DeviceBeckman Coulter, Inc.Worldwide distribution - US Nationwide, including Puerto Rico and the…Official notice →
  110. 2024-02-14
  111. US
    Drugs & medicalClass II

    Beckman Coulter, Inc. recalls DxI 9000 Access Immunoassay Analyzer running software version 1.16 or prior,

    There are potential performance issues found in the DxI 9000 Access Immunoassay Analyzer, including: 1) Access Ultrasensitive Insulin users who configure SI units (pmol/L) applies an incorrect conversion factor; 2)…

    FDA · DeviceBeckman Coulter, Inc.Worldwide distribution - US Nationwide and the countries of Australia,…Official notice →
  112. US
    Drugs & medicalClass II

    Beckman Coulter, Inc. recalls Access Ultrasensitive Insulin only used on the DxI 9000 Access Immunoassay Analyzer, Catalog Number 33410

    There is an issue with the Access Ultrasensitive Insulin assay protocol file (APF) for use with the DxI 9000 Access Immunoassay Analyzer (APF 182 test definition version 6.7 and below). The Access Ultrasensitive…

    FDA · DeviceBeckman Coulter, Inc.Worldwide distribution - US Nationwide and the countries of Australia,…Official notice →
  113. 2024-01-31
  114. US
    Drugs & medicalClass II

    Beckman Coulter, Inc. recalls Beckman Coulter DxU 850m/840m Microscopy Analyzers, Part Numbers: C49513, C76947

    Beckman Coulter has become aware that the cannula, which is found in the Flowcell and the iQClear Rinse Waste Well Assembly, may rust. Rust from the cannula has the potential to generate erroneous results.

    FDA · DeviceBeckman Coulter, Inc.US and Australia, Brazil, Canada, China, France, Germany, Ghana, India,…Official notice →
  115. 2023-10-18
  116. US
    Drugs & medicalClass II

    Beckman Coulter, Inc. recalls DxI 9000 Access Immunoassay Analyzer, REF C11137, Sofware Versions 1.15 and below; invitro diagnostic

    The firm has investigated and confirmed that when DxI 9000 is connected to Laboratory Information System (LIS, host), the sample barcode presented may be ignored and replaced with a sample ID that was previously in the…

    FDA · DeviceBeckman Coulter, Inc.Worldwide - US Nationwide distribution in the state of Florida, and the…Official notice →
  117. US
    Drugs & medicalClass II

    Beckman Coulter, Inc. recalls Access Thyroglobulin Calibrators

    It has been determined through an internal investigation that the alternate Access Thyroglobulin calibrator cards for use with the Thyg reagent packs distributed in April 2023 will cause a decrease of approximately 11%…

    FDA · DeviceBeckman Coulter, Inc.worldwide except US, Canada, and ChinaOfficial notice →
  118. 2023-09-06
  119. US
    Drugs & medicalClass II

    Beckman Coulter Inc. recalls Beckman Coulter, IVD, REF: OSR61171, IgA, 4 x 14 mL R1, 4 x 11 mL R2

    Lipemic interference failed to meet the performance specification listed within the IFU.

    FDA · DeviceBeckman Coulter Inc.US: AL AR AZ CA CO CT DC DE FL GA IA IL IN KS KY LA MA MD MI MN MO MS NC ND NE…Official notice →
  120. 2023-08-02
  121. US
    Drugs & medicalClass II

    Beckman Coulter, Inc. recalls Ringed DxI Reaction Vessels (RVs)

    Beckman Coulter, Inc. is recalling Ringed DxI Reaction Vessels (RVs) as part of Unicel DxI Immunoassay System (600, 800 series). The reason for the recall is that the firm has identified 3 lots of RVs that failed an…

    FDA · DeviceBeckman Coulter, Inc.Worldwide - US Nationwide distribution in the states of CA, MA, TX, NJ, MI,…Official notice →
  122. 2023-07-19
  123. US
    Drugs & medicalClass II

    Beckman Coulter, Inc. recalls COULTER DxH Diluent, REF 628017; diluent used for counting and sizing blood cells

    Specific diluent lot numbers showed that conductivity, osmolality, and pH were out of manufacturing specifications.

