Drugs & medicalClass II

Beckman Coulter, Inc. recalls Access 2 Immunoassay Analyzer, Catalog Number 81600N

Beckman Coulter identified that an issue for Access 2 Sample Motors (B48016) due to improper alignment, causing the motors to rub against the sample wheelbase plate and index sensor. The issue was traced to a worn tool…

Read the official notice →
Date
2025-03-26
Country
United States
Brand
Beckman Coulter
Product
Access 2 Immunoassay Analyzer, Catalog Number 81600N
Classification
Class II
Units
Worldwide - US Nationwide distribution.
Category
Drugs & medical
Source
FDA · Device

Copied automatically from the US Food and Drug Administration. Details may be shortened or out of date; the official notice is the authoritative source. Disclaimer

What to do

  1. Check the product name, strength and lot number against the official notice.
  2. Do not stop a prescribed medicine without talking to your doctor or pharmacist first.
  3. Ask the pharmacy or supplier about a replacement or refund.

General guidance from Recall Atlas. The official notice from the US Food and Drug Administration has the exact models, batch codes and remedy, and it takes priority.

More Beckman Coulter recalls

  1. 2025-04-09
  2. US
    Drugs & medicalClass II

    Beckman Coulter Inc. recalls DxC 500 AU Clinical Chemistry Analyzer, Catalog Numbers/UDI codes: C63519 / 14987666545058 C63520 / 14987666545065 The Beckman Coulter DxC…

    Beckman Coulter has determined that device software versions V1.3, V1.4, V1.4.1, V1.4.2, V1.4.3. have a defect due of which, when all constituent tests of a calculated result are rerun, either manually ordered by the…

    FDA · DeviceBeckman Coulter Inc.Worldwide Distribution: U.S. (nationwide) to states of: AL, CA, CO, GA, ID,…Official notice →
  3. 2025-03-26
  4. US
    Drugs & medicalClass II

    Beckman Coulter, Inc. recalls DxI 9000 Access Immunoassay Analyzer, Catalog Number C11137.

    the DxI 9000 encoder wheel was misaligned on the motor shaft, leading to increased slippage errors and causing the instrument to stop, posing a high safety risk due to potential delays in patient results.

    FDA · DeviceBeckman Coulter, Inc.US Nationwide. Global Distribution.Official notice →
  5. US
    Drugs & medicalClass II

    Beckman Coulter, Inc. recalls DxI 9000 Access Immunoassay Analyzer, Catalog Number C11137, with Component PCB, TSI/RSI firmware C29278

    Beckman Coulter identified that an issue for DxI 9000 tip stocker interface (TSI) and reagent storage interface (RSI) boards with the firmware version 1.1.0.0 can cause sub-system communication failures. This failure…

    FDA · DeviceBeckman Coulter, Inc.Worldwide - US Nationwide distribution in the states of FL, TX, WA.Official notice →
  6. 2025-03-19
  7. US
    Drugs & medicalClass II

    Beckman Coulter, Inc. recalls DxI 9000 Access Immunoassay Analyzer, Catalog Number C11137

    The DxI 9000 Access Immunoassay Analyzer wash wheel mixer may not spin freely within the idler pulley housing. When the problem occurs, the analyzer may generate a system event for a Wash Wheel Dispense Arm Spin Mixer…

    FDA · DeviceBeckman Coulter, Inc.US Nationwide. Global Distribution.Official notice →
  8. 2025-03-12
  9. US
    Drugs & medicalClass II

    Beckman Coulter, Inc. recalls Access Erythropoietin (EPO), Catalog Number - A16364 a glycoprotein (~30,400 Daltons) produced primarily by the kidney, is the principal…

    Affected lot (439363) exhibited a negative dose drop of -22% with native patient samples compared to alternate reagent lots. Patient samples tested with the affected lot may demonstrate repeatable falsely decreased…

    FDA · DeviceBeckman Coulter, Inc.Domestic distribution to AL AR AZ CA CT FL ID IL IN KS MA MI NC NE NJ NY OH OR…Official notice →
  10. 2025-02-12
  11. US
    Drugs & medicalClass II

    Beckman Coulter, Inc. recalls Beckman Coulter DxI 9000 Access Immunoassay Analyzer, catalog number C11137; in-vitro diagnostic instrument

    A software error caused an increase in tray gripper motion errors that causes intermittent hard stops and a delay in patient results.

