Drugs & medicalClass II

Beckman Coulter, Inc. recalls Beckman Coulter DxI 9000 Access Immunoassay Analyzer, catalog number C11137; in-vitro diagnostic instrument

A software error caused an increase in tray gripper motion errors that causes intermittent hard stops and a delay in patient results.

Read the official notice →
Date
2025-02-12
Country
United States
Brand
Beckman Coulter
Product
Beckman Coulter DxI 9000 Access Immunoassay Analyzer, catalog number C11137; in-vitro diagnostic instrument
Classification
Class II
Units
Worldwide distribution - US Nationwide and the countries of Australia,…
Category
Drugs & medical
Source
FDA · Device

Copied automatically from the US Food and Drug Administration. Details may be shortened or out of date; the official notice is the authoritative source. Disclaimer

What to do

  1. Check the product name, strength and lot number against the official notice.
  2. Do not stop a prescribed medicine without talking to your doctor or pharmacist first.
  3. Ask the pharmacy or supplier about a replacement or refund.

General guidance from Recall Atlas. The official notice from the US Food and Drug Administration has the exact models, batch codes and remedy, and it takes priority.

More Beckman Coulter recalls

  1. 2025-03-26
  2. US
    Drugs & medicalClass II

    Beckman Coulter, Inc. recalls Access 2 Immunoassay Analyzer, Catalog Number 81600N

    Beckman Coulter identified that an issue for Access 2 Sample Motors (B48016) due to improper alignment, causing the motors to rub against the sample wheelbase plate and index sensor. The issue was traced to a worn tool…

    FDA · DeviceBeckman Coulter, Inc.Worldwide - US Nationwide distribution.Official notice →
  3. 2025-03-19
  4. US
    Drugs & medicalClass II

    Beckman Coulter, Inc. recalls DxI 9000 Access Immunoassay Analyzer, Catalog Number C11137

    The DxI 9000 Access Immunoassay Analyzer wash wheel mixer may not spin freely within the idler pulley housing. When the problem occurs, the analyzer may generate a system event for a Wash Wheel Dispense Arm Spin Mixer…

    FDA · DeviceBeckman Coulter, Inc.US Nationwide. Global Distribution.Official notice →
  5. 2025-03-12
  6. US
    Drugs & medicalClass II

    Beckman Coulter, Inc. recalls Access Erythropoietin (EPO), Catalog Number - A16364 a glycoprotein (~30,400 Daltons) produced primarily by the kidney, is the principal…

    Affected lot (439363) exhibited a negative dose drop of -22% with native patient samples compared to alternate reagent lots. Patient samples tested with the affected lot may demonstrate repeatable falsely decreased…

    FDA · DeviceBeckman Coulter, Inc.Domestic distribution to AL AR AZ CA CT FL ID IL IN KS MA MI NC NE NJ NY OH OR…Official notice →
  7. 2025-02-12
  8. US
    Drugs & medicalClass II

    Beckman Coulter Inc. recalls Access PCT Reagent Pack, Catalog Number C53987, when used on Access 2 and UniCel Dxl Immunoassay Systems

    Identified reagent lots are experiencing a high rate of calibration failures with failure code of "Max Iterate" on Access 2 and UniCel Dxl Immunoassay Systems. Reporting of results may be delayed due to the time…

    FDA · DeviceBeckman Coulter Inc.Worldwide distribution - US Nationwide and the countries of Austria, Bahrain,…Official notice →
  9. 2025-01-15
  10. US
    Drugs & medicalClass II

    Beckman Coulter, Inc. recalls Access Intact PTH assay, a paramagnetic particle, chemiluminescent immunoassay for the quantitative determination of intact parathyroid…

    Beckman Coulter has identified that some Access PTH reagent packs from lot 339071 may be over- or under-filled in well 0. The dispensed material in well 0 could be above or below specification, risking sealing issues,…

    FDA · DeviceBeckman Coulter, Inc.Worldwide distribution - US Nationwide and the country of Canada.Official notice →
  11. 2025-01-01
  12. US
    Drugs & medicalClass II

    Beckman Coulter, Inc. recalls Access hsTnI Reagent, REF B52699, Immunoassay Method, Troponin Subunit. For the quantitative determination of cardiac troponin I (cTnI)…

    Beckman Coulter, Inc. is recalling their Access hsTnI assay, a medical device by correction. The reason for the recall is: high troponin antigen (>55,000 pg/mL) can contaminate an analyzers probe and cause inter-assay…

    FDA · DeviceBeckman Coulter, Inc.WorldwideOfficial notice →
All 84 Beckman Coulter recalls →

Other drugs & medical recalls in the United States

  1. 2025-02-19
  2. US
    Drugs & medicalClass I

    Olympus Corporation of the Americas recalls Olympus Single Use Guide Sheath Kits with the following product descriptions and model numbers: 1. Single Use Guide Sheath…

    Potential for the radiopaque tip of the guide sheath component to disassociate and fall off into the patient.

    FDA · DeviceOlympus Corporation of the AmericasUS nationwide distribution.Official notice →
  3. US
  4. 2025-02-12
  5. US
    Drugs & medicalClass II

    Medicrea International recalls UNiD" Adaptive Spine Intelligence UNiD Spine Analyzer Product Number SW3002 UDI-DI code 03613720286929 Two cloud-based software…

    Due to software anomalies that may impact on rod planning specifically, certain optional surgical parameters may have had errors that resulted i incorrect calculations displayed on system. Impacted parameters includes…

    FDA · DeviceMedicrea InternationalU.S.: AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY,…Official notice →
  6. US
    Drugs & medicalClass II

    US Endodontics, LLC recalls Brasseler ESR Rotary/Reciprocating Endodontic File Device SKU AP120213PK AP120253PK AP120313PK AP125213PK AP125253PK AP125313PK AP135213PK…

    Due to devices being irradiated above the established specification which my result in package integrity issues.

    FDA · DeviceUS Endodontics, LLCWorldwide - US Nationwide distribution in the state of GA and the countries of…Official notice →
  7. US
    Drugs & medicalClass II

    Sysmex America, Inc. recalls Sysmex PS-10 Sample Preparation System with software version 1.5 and 1.6., Catalog Number BQ716341. Product usage for automated pipetting,…

    Sysmex was made aware of a report of false results due to carry-over caused by the sporadic omission of the probe wash step during antibody pipetting on the PS-10 Sample Preparation System.

    FDA · DeviceSysmex America, Inc.US Nationwide.Official notice →
  8. US
    Drugs & medicalClass II

    Beckman Coulter Inc. recalls Access PCT Reagent Pack, Catalog Number C53987, when used on Access 2 and UniCel Dxl Immunoassay Systems

    Identified reagent lots are experiencing a high rate of calibration failures with failure code of "Max Iterate" on Access 2 and UniCel Dxl Immunoassay Systems. Reporting of results may be delayed due to the time…

    FDA · DeviceBeckman Coulter Inc.Worldwide distribution - US Nationwide and the countries of Austria, Bahrain,…Official notice →