Drugs & medicalClass II

Beckman Coulter, Inc. recalls DxI 9000 Access Immunoassay Analyzer, Catalog Number C11137

The DxI 9000 Access Immunoassay Analyzer wash wheel mixer may not spin freely within the idler pulley housing. When the problem occurs, the analyzer may generate a system event for a Wash Wheel Dispense Arm Spin Mixer…

Read the official notice →
Date
2025-03-19
Country
United States
Brand
Beckman Coulter
Product
DxI 9000 Access Immunoassay Analyzer, Catalog Number C11137
Classification
Class II
Units
US Nationwide. Global Distribution.
Category
Drugs & medical
Source
FDA · Device

Copied automatically from the US Food and Drug Administration. Details may be shortened or out of date; the official notice is the authoritative source. Disclaimer

What to do

  1. Check the product name, strength and lot number against the official notice.
  2. Do not stop a prescribed medicine without talking to your doctor or pharmacist first.
  3. Ask the pharmacy or supplier about a replacement or refund.

General guidance from Recall Atlas. The official notice from the US Food and Drug Administration has the exact models, batch codes and remedy, and it takes priority.

More Beckman Coulter recalls

  1. 2025-03-26
  2. US
    Drugs & medicalClass II

    Beckman Coulter, Inc. recalls DxI 9000 Access Immunoassay Analyzer, Catalog Number C11137.

    the DxI 9000 encoder wheel was misaligned on the motor shaft, leading to increased slippage errors and causing the instrument to stop, posing a high safety risk due to potential delays in patient results.

    FDA · DeviceBeckman Coulter, Inc.US Nationwide. Global Distribution.Official notice →
  3. US
    Drugs & medicalClass II

    Beckman Coulter, Inc. recalls DxI 9000 Access Immunoassay Analyzer, Catalog Number C11137, with Component PCB, TSI/RSI firmware C29278

    Beckman Coulter identified that an issue for DxI 9000 tip stocker interface (TSI) and reagent storage interface (RSI) boards with the firmware version 1.1.0.0 can cause sub-system communication failures. This failure…

    FDA · DeviceBeckman Coulter, Inc.Worldwide - US Nationwide distribution in the states of FL, TX, WA.Official notice →
  4. US
    Drugs & medicalClass II

    Beckman Coulter, Inc. recalls Access 2 Immunoassay Analyzer, Catalog Number 81600N

    Beckman Coulter identified that an issue for Access 2 Sample Motors (B48016) due to improper alignment, causing the motors to rub against the sample wheelbase plate and index sensor. The issue was traced to a worn tool…

    FDA · DeviceBeckman Coulter, Inc.Worldwide - US Nationwide distribution.Official notice →
  5. 2025-03-12
  6. US
    Drugs & medicalClass II

    Beckman Coulter, Inc. recalls Access Erythropoietin (EPO), Catalog Number - A16364 a glycoprotein (~30,400 Daltons) produced primarily by the kidney, is the principal…

    Affected lot (439363) exhibited a negative dose drop of -22% with native patient samples compared to alternate reagent lots. Patient samples tested with the affected lot may demonstrate repeatable falsely decreased…

    FDA · DeviceBeckman Coulter, Inc.Domestic distribution to AL AR AZ CA CT FL ID IL IN KS MA MI NC NE NJ NY OH OR…Official notice →
  7. 2025-02-12
  8. US
    Drugs & medicalClass II

    Beckman Coulter, Inc. recalls Beckman Coulter DxI 9000 Access Immunoassay Analyzer, catalog number C11137; in-vitro diagnostic instrument

    A software error caused an increase in tray gripper motion errors that causes intermittent hard stops and a delay in patient results.

    FDA · DeviceBeckman Coulter, Inc.Worldwide distribution - US Nationwide and the countries of Australia,…Official notice →
  9. US
    Drugs & medicalClass II

    Beckman Coulter Inc. recalls Access PCT Reagent Pack, Catalog Number C53987, when used on Access 2 and UniCel Dxl Immunoassay Systems

    Identified reagent lots are experiencing a high rate of calibration failures with failure code of "Max Iterate" on Access 2 and UniCel Dxl Immunoassay Systems. Reporting of results may be delayed due to the time…

    FDA · DeviceBeckman Coulter Inc.Worldwide distribution - US Nationwide and the countries of Austria, Bahrain,…Official notice →
All 84 Beckman Coulter recalls →

Other drugs & medical recalls in the United States

  1. 2025-03-19
  2. US
    Drugs & medicalClass II

    Zimmer, Inc. recalls Z1 Femoral Hip System, Product Number 611777612

    There is a potential design issue with the Z1 Offset Stem Inserter leading to fracturing of the distal tip.

    FDA · DeviceZimmer, Inc.US Nationwide distribution.Official notice →
  3. US
    Drugs & medicalClass II

    Stryker Corporation recalls Stryker CinchLock Flex Knotless Anchor with Inserter and Suture, REF CAT02643, Sterile EO, Rx Only MR Safe CE 0197

    Increase in complaints concerning "pullwire" breakage that may result in the broken pullwire remaining in the anchor implant body after deployment.

    FDA · DeviceStryker CorporationWorldwide - US Nationwide distribution including in the states of Alabama,…Official notice →
  4. US
    Drugs & medicalClass II

    Spectrum Medical Inc recalls Brand Name: Quantum Perfusion Centrifugal Blood Pump CP22 with Integrated Sensors Product Name: Pump, Blood, Cardiopulmonary Bypass,…

    Their is a potential of fluid leakage from the centrifugal blood pump pump during the priming procedure or during clinical procedures.

    FDA · DeviceSpectrum Medical IncUS Nationwide distribution including in the states of Alabama, Arizona,…Official notice →
  5. US
    Drugs & medicalClass II

    Intuitive Surgical, Inc. recalls Brand Name: Da Vinci 5 Product Name: ASSY,DV5 CONSOLE,IS5000 Foot Tray Pedals Model/Catalog Number: 380730 Software Version: N/A . Refer…

    Due to an increase in complaints concerning foot tray pedal spring failing resulting in the pedal remaining pressed

    FDA · DeviceIntuitive Surgical, Inc.Worldwide distribution - US Nationwide and the countries of Alaska, Arizona,…Official notice →
  6. US
    Drugs & medicalClass II

    Philips Medical Systems Nederland B.V. recalls Spectral CT on Rails, Software Version Number 5.1.0, Model number: 728334;

    Multiple problems identified with the software version leading to various scanning and image issues, and unintended device movement.

    FDA · DevicePhilips Medical Systems Nederland B.V.Domestic: MNOfficial notice →
  7. US
    Drugs & medicalClass II

    ZOLL Medical Corporation recalls Brand Name: ZOLL Product Name: Powerheart G5 AED Model/Catalog Number: G5S-80C; Software Version: NA Product Description: The Powerheart…

    The potential for devices to fail their self-test as a result of prolonged exposure to extreme environments (temperature and/or humidity).

    FDA · DeviceZOLL Medical CorporationDomestic: Nationwide Distribution International: AE, AI, AM, AR, AT, AU, BD,…Official notice →
See all drugs & medical recalls →