Waldemar Link GmbH & Co. KG (Mfg Site) recalls Modular Stem cylindrical, Tilastan¿, press-fit cementless, L= 80 mm, DIA = 11 mm, Product Code: 880-601/11
A problem was discovered during the scanning process of the outer carton label because of an incorrect GTIN on the label, which is a GTIN for a different product.
Read the official notice →- Date
- 2025-03-26
- Country
- United States
- Brand
- Waldemar Link GmbH & Co. KG
- Product
- Modular Stem cylindrical, Tilastan¿, press-fit cementless, L= 80 mm, DIA = 11 mm, Product Code: 880-601/11
- Classification
- Class II
- Units
- Worldwide - US Nationwide distribution in the states of AL, CA, CO, FL, GA,…
- Category
- Drugs & medical
- Source
- FDA · Device
Copied automatically from the US Food and Drug Administration. Details may be shortened or out of date; the official notice is the authoritative source. Disclaimer
What to do
- Check the product name, strength and lot number against the official notice.
- Do not stop a prescribed medicine without talking to your doctor or pharmacist first.
- Ask the pharmacy or supplier about a replacement or refund.
General guidance from Recall Atlas. The official notice from the US Food and Drug Administration has the exact models, batch codes and remedy, and it takes priority.
More Waldemar Link GmbH & Co. KG recalls
- 2025-09-17
US
Waldemar Link GmbH & Co. KG (Mfg Site) recalls LinkSymphoKnee System, Tibial Component, Modular, Symmetric, Sz. 5, CoCrMo/TiNbN, cemented, Fixed. REF Number: 880-100/50
Some modular LinkSymphoKnee TiNbN Tibial Components were manufactured with two uncoated security screws, which is incorrect.
FDA · DeviceWaldemar Link GmbH & Co. KG (Mfg Site)Worldwide - US Nationwide distribution in the states of CO, IN, NJ, TX and the…Official notice →- 2025-08-20
US
Waldemar Link GmbH & Co. KG (Mfg Site) recalls Prosthesis Head B, 28 mm, 7 mm neck length. Item Number: 198-828/04.
Inconsistent size terminology and color coding used on labeling
FDA · DeviceWaldemar Link GmbH & Co. KG (Mfg Site)Worldwide - US Nationwide distribution in the states of AZ, CA, FL, IN, KS,…Official notice →- 2025-05-07
US
Waldemar Link GmbH & Co. KG (Mfg Site) recalls Posterior Femoral Augment; Item Numbers: (1) 880-310/11 (Sz. 0 - cemented sm), (2) 880-310/21 (Sz. 0 - cemented lg), (3)…
The device was delivered with a preassembled Femoral Augment screw that was missing its thread.
FDA · DeviceWaldemar Link GmbH & Co. KG (Mfg Site)US Nationwide distribution in the states of AL, AZ, CA, CO, FL, GA, IN, IL,…Official notice →- 2025-01-01
US
Waldemar Link GmbH & Co. KG (Mfg Site) recalls LINK MP Reconstruction System Neck Segment (with suture holes, standard, uncemented, L= 35 mm, CCD 126¿, Offset= 31 mm,…
Error on the main label and the patient label. The labels indicate a size XXL, this is a standard-size neck segment.. All further information on the labels is correct (Length, CCD and Offset).
FDA · DeviceWaldemar Link GmbH & Co. KG (Mfg Site)US Nationwide distribution in the states of AL, FL.Official notice →- 2024-10-02
US
Waldemar Link GmbH & Co. KG (Corp. Hq.) recalls LINK Universal Handle, with quick coupling, Stainless Steel, straight, Item Number 130-394/01
Some lots of the LINK Universal Handle, with quick coupling, Stainless Steel, straight may suffer an immediate material failure intraoperatively when the instrument is used for the first time. The instrument can break…
FDA · DeviceWaldemar Link GmbH & Co. KG (Corp. Hq.)US Nationwide distribution in the states of FL, KS, MI, NJ, VA.Official notice →- 2024-07-24
US
Waldemar Link GmbH & Co. KG (Mfg Site) recalls Endo Model SL Tibial Components 16-2817/02 Tibial Component, Small, W: 60mm 16-2817/05 Tibial Component, Medium, W: 65mm…
Blind screws of the modular tibial component cannot be loosened intraoperatively,prolongation surgery due to intraoperatively change in procedure, probably to cementing technique
FDA · DeviceWaldemar Link GmbH & Co. KG (Mfg Site)Nationwide Foreign: Algeria Argentina Australia Austria Belarus Belgium Brazil…Official notice →
Other drugs & medical recalls in the United States
- 2025-03-26
US
Beckman Coulter, Inc. recalls Access 2 Immunoassay Analyzer, Catalog Number 81600N
Beckman Coulter identified that an issue for Access 2 Sample Motors (B48016) due to improper alignment, causing the motors to rub against the sample wheelbase plate and index sensor. The issue was traced to a worn tool…
US
Oridion Medical 1987 Ltd. recalls Microstream Instructions for Use and Part Number used with - Product Description, REF: Filter Line Sets Adult/Pediatric &…
Neonatal-Infant and Adult-Pediatric CO2 Sampling Line/Airway Adapter/CO2 Oral-Nasal Filter Line/Intubated Filter (CO2) Line device users may experience difficulty/inability to disconnect adapter from endotracheal tube…
FDA · DeviceOridion Medical 1987 Ltd.Worldwide - US Nationwide distribution including in the states of TX, OH, FL,…Official notice →US
CareFusion 303, Inc. recalls BD Pyxis: Anesthesia Station ES, REF: 327, MedStation ES, REF: 323, MedStation ES Tower, REF: 352, CII Safe ES, REF: 1116-00, CII Safe ES…
Sync timeout may occur between the server and automated dispensing device, due to a server software error which may occur if syncing a large amount of data that takes excessive time, which may cause the dispensing…
FDA · DeviceCareFusion 303, Inc.Worldwide - US Nationwide distribution including in the states of KY, OH, NY,…Official notice →US
Microbiologics Inc recalls Helix Elite" Inactivated Standard: Inactivated SARS-CoV-2 Whole Virus (Pellet), Multi-analyte, unassayed quality control material, Catalog…
The A549 human cell target (human cells for sample adequacy control) may give a late Ct value.
US
MEDLINE INDUSTRIES, LP - Northfield recalls Proxima Drape labeled as PACK,BASIC,SIRUS,V,5/CS; PACK,BASIC IV,SIRUS,5/CS; PACK,SET UP,NO DRAPE,12/CS; PACK,SET UP,NO…
Potential breach in pouch packaging which could lead to loss of sterility.
US
American Contract Systems Inc recalls MAJOR VASCULAR PACK , Model No UTMV78X-01 UTMV78Y UTMV78Y-01 UTMV78X-01 UTMV78Y-02
ACS is unable to confirm product sterilization assurance requirements were met. Possible loss or lack of functionality may lead to delay or prolonged treatment.
FDA · DeviceAmerican Contract Systems IncUS Nationwide distribution in the state of OH.Official notice →