Drugs & medicalClass II

Beckman Coulter, Inc. recalls Beckman Coulter DxI 9000 Access Immunoassay Analyzer, Model Number C21510, RSM Pneumatic Tubing Kit item #2, REF C11137; in vitro…

Beckman Coulter is recalling the DxI 9000 Access Immunoassay Analyzer due to undersized Pneumatics Tubing in the reagent storage module causing air leaks, compressor failures or sample processing errors. Four…

Read the official notice →
Date
2024-09-25
Country
United States
Brand
Beckman Coulter
Product
Beckman Coulter DxI 9000 Access Immunoassay Analyzer, Model Number C21510, RSM Pneumatic Tubing Kit item #2, REF C11137; in vitro…
Classification
Class II
Units
Worldwide distribution - US Nationwide and the countries of Australia,…
Category
Drugs & medical
Source
FDA · Device

Copied automatically from the US Food and Drug Administration. Details may be shortened or out of date; the official notice is the authoritative source. Disclaimer

What to do

  1. Check the product name, strength and lot number against the official notice.
  2. Do not stop a prescribed medicine without talking to your doctor or pharmacist first.
  3. Ask the pharmacy or supplier about a replacement or refund.

General guidance from Recall Atlas. The official notice from the US Food and Drug Administration has the exact models, batch codes and remedy, and it takes priority.

More Beckman Coulter recalls

  1. 2024-10-23
  2. US
    Drugs & medicalClass II

    Beckman Coulter, Inc. recalls The Access TSH (3rd IS) assay is a paramagnetic particle, chemiluminescent immunoassay for the quantitative etermdination of human…

    Certain lots of Access TSH reagent packs may produce erroneously low TSH (thyroid-stimulating hormone) results.

    FDA · DeviceBeckman Coulter, Inc.Worldwide distribution - US Nationwide and the countries of Austria, Belgium,…Official notice →
  3. US
    Drugs & medicalClass III

    Beckman Coulter, Inc. recalls Beckman Coulter IRISPEC CA/CB/CC Urine Chemistry Controls, Catalog Number REF 800-7211 and 800-7702, in vitro diagnostic device.

    Beckman Coulter has become aware of an increase in customer complaints for intermittent IRISpec CB negative glucose control failure, resulting in false positive control results when used in conjunction with…

    FDA · DeviceBeckman Coulter, Inc.worldwideOfficial notice →
  4. 2024-10-16
  5. US
    Drugs & medicalClass II

    Beckman Coulter, Inc. recalls Beckman Coulter Dxl 9000 Access Immunoassay Analyzer, REF #C11137. An in vitro diagnostic device used for the quantitative,…

    When the DxI 9000 Access Immunoassay Analyzer with system software version 1.16.2 and prior is configured to automatically request a calibration order, the instrument can process the order with expired calibrator…

    FDA · DeviceBeckman Coulter, Inc.Distribution was made to FL, GA, IN, KS, NH, NJ, OK, PA, and TX. There was…Official notice →
  6. 2024-09-11
  7. US
    Drugs & medicalClass II

    Beckman Coulter, Inc. recalls DxI 9000 Access Immunoassay Analyzer, REF C11137, vitro diagnostic device used for the quantitative, semi-quantitative, or qualitative…

    Beckman Coulter manufacturing discovered cracks around a threaded insert on the bottom of the Internal Wash Buffer Tank that is used within the Wash Buffer Supply Drawer on the DxI 9000.This can result in leaking,…

    FDA · DeviceBeckman Coulter, Inc.Worldwide - US Nationwide distribution in the state of New Hampshire and the…Official notice →
  8. 2024-09-04
  9. US
    Drugs & medicalClass II

    Beckman Coulter, Inc. recalls DxI 9000 Access Immunoassay Analyzer, Catalog #C11137.

    When a DxI 9000 Access Immunoassay Analyzer is connected to a Laboratory Automation System (LAS) and is running with the system software versions 1.17.0 and below, the analyzer does not send a message to the LAS when…

    FDA · DeviceBeckman Coulter, Inc.Worldwide - US Nationwide distribution in the states of CA, FL, GA, IN, KS,…Official notice →
  10. 2024-08-21
  11. US
    Drugs & medicalClass II

    Beckman Coulter, Inc. recalls Discrete photometric chemistry analyzer for clinical use.

    Beckman Coulter, Inc. has identified a rise in reagent carousel motion errors in Access 2 Immunoassay System Analyzers that appear to be caused by misaligned index wheels housed within the reagent carousel module.…

    FDA · DeviceBeckman Coulter, Inc.Worldwide - US Nationwide distribution in the states of NE, MT, LA, TX, NY,…Official notice →
All 84 Beckman Coulter recalls →

Other drugs & medical recalls in the United States

  1. 2024-09-25
  2. US
    Drugs & medicalClass II

    Unomedical A/S recalls VariSoft Infusion Set, Single Use.

    Specific VariSoftTM Infusion Sets were packaged using incorrect secondary packaging (boxes) branded "Trusteel Infusion Set".

    FDA · DeviceUnomedical A/SUS Nationwide distribution in the state of California.Official notice →
  3. US
    Drugs & medicalClass II

    Cypress Medical Products LLC recalls BinaxNOW Respiratory Strep A Test Kit, CLIA Waived, Catalog Number 730025

    Due to a classification error in our inventory management system, these medical devices were shipped to customers that did not have the requisite license to purchase such devices.

    FDA · DeviceCypress Medical Products LLCUS Nationwide distribution.Official notice →
  4. US
    Drugs & medicalClass II

    Safeguard US Operating LLC recalls HAWK Advanced Hypothermia Management Set, Item Number: 59-320

    Incorrect expiration date.

    FDA · DeviceSafeguard US Operating LLCUS Nationwide distribution in the states of AK, CA, CO, DC, GA, IL, NC, NM,…Official notice →
  5. US
    Drugs & medicalClass II

    B Braun Medical Inc recalls SL BLOODLINE FOR FMC 2008 SERIES-Tubing to be used to transfer blood between a patient and a haemodialyser during haemodialysis.…

    Incomplete insertion of the patient connector of the Streamline Airless System Hemodialysis Bloodline to the Locksite may result in the potential for occlusion during prograde rinseback procedures. If the patient…

    FDA · DeviceB Braun Medical IncUS Nationwide distribution.Official notice →
  6. US
    Drugs & medicalClass II

    Smith & Nephew, Inc. recalls DIAMOND POINT, BOTH ENDS K-WIRE, REF 128042, metallic bone fixation fastener

    Smith & Nephew, Inc., has initiated a field action to voluntarily remove two batches of K-Wires due to a packaging error. A complaint was received indicating that a package contained K-WIRE .062X9 2PT DM wires instead…

    FDA · DeviceSmith & Nephew, Inc.US, Canada, NorwayOfficial notice →
  7. US
    Drugs & medicalClass II

    Siemens Healthcare Diagnostics, Inc. recalls Atellica CH Microalbumin_2 (¿ALB_2)-For in vitro diagnostic use in the quantitative measurement of microalbumin in human…

    Atellica CH Microalbumin_2 (¿ALB_2) and the ADVIA Chemistry Microalbumin_2 (¿ALB_2) are not meeting the High-Dose Hook Effect claim/Prozone Effect claim as stated in the Instructions for Use (IFU) on the Atellica CH,…

    FDA · DeviceSiemens Healthcare Diagnostics, Inc.Worldwide distribution - US Nationwide and the countries of Argentina,…Official notice →
See all drugs & medical recalls →