Beckman Coulter, Inc. recalls Beckman Coulter Dxl 9000 Access Immunoassay Analyzer, REF #C11137. An in vitro diagnostic device used for the quantitative,…
When the DxI 9000 Access Immunoassay Analyzer with system software version 1.16.2 and prior is configured to automatically request a calibration order, the instrument can process the order with expired calibrator…
Read the official notice →- Date
- 2024-10-16
- Country
- United States
- Brand
- Beckman Coulter
- Product
- Beckman Coulter Dxl 9000 Access Immunoassay Analyzer, REF #C11137. An in vitro diagnostic device used for the quantitative,…
- Classification
- Class II
- Units
- Distribution was made to FL, GA, IN, KS, NH, NJ, OK, PA, and TX. There was…
- Category
- Drugs & medical
- Source
- FDA · Device
Copied automatically from the US Food and Drug Administration. Details may be shortened or out of date; the official notice is the authoritative source. Disclaimer
What to do
- Check the product name, strength and lot number against the official notice.
- Do not stop a prescribed medicine without talking to your doctor or pharmacist first.
- Ask the pharmacy or supplier about a replacement or refund.
General guidance from Recall Atlas. The official notice from the US Food and Drug Administration has the exact models, batch codes and remedy, and it takes priority.
More Beckman Coulter recalls
- 2024-10-23
US
Beckman Coulter Inc. recalls IMMAGE Immunochemistry Systems IGM Immunoglobulin M Reagent REF 447610 IGM reagent, when used in conjunction with IMMAGE Immunochemistry…
Firm identified two lots of reagents that are not meeting the sensitivity statement in the IFU
FDA · DeviceBeckman Coulter Inc.U.S.: AZ, CA, CO, FL, GA, HI, IL, MA, MI, MO, ND, NY, OH, PA, and TX O.U.S.:…Official notice →US
Beckman Coulter, Inc. recalls The Access TSH (3rd IS) assay is a paramagnetic particle, chemiluminescent immunoassay for the quantitative etermdination of human…
Certain lots of Access TSH reagent packs may produce erroneously low TSH (thyroid-stimulating hormone) results.
FDA · DeviceBeckman Coulter, Inc.Worldwide distribution - US Nationwide and the countries of Austria, Belgium,…Official notice →US
Beckman Coulter, Inc. recalls Beckman Coulter IRISPEC CA/CB/CC Urine Chemistry Controls, Catalog Number REF 800-7211 and 800-7702, in vitro diagnostic device.
Beckman Coulter has become aware of an increase in customer complaints for intermittent IRISpec CB negative glucose control failure, resulting in false positive control results when used in conjunction with…
- 2024-09-25
US
Beckman Coulter, Inc. recalls Beckman Coulter DxI 9000 Access Immunoassay Analyzer, Model Number C21510, RSM Pneumatic Tubing Kit item #2, REF C11137; in vitro…
Beckman Coulter is recalling the DxI 9000 Access Immunoassay Analyzer due to undersized Pneumatics Tubing in the reagent storage module causing air leaks, compressor failures or sample processing errors. Four…
FDA · DeviceBeckman Coulter, Inc.Worldwide distribution - US Nationwide and the countries of Australia,…Official notice →- 2024-09-11
US
Beckman Coulter, Inc. recalls DxI 9000 Access Immunoassay Analyzer, REF C11137, vitro diagnostic device used for the quantitative, semi-quantitative, or qualitative…
Beckman Coulter manufacturing discovered cracks around a threaded insert on the bottom of the Internal Wash Buffer Tank that is used within the Wash Buffer Supply Drawer on the DxI 9000.This can result in leaking,…
FDA · DeviceBeckman Coulter, Inc.Worldwide - US Nationwide distribution in the state of New Hampshire and the…Official notice →- 2024-09-04
US
Beckman Coulter, Inc. recalls DxI 9000 Access Immunoassay Analyzer, Catalog #C11137.
When a DxI 9000 Access Immunoassay Analyzer is connected to a Laboratory Automation System (LAS) and is running with the system software versions 1.17.0 and below, the analyzer does not send a message to the LAS when…
FDA · DeviceBeckman Coulter, Inc.Worldwide - US Nationwide distribution in the states of CA, FL, GA, IN, KS,…Official notice →
Other drugs & medical recalls in the United States
- 2024-10-16
US
MEDLINE INDUSTRIES, LP - Northfield recalls MEDLINE Automatic Digital Blood Pressure Monitor, REF MDS1001UT.
It has been determined that certain Blood Pressure Monitors containing a faulty microchip have failed to power on or have shown display screen issues after the initial few uses, as well as overheating at times making…
FDA · DeviceMEDLINE INDUSTRIES, LP - NorthfieldWorldwide distribution - US Nationwide and the countries of Panama and Jamaica.Official notice →US
Windstone Medical Packaging, Inc. recalls Aligned Medical Septo Pack REF AMS11480 that contains the Cardinal Health Salem Sump PVC tubes. Product packaged in a…
Due to convenience kits containing a kit component that is under a recall action for improper use.
FDA · DeviceWindstone Medical Packaging, Inc.U.S. Nationwide distribution in the states of CA, MO, OH, VA, and WY.Official notice →US
Becton Dickinson & Co. recalls BD Synapsys Informatics Solution with software versions: 4.20, 4.30, 5.10, 5.20 and 5.30- IVD standalone laboratory software solution…
Synapsys software versions 4.20, 4.30, 5.10, 5.20, or 5.30 and utilizing the in-line breakpoint feature, an error is occurring where BD Synapsys will incorrectly display breakpoints for organisms that do not have…
FDA · DeviceBecton Dickinson & Co.Nationwide Foreign: Australia, New Zealand, Singapore, Hong Kong, Taiwan,…Official notice →US
Atrium Medical Corporation recalls iCast Covered Stent, 6MMx16MMx120CM, Model Number 42616
There is an error in the Unique Device Identifier (UDI) printed on the product's distribution label.
FDA · DeviceAtrium Medical CorporationUS Nationwide distribution in the states of FL, MA, MI, OH, WA.Official notice →US
Howmedica Osteonics Corp. recalls Triathlon Universal TS Baseplate Size 2 - A single use sterile implantable device intended for cemented applications during primary and…
Anterior locking tabs on the Triathlon Universal Baseplate are out of specification which may prevent engagement of either of the two locking tabs with the locking wire of the tibial insert.
FDA · DeviceHowmedica Osteonics Corp.US Nationwide distribution in the states of AZ, FL, NC, OH, PA, TN, WI.…Official notice →US
Qualigen Inc recalls FastPack TSH Calibrator Kit, REF: 25000024, For use in conjunction with the FastPack IP TSH Immunoassay and FastPack IP System Analyzer
Thyroid Stimulating Hormone (TSH) calibrator card has an incorrect barcode linked to incorrect information, that may allow successful system calibration, but the TSH results obtained could be higher than expected.
FDA · DeviceQualigen IncUS: TX, MS, NY, OH, MT, WA, PA, MN, SC, GA, NC, NV, WY, AK, SD, KS, NE, VT,…Official notice →