    FDA · DeviceBeckman Coulter, Inc.Worldwide distribution - US Nationwide and the countries of Australia,…Official notice →
All 84 Beckman Coulter recalls →

Other drugs & medical recalls in the United States

  1. 2025-03-26
  2. US
    Drugs & medicalClass II

    Immuno-Mycologics, Inc recalls Candida ID Antigen, REAG, AG, REF C50110, Vol 1 mL REF ID1001, Immy Fungal Antigens, Positive Controls, & Immunodiffusion (ID) plates for…

    Candida Immunodiffusion Antigen may have reduced sensitivity which could result in false negative results.

    FDA · DeviceImmuno-Mycologics, IncWorldwide - US Nationwide distribution in the states of CA and OK and the…Official notice →
  3. US
    Drugs & medicalClass II

    Beckman Coulter, Inc. recalls DxI 9000 Access Immunoassay Analyzer, Catalog Number C11137.

    the DxI 9000 encoder wheel was misaligned on the motor shaft, leading to increased slippage errors and causing the instrument to stop, posing a high safety risk due to potential delays in patient results.

    FDA · DeviceBeckman Coulter, Inc.US Nationwide. Global Distribution.Official notice →
  4. US
    Drugs & medicalClass II

    Beckman Coulter, Inc. recalls DxI 9000 Access Immunoassay Analyzer, Catalog Number C11137, with Component PCB, TSI/RSI firmware C29278

    Beckman Coulter identified that an issue for DxI 9000 tip stocker interface (TSI) and reagent storage interface (RSI) boards with the firmware version 1.1.0.0 can cause sub-system communication failures. This failure…

    FDA · DeviceBeckman Coulter, Inc.Worldwide - US Nationwide distribution in the states of FL, TX, WA.Official notice →
  5. US
    Drugs & medicalClass II

    Oridion Medical 1987 Ltd. recalls Microstream Instructions for Use and Part Number used with - Product Description, REF: Filter Line Sets Adult/Pediatric &…

    Neonatal-Infant and Adult-Pediatric CO2 Sampling Line/Airway Adapter/CO2 Oral-Nasal Filter Line/Intubated Filter (CO2) Line device users may experience difficulty/inability to disconnect adapter from endotracheal tube…

    FDA · DeviceOridion Medical 1987 Ltd.Worldwide - US Nationwide distribution including in the states of TX, OH, FL,…Official notice →
  6. US
    Drugs & medicalClass II

    CareFusion 303, Inc. recalls BD Pyxis: Anesthesia Station ES, REF: 327, MedStation ES, REF: 323, MedStation ES Tower, REF: 352, CII Safe ES, REF: 1116-00, CII Safe ES…

    Sync timeout may occur between the server and automated dispensing device, due to a server software error which may occur if syncing a large amount of data that takes excessive time, which may cause the dispensing…

    FDA · DeviceCareFusion 303, Inc.Worldwide - US Nationwide distribution including in the states of KY, OH, NY,…Official notice →
  7. US
    Drugs & medicalClass II

    Waldemar Link GmbH & Co. KG (Mfg Site) recalls Modular Stem cylindrical, Tilastan¿, press-fit cementless, L= 80 mm, DIA = 11 mm, Product Code: 880-601/11

    A problem was discovered during the scanning process of the outer carton label because of an incorrect GTIN on the label, which is a GTIN for a different product.

    FDA · DeviceWaldemar Link GmbH & Co. KG (Mfg Site)Worldwide - US Nationwide distribution in the states of AL, CA, CO, FL, GA,…Official notice →
See all drugs & medical recalls →