    FDA · DeviceBeckman Coulter, Inc.Worldwide distribution - US Nationwide and the countries of Argentina, Bolivia…Official notice →
  124. 2023-07-05
  125. US
    Drugs & medicalClass II

    Beckman Coulter, Inc. recalls Access Hybritech p2PSA reagent, REF B03704 and REF A49752.

    Reagent lot numbers include an insufficient concentration of blocking reagent.

    FDA · DeviceBeckman Coulter, Inc.Distribution was made to CA, CO, CT, FL, GA, IL, IN, MA, MI, NC, NE, NJ, NY,…Official notice →
  126. 2023-05-24
  127. US
    Drugs & medicalClass II

    Beckman Coulter Inc. recalls Beckman Coulter Bicarbonate, REF: OSR6137, 4x25 mL

    There is a potential for a decrease in calibration ODs for Bicarbonate reagent which could result in calibration failures and/or quality control failures.

    FDA · DeviceBeckman Coulter Inc.Worldwide Distribution: US (nationwide): AK, AL, AR, AZ, CA, CO, CT, DC, DE,…Official notice →
  128. US
    Drugs & medicalClass II

    Beckman Coulter, Inc. recalls DxU 840m Iris Urine Microscopy Analyzer, Catalog Number C76947, in-vitro diagnostic device, urine analysis

    Beckman Coulter has become aware of an intermittent issue in which the optional flag "Previous Sample Had Sperm" was enabled but not displayed so that a carryover event could have been investigated.

    FDA · DeviceBeckman Coulter, Inc.Worldwide DistributionOfficial notice →
  129. 2023-04-26
  130. US
    Drugs & medicalClass II

    Beckman Coulter, Inc. recalls Beckman Coulter Access Free T3, Triiodothyronine, Free, Catalog #A13422, each box contains 2 reagent packs.

    The reagent lot produces elevated results (approximately >30%) in some (15-20%)patient samples that do not correlate with the results of other thyroid markers nor Free T3 results generated on other platforms. This…

    FDA · DeviceBeckman Coulter, Inc.Worldwide distribution - US Nationwide, including Puerto Rico and the…Official notice →
  131. 2023-03-22
  132. US
    Drugs & medicalClass II

    Beckman Coulter, Inc. recalls Access Immunoassay Systems (1) Access 2 Immunoassay Analyzer (81600N); (2) Access 2 Immunoassay Analyzer, Refurbished (386220); (3) Access…

    An increased incidence of damage to the nut band from a new vendor (Boamax) has led to an increased volume of complaints from consignees regarding wash arm motion failures, system check failures, and/or assay…

    FDA · DeviceBeckman Coulter, Inc.Worldwide and Nationwide Distribution. There was government/military…Official notice →
  133. 2023-01-11
  134. US
    Drugs & medicalClass II

    Beckman Coulter, Inc. recalls BECKMAN COULTER IRISpec CA/CB/CC control, REF 900-7211, For in vitro diagnostic use only.

    Beckman Coulter has become aware of an increase in customer complaints generated for intermittent IRISpec CB glucose control false positive results when used in conjunction with iChemVelocity Urine Chemistry Strips PN…

    FDA · DeviceBeckman Coulter, Inc.US and Canada, Brazil, Korea, Mexico, Puerto Rico, TaiwanOfficial notice →
  135. 2022-11-30
  136. US
    Drugs & medicalClass II

    Beckman Coulter, Inc. recalls MicroScan autoSCAN-4 Instrument, Catalog Number B1018-280

    Due to MicroScan autoSCAN-4 Instruments being manufactured with an incorrect diffuser plate.

