Beckman Coulter, Inc. recalls DxI 9000 Access Immunoassay Analyzer, catalog number C11137 in vitro diagnostic device used for the quantitative, semi-quantitative, or…
Ground screws on Dxl 9000 PCB boards may cause electric shorts, leading to smoke and potentially delaying patient results. Smoke also poses a hazard to equipment and users.
Read the official notice →- Date
- 2025-04-16
- Country
- United States
- Brand
- Beckman Coulter
- Product
- DxI 9000 Access Immunoassay Analyzer, catalog number C11137 in vitro diagnostic device used for the quantitative, semi-quantitative, or…
- Classification
- Class II
- Units
- OUS (Foreign) distribution to countries of: Australia, Croatia, Czechia,…
- Category
- Drugs & medical
- Source
- FDA · Device
Copied automatically from the US Food and Drug Administration. Details may be shortened or out of date; the official notice is the authoritative source. Disclaimer
What to do
- Check the product name, strength and lot number against the official notice.
- Do not stop a prescribed medicine without talking to your doctor or pharmacist first.
- Ask the pharmacy or supplier about a replacement or refund.
General guidance from Recall Atlas. The official notice from the US Food and Drug Administration has the exact models, batch codes and remedy, and it takes priority.
More Beckman Coulter recalls
- 2025-05-21
US
Beckman Coulter Inc. recalls DxC 500 AU Clinical Chemistry Analyzer, Catalog Numbers/UDI codes: C63519 / 14987666545058 C63520 / 14987666545065. The Beckman Coulter…
Beckman Coulter is recalling their DxC 500 AU Clinical Chemistry Analyzer because a software error causes the analyzer to not run a requested calibration order in the following scenario: when a reagent blank or…
FDA · DeviceBeckman Coulter Inc.U.S: AL, CA, GA, ID, IN, LA, MA, MI, MN, MT, NC, NV, PA, PR, TN , TX, and WI…Official notice →- 2025-05-07
US
Beckman Coulter Inc. recalls DxC 500 AU Clinical Chemistry Analyzer, Catalog Numbers/UDI codes: C63519 / 14987666545058 C63520 / 14987666545065 Software versions:…
The reason for the recall is incorrect sample barcode identification and processing errors that result in the reporting of erroneous and delayed results.
FDA · DeviceBeckman Coulter Inc.Worldwide - U.S. Nationwide distribution including in the states of AL, CA,…Official notice →- 2025-04-30
US
Beckman Coulter, Inc. recalls Beckman Coulter DxI 9000 Access Immunoassay Analyzer
When DxI 9000 Access Immunoassay Analyzer is connected to a host system (laboratory information system (LIS) or middleware) and has accumulated canceled QC test results that were not sent by LIS, the analyzer may lose…
- 2025-04-09
US
Beckman Coulter Inc. recalls DxC 500 AU Clinical Chemistry Analyzer, Catalog Numbers/UDI codes: C63519 / 14987666545058 C63520 / 14987666545065 The Beckman Coulter DxC…
Beckman Coulter has determined that device software versions V1.3, V1.4, V1.4.1, V1.4.2, V1.4.3. have a defect due of which, when all constituent tests of a calculated result are rerun, either manually ordered by the…
FDA · DeviceBeckman Coulter Inc.Worldwide Distribution: U.S. (nationwide) to states of: AL, CA, CO, GA, ID,…Official notice →- 2025-03-26
US
Beckman Coulter, Inc. recalls DxI 9000 Access Immunoassay Analyzer, Catalog Number C11137.
the DxI 9000 encoder wheel was misaligned on the motor shaft, leading to increased slippage errors and causing the instrument to stop, posing a high safety risk due to potential delays in patient results.
US
Beckman Coulter, Inc. recalls DxI 9000 Access Immunoassay Analyzer, Catalog Number C11137, with Component PCB, TSI/RSI firmware C29278
Beckman Coulter identified that an issue for DxI 9000 tip stocker interface (TSI) and reagent storage interface (RSI) boards with the firmware version 1.1.0.0 can cause sub-system communication failures. This failure…
FDA · DeviceBeckman Coulter, Inc.Worldwide - US Nationwide distribution in the states of FL, TX, WA.Official notice →
Other drugs & medical recalls in the United States
- 2025-04-16
US
Tornier S.A.S. recalls stryker Blueprint Mixed Reality Coracoid Clamp, Catalog Number MRUE202, packaged in stryker Blueprint Mixed Reality Instrument Kit,…
Stryker identified an elevated complaint rate related to difficulties using the Blueprint Mixed Reality Guidance Instrumentation system is multifactorial, but in each instance prohibits the system from meeting its…
FDA · DeviceTornier S.A.S.US: TX, MN, AK, PA, MA, TN, ID, and France, Canada, and IrelandOfficial notice →US
Omnicell, Inc. recalls IVX Station Reconstitution Transfer Set. Model Number: 298921215. Packaging: 1 set per White Tyvek Pouch, 5 Tyvek Pouches per shelf box,…
IVX Station Reconstitution Transfer Sets were shipped in advance of the IVX Station upgrade and not intended for clinical use but are engineering devices intended for use during set up and system calibration. The items…
US
MEDLINE INDUSTRIES, LP - Northfield recalls Wayson Hydrocolloid Models: 1) MJ BANDAGE HYDROCOLLOID HEEL OVAL 6CT, Model Number: MJ18004V2; 2) MJ BANDAGE HYDROCOLLOID…
There is a potential for the sterility of the device to be compromised.
FDA · DeviceMEDLINE INDUSTRIES, LP - NorthfieldWorldwide Distribution: US (Nationwide) and OUS (Foreign) to countries of:…Official notice →US
BioFire Diagnostics, LLC recalls FilmArra Pneumonia Panel plus (Pneumoplus), REF: RFIT-LBL-0374
Increased risk of control failures and false negative test results with multiplexed nucleic acid test.
FDA · DeviceBioFire Diagnostics, LLCOUS International Distribution to countries of: France, Italia, Spain,…Official notice →US
Qapel Medical Inc. recalls Hippo 072 Aspiration System including Cheetah Delivery Tool and Aspiration Tubing, REF: APT6072-132. Hippo 072 Aspiration System including…
Aspiration catheter distal tip features and characteristics may not be in the scope of FDA clearance; tip detachment, vessel rupture, and vasospasm could occur.
FDA · DeviceQapel Medical Inc.US: CA, IL, NE, TN, TX, NJ, NC, SD, FL, MI, PA, GA, AL, IN, NY, WI, AZ, MN,…Official notice →US
MEDLINE INDUSTRIES, LP - Northfield recalls Medline medical convenience kits labeled as: 1) NEONATAL INTUBATION TRAY, REF ACC010371C; 2) NEWBORN RESUSCITATION KIT, REF…
Specific Medline Kits were manufactured using Intubation ORAL/NASAL Endotracheal Tubes which were subsequently recalled by Smiths Medical.