Beckman Coulter, Inc. recalls Discrete photometric chemistry analyzer for clinical use.
Beckman Coulter, Inc. has identified a rise in reagent carousel motion errors in Access 2 Immunoassay System Analyzers that appear to be caused by misaligned index wheels housed within the reagent carousel module.…
Read the official notice →- Date
- 2024-08-21
- Country
- United States
- Brand
- Beckman Coulter
- Product
- Discrete photometric chemistry analyzer for clinical use.
- Classification
- Class II
- Units
- Worldwide - US Nationwide distribution in the states of NE, MT, LA, TX, NY,…
- Category
- Drugs & medical
- Source
- FDA · Device
Copied automatically from the US Food and Drug Administration. Details may be shortened or out of date; the official notice is the authoritative source. Disclaimer
What to do
- Check the product name, strength and lot number against the official notice.
- Do not stop a prescribed medicine without talking to your doctor or pharmacist first.
- Ask the pharmacy or supplier about a replacement or refund.
General guidance from Recall Atlas. The official notice from the US Food and Drug Administration has the exact models, batch codes and remedy, and it takes priority.
More Beckman Coulter recalls
- 2024-09-25
US
Beckman Coulter, Inc. recalls Beckman Coulter DxI 9000 Access Immunoassay Analyzer, Model Number C21510, RSM Pneumatic Tubing Kit item #2, REF C11137; in vitro…
Beckman Coulter is recalling the DxI 9000 Access Immunoassay Analyzer due to undersized Pneumatics Tubing in the reagent storage module causing air leaks, compressor failures or sample processing errors. Four…
FDA · DeviceBeckman Coulter, Inc.Worldwide distribution - US Nationwide and the countries of Australia,…Official notice →- 2024-09-11
US
Beckman Coulter, Inc. recalls DxI 9000 Access Immunoassay Analyzer, REF C11137, vitro diagnostic device used for the quantitative, semi-quantitative, or qualitative…
Beckman Coulter manufacturing discovered cracks around a threaded insert on the bottom of the Internal Wash Buffer Tank that is used within the Wash Buffer Supply Drawer on the DxI 9000.This can result in leaking,…
FDA · DeviceBeckman Coulter, Inc.Worldwide - US Nationwide distribution in the state of New Hampshire and the…Official notice →- 2024-09-04
US
Beckman Coulter, Inc. recalls DxI 9000 Access Immunoassay Analyzer, Catalog #C11137.
When a DxI 9000 Access Immunoassay Analyzer is connected to a Laboratory Automation System (LAS) and is running with the system software versions 1.17.0 and below, the analyzer does not send a message to the LAS when…
FDA · DeviceBeckman Coulter, Inc.Worldwide - US Nationwide distribution in the states of CA, FL, GA, IN, KS,…Official notice →- 2024-08-21
US
Beckman Coulter, Inc. recalls Beckman Coulter DxI 9000 Access Immunoassay Analyzer, Catalog #C11137.
The reason for this recall is Beckman Coulter has determined that when a DxI 9000 Access Immunoassay Analyzer is connected to a Laboratory Automation System (LAS); is running with the software versions 1.17 and below;…
FDA · DeviceBeckman Coulter, Inc.Worldwide distribution - US Nationwide distribution in the states of CA, FL,…Official notice →- 2024-07-31
US
Beckman Coulter Inc. recalls Beckman Coulter, REF: OSR61173, IGM 4x14 mL R1, 4x11 mL R2
LIH Influence check settings for Lipemia are incorrect which may result patient samples being reported as false low results or cause a high result to report as normal
FDA · DeviceBeckman Coulter Inc.Worldwide - US Nationwide distribution including in the states of AL, AR, AZ,…Official notice →US
Beckman Coulter, Inc. recalls Dxl 9000 Access Immunoassay Analyzer, Catalog #C11137.
The analyzers were built with nonconforming USB cables made with brass contacts instead of specified gold contacts. Brass contacts may corrode over time, leading to loss of internal instrument communication and delayed…
FDA · DeviceBeckman Coulter, Inc.Worldwide - US Nationwide distribution in the states of FL, IN, KS, NH, PA,…Official notice →
Other drugs & medical recalls in the United States
- 2024-08-21
US
Thoratec LLC recalls HeartMate 3 System Controllers provided within the following HeartMate 3 Left Ventricular Assist System (LVAS) Implant Kits or distributed…
Left ventricular assist system controller UI membrane/screen may lift along the edge of the controller housing, and if controller is exposed to fluid, ingress may occur, which may damage printed circuit board…
FDA · DeviceThoratec LLCUS: MN, PA, FL, IL, NY, DE, AR, NJ, NC, GA, WI, AZ, TN, MO, TX, MA, OH, VA,…Official notice →US
Ortho-Clinical Diagnostics, Inc. recalls VITROS Immunodiagnostic VITROS Folate Reagent Pack 1/2
It was confirmed that fibrinogen in patient plasma samples precipitates out of solution upon the addition of folate stabilizer reagent as part of the pre-treatment process causing the increase in "TM5-4MB" condition…
FDA · DeviceOrtho-Clinical Diagnostics, Inc.Worldwide distribution - US Nationwide and the countries of Australia,…Official notice →US
Merit Medical Systems, Inc. recalls Various Merit Medical convenience kits that contain recalled Jiangsu Shenli Medical Production Co. Ltd. syringes: Rapid Response…
Medical convenience kits contain plastic syringes affected by the FDA Safety Alert issued on March 19, 2024. Leaks, breakage, and/ other quality issues have been identified that may pose a risk to patient health.
FDA · DeviceMerit Medical Systems, Inc.U.S. Nationwide distribution including in the states of AK, AL, AR, AZ, CA,…Official notice →US
Beckman Coulter, Inc. recalls Beckman Coulter DxI 9000 Access Immunoassay Analyzer, Catalog #C11137.
The reason for this recall is Beckman Coulter has determined that when a DxI 9000 Access Immunoassay Analyzer is connected to a Laboratory Automation System (LAS); is running with the software versions 1.17 and below;…
FDA · DeviceBeckman Coulter, Inc.Worldwide distribution - US Nationwide distribution in the states of CA, FL,…Official notice →US
Defibtech, LLC recalls RMU-2000 Automated Chest Compression Device
Problem in the device's motor may cause the device to stop compressions. This may lead to a delay in therapy if the device fails to operate, and could lead to patient injury or death.
FDA · DeviceDefibtech, LLCDomestic distribution to CT, FL, NJ, SC, TN. Foreign distribution to JapanOfficial notice →US
ICU Medical Inc recalls Replacement Battery List Number SUB0000594 and SUB0000864 found in Plum A+ & Plum A+3 Infusion Systems, Pump List Numbers: 11005, 11971,…
Due to a manufacturing defect from the battery supplier, affected batteries may experience a loss of capacity earlier in the battery lifecycle than expected and overall battery runtime may decrease earlier in the…
FDA · DeviceICU Medical IncWorldwide Distribution: US (nationwide) and OUS (foreign) countries of:…Official notice →