    FDA · DeviceBeckman Coulter, Inc.Worldwide - U.S. Nationwide distribution in the states of AK, FL, IA, ID, MI…Official notice →
  137. 2022-05-25
  138. US
    Drugs & medicalClass II

    Beckman Coulter Inc. recalls MicroScan WalkAway-40 plus Instrument REF B1018-283

    WalkAway Instruments may have been assembled with the incorrect AC power entry connector inlet with is rated at 1 ampere current rating rather than the correct current rating of 10 ampere

    FDA · DeviceBeckman Coulter Inc.Worldwide distribution - U.S. Nationwide distribution in the states of Alaska,…Official notice →
  139. 2022-04-27
  140. US
    Drugs & medicalClass III

    Beckman Coulter, Inc. recalls Beckman Coulter IRISpec CA/CB/CC control, Catalog #800-7211, containing 3 x 100 mL bottles each of CA Control, CB Control, and CC Control.

    IRISpec CA failures for glucose.

    FDA · DeviceBeckman Coulter, Inc.Worldwide distribution - US Nationwide, including Puerto Rico and the…Official notice →
  141. 2022-03-09
  142. US
    Drugs & medicalClass II

    Beckman Coulter Inc. recalls AU/DxC AU Magnesium, Catalog/REF Number OSR6189, UDI 15099590010850

    Lipemic interference for the Magnesium serum application failed to meet the performance claim as defined in the IFU. Use may cause a maximum positive bias up to 30.38% in low magnesium patient samples. The impact is…

    FDA · DeviceBeckman Coulter Inc.US: Alabama, Alaska, Arizona, Arkansas, California, Colorado, Connecticut,…Official notice →
  143. 2021-12-29
  144. US
    Drugs & medicalClass II

    Beckman Coulter Inc. recalls SYNCHRON Systems Cholesterol (CHOL) Reagent, REF: 467825

    Cholesterol Reagent may fail with Out of Calibration Range Low (OCR Low) and/or Back to Back (B/B) errors. Low quality control recovery failures may be observed after passing calibration. Brown particulates in…

    FDA · DeviceBeckman Coulter Inc.US Nationwide distribution in the states of Louisiana, Ohio, North Carolina,…Official notice →
  145. 2021-10-20
  146. US
    Drugs & medicalClass II

    Beckman Coulter Inc. recalls Power Processor 3K, 3K High Speed Stockyards Module Nameplate: PP 3K Stockyard (Refrig) 9707 REF A88116

    Missing splash guard which may expose users to hazardous materials. .

    FDA · DeviceBeckman Coulter Inc.U.S. Nationwide Distribution in the states of: CA, GA, IA, ID, IL, IN, LA, MA,…Official notice →
  147. 2021-10-13
  148. US
    Drugs & medicalClass II

    Beckman Coulter Inc. recalls Access hsTnI High Sensitivity Troponin I, REF B52699 Access Immunoassay System REAGENT PACK

    Through customer feedback, testing a sample with cardiac troponin I concentration that are >270,000 pg/mL may cause carryover into the hsTnl reagent pack. The carryover is caused by sample-to-pack carryover…

    FDA · DeviceBeckman Coulter Inc.U.S.: AL, AR, AZ, CA, CO, CT, DE, DC, FL, GA, Guam, HI, IA, ID, IL, IN, KS,…Official notice →
  149. 2021-10-06
  150. US
    Drugs & medicalClass II

    Beckman Coulter Inc. recalls DxA 5000 (DxA Automation System, DxA 5000 fit), "For in Vitro Diagnostic Use", Reference Numbers: B87352 with B57634, UDI: 15099590707118

    There is a potential that sample carriers (w or w/o samples) within automated sampling system may transport samples to the outer lane which does not contain a physical barrier to restrain samples. Samples are dropped…

    FDA · DeviceBeckman Coulter Inc.US Nationwide distribution in the states of AL, CA, IL, In, KY, MN, MI, OH,…Official notice →