Beckman Coulter, Inc. recalls DxI 9000 Access Immunoassay Analyzer, Catalog #C11137.

When a DxI 9000 Access Immunoassay Analyzer is connected to a Laboratory Automation System (LAS) and is running with the system software versions 1.17.0 and below, the analyzer does not send a message to the LAS when…

Read the official notice →
Date
2024-09-04
Country
United States
Brand
Beckman Coulter
Product
DxI 9000 Access Immunoassay Analyzer, Catalog #C11137.
Classification
Class II
Units
Worldwide - US Nationwide distribution in the states of CA, FL, GA, IN, KS,…
Category
Drugs & medical
Source
FDA · Device

Copied automatically from the US Food and Drug Administration. Details may be shortened or out of date; the official notice is the authoritative source. Disclaimer

What to do

  1. Check the product name, strength and lot number against the official notice.
  2. Do not stop a prescribed medicine without talking to your doctor or pharmacist first.
  3. Ask the pharmacy or supplier about a replacement or refund.

General guidance from Recall Atlas. The official notice from the US Food and Drug Administration has the exact models, batch codes and remedy, and it takes priority.

More Beckman Coulter recalls

  1. 2024-10-16
  2. US
    Drugs & medicalClass II

    Beckman Coulter, Inc. recalls Beckman Coulter Dxl 9000 Access Immunoassay Analyzer, REF #C11137. An in vitro diagnostic device used for the quantitative,…

    When the DxI 9000 Access Immunoassay Analyzer with system software version 1.16.2 and prior is configured to automatically request a calibration order, the instrument can process the order with expired calibrator…

    FDA · DeviceBeckman Coulter, Inc.Distribution was made to FL, GA, IN, KS, NH, NJ, OK, PA, and TX. There was…Official notice →
  3. 2024-09-25
  4. US
    Drugs & medicalClass II

    Beckman Coulter, Inc. recalls Beckman Coulter DxI 9000 Access Immunoassay Analyzer, Model Number C21510, RSM Pneumatic Tubing Kit item #2, REF C11137; in vitro…

    Beckman Coulter is recalling the DxI 9000 Access Immunoassay Analyzer due to undersized Pneumatics Tubing in the reagent storage module causing air leaks, compressor failures or sample processing errors. Four…

    FDA · DeviceBeckman Coulter, Inc.Worldwide distribution - US Nationwide and the countries of Australia,…Official notice →
  5. 2024-09-11
  6. US
    Drugs & medicalClass II

    Beckman Coulter, Inc. recalls DxI 9000 Access Immunoassay Analyzer, REF C11137, vitro diagnostic device used for the quantitative, semi-quantitative, or qualitative…

    Beckman Coulter manufacturing discovered cracks around a threaded insert on the bottom of the Internal Wash Buffer Tank that is used within the Wash Buffer Supply Drawer on the DxI 9000.This can result in leaking,…

    FDA · DeviceBeckman Coulter, Inc.Worldwide - US Nationwide distribution in the state of New Hampshire and the…Official notice →
  7. 2024-08-21
  8. US
    Drugs & medicalClass II

    Beckman Coulter, Inc. recalls Discrete photometric chemistry analyzer for clinical use.

    Beckman Coulter, Inc. has identified a rise in reagent carousel motion errors in Access 2 Immunoassay System Analyzers that appear to be caused by misaligned index wheels housed within the reagent carousel module.…

    FDA · DeviceBeckman Coulter, Inc.Worldwide - US Nationwide distribution in the states of NE, MT, LA, TX, NY,…Official notice →
  9. US
    Drugs & medicalClass II

    Beckman Coulter, Inc. recalls Beckman Coulter DxI 9000 Access Immunoassay Analyzer, Catalog #C11137.

    The reason for this recall is Beckman Coulter has determined that when a DxI 9000 Access Immunoassay Analyzer is connected to a Laboratory Automation System (LAS); is running with the software versions 1.17 and below;…

    FDA · DeviceBeckman Coulter, Inc.Worldwide distribution - US Nationwide distribution in the states of CA, FL,…Official notice →
  10. 2024-07-31
  11. US
    Drugs & medicalClass II

    Beckman Coulter Inc. recalls Beckman Coulter, REF: OSR61173, IGM 4x14 mL R1, 4x11 mL R2

    LIH Influence check settings for Lipemia are incorrect which may result patient samples being reported as false low results or cause a high result to report as normal

    FDA · DeviceBeckman Coulter Inc.Worldwide - US Nationwide distribution including in the states of AL, AR, AZ,…Official notice →
All 84 Beckman Coulter recalls →

Other drugs & medical recalls in the United States

  1. 2024-09-04
  2. US
    Drugs & medicalClass II

    Agfa N.V. recalls DR 800. Digital Radiography X-ray System.

    Potential for the front lever chain of the DR 800 table to fail.

    FDA · DeviceAgfa N.V.Worldwide - US Nationwide distribution in the states of CA, FL, IL, IN, KS,…Official notice →
  3. US
    Drugs & medicalClass II

    RONAN MEDICAL LLC recalls BrainFeedback Pro 2 Device - is intended as an instrument that provides a visual and auditory signal corresponding to the status of a…

    Neurofeedback systems removed due to missing design and development documentation and devices may not perform as intended.

    FDA · DeviceRONAN MEDICAL LLCWorldwide - US Nationwide distribution in the states of CA, FL, TN, MI, IL, CT…Official notice →
  4. US
    Drugs & medicalClass II

    Abbott Diagnostics Technologies AS recalls Afinion 2 Analyzer, model #1116986, US refurbished, CLIA waived.

    Five Afinion 2 (AF2) Analyzer were incorrectly market configured as Moderate Complexity and distributed to customers. Use of moderately complex assays by waived laboratories could result in erroneous results.

    FDA · DeviceAbbott Diagnostics Technologies ASUS Nationwide distribution in the states of AZ, IN, MA, MN, and OH.Official notice →
  5. US
    Drugs & medicalClass II

    Schiller, Ag recalls CARDIOVIT AT-180. Model Number: 0A.110000; Catalog number: 3.920570

    Potential for high-frequency signal artifacts is recorded during an ECG acquisition performed by CARDIOVIT AT-180 electrocardiographs.

    FDA · DeviceSchiller, AgWorldwide - US distribution in the state of Florida and the countries of…Official notice →
  6. US
    Drugs & medicalClass II

    GE Healthcare (China) Co., Ltd. recalls GE Healthcare Proteus XR/a, stationary X-Ray System, Model/Catalog Numbers: 1) 2259973; 2) 2259988; 3) 2266999; 4) 2373519; 5)…

    The XR system, Proteus XR/a, does not have a de-installation manual describing process steps for de-installing the devices. De-installation instruction process steps must be followed when de-installing these devices to…

    FDA · DeviceGE Healthcare (China) Co., Ltd.Worldwide distribution.Official notice →
  7. US
    Drugs & medicalClass II

    Howmedica Osteonics Corp. recalls Custom Made Device - GMRS EXTENSION Piece 40MM LEFT, 35 Deg EXT Rotated Tabs (Patient Specific); Catalog Number: C-MM100434-00;

    A patient specific, custom-made device was supplied for a surgical procedure that did not match the design proposal.

    FDA · DeviceHowmedica Osteonics Corp.International distribution to the country of Australia.Official notice →
See all drugs & medical recalls →