Drugs & medicalClass II

Beckman Coulter, Inc. recalls BECKMAN COULTER IRISpec CA/CB/CC control, REF 900-7211, For in vitro diagnostic use only.

Beckman Coulter has become aware of an increase in customer complaints generated for intermittent IRISpec CB glucose control false positive results when used in conjunction with iChemVelocity Urine Chemistry Strips PN…

Read the official notice →
Date
2023-01-11
Country
United States
Brand
Beckman Coulter
Product
BECKMAN COULTER IRISpec CA/CB/CC control, REF 900-7211, For in vitro diagnostic use only.
Classification
Class II
Units
US and Canada, Brazil, Korea, Mexico, Puerto Rico, Taiwan
Category
Drugs & medical
Source
FDA · Device

Copied automatically from the US Food and Drug Administration. Details may be shortened or out of date; the official notice is the authoritative source. Disclaimer

What to do

  1. Check the product name, strength and lot number against the official notice.
  2. Do not stop a prescribed medicine without talking to your doctor or pharmacist first.
  3. Ask the pharmacy or supplier about a replacement or refund.

General guidance from Recall Atlas. The official notice from the US Food and Drug Administration has the exact models, batch codes and remedy, and it takes priority.

More Beckman Coulter recalls

  1. 2023-05-24
  2. US
    Drugs & medicalClass II

    Beckman Coulter, Inc. recalls DxU 840m Iris Urine Microscopy Analyzer, Catalog Number C76947, in-vitro diagnostic device, urine analysis

    Beckman Coulter has become aware of an intermittent issue in which the optional flag "Previous Sample Had Sperm" was enabled but not displayed so that a carryover event could have been investigated.

    FDA · DeviceBeckman Coulter, Inc.Worldwide DistributionOfficial notice →
  3. 2023-04-26
  4. US
    Drugs & medicalClass II

    Beckman Coulter, Inc. recalls Beckman Coulter Access Free T3, Triiodothyronine, Free, Catalog #A13422, each box contains 2 reagent packs.

    The reagent lot produces elevated results (approximately >30%) in some (15-20%)patient samples that do not correlate with the results of other thyroid markers nor Free T3 results generated on other platforms. This…

    FDA · DeviceBeckman Coulter, Inc.Worldwide distribution - US Nationwide, including Puerto Rico and the…Official notice →
  5. 2023-03-22
  6. US
    Drugs & medicalClass II

    Beckman Coulter, Inc. recalls Access Immunoassay Systems (1) Access 2 Immunoassay Analyzer (81600N); (2) Access 2 Immunoassay Analyzer, Refurbished (386220); (3) Access…

    An increased incidence of damage to the nut band from a new vendor (Boamax) has led to an increased volume of complaints from consignees regarding wash arm motion failures, system check failures, and/or assay…

    FDA · DeviceBeckman Coulter, Inc.Worldwide and Nationwide Distribution. There was government/military…Official notice →
  7. 2022-11-30
  8. US
    Drugs & medicalClass II

    Beckman Coulter, Inc. recalls MicroScan autoSCAN-4 Instrument, Catalog Number B1018-280

    Due to MicroScan autoSCAN-4 Instruments being manufactured with an incorrect diffuser plate.

    FDA · DeviceBeckman Coulter, Inc.Worldwide - U.S. Nationwide distribution in the states of AK, FL, IA, ID, MI…Official notice →
  9. 2022-05-25
  10. US
    Drugs & medicalClass II

    Beckman Coulter Inc. recalls MicroScan WalkAway-40 plus Instrument REF B1018-283

    WalkAway Instruments may have been assembled with the incorrect AC power entry connector inlet with is rated at 1 ampere current rating rather than the correct current rating of 10 ampere

    FDA · DeviceBeckman Coulter Inc.Worldwide distribution - U.S. Nationwide distribution in the states of Alaska,…Official notice →
  11. 2022-04-27
  12. US
    Drugs & medicalClass III

    Beckman Coulter, Inc. recalls Beckman Coulter IRISpec CA/CB/CC control, Catalog #800-7211, containing 3 x 100 mL bottles each of CA Control, CB Control, and CC Control.

    IRISpec CA failures for glucose.

    FDA · DeviceBeckman Coulter, Inc.Worldwide distribution - US Nationwide, including Puerto Rico and the…Official notice →
All 84 Beckman Coulter recalls →

Other drugs & medical recalls in the United States

  1. 2023-01-11
  2. US
    Drugs & medicalClass II

    Biomerieux Inc recalls PPM LOMBARD PRODUCTS TSA 10PLT, CATALOG M1040

    As the temperature and time out of range was exceeded, product performances cannot be guaranteed

    FDA · DeviceBiomerieux IncUS Nationwide distribution.Official notice →
  3. US
    Drugs & medicalClass II

    Beaver Visitec International, Inc. recalls Beaver EdgeAhead Stiletto/MVR Knife .90mm Straight (20G)-for ocular paracentesis incisions in cataract procedures. Part…

    Packaging sterile barrier has been compromised. Open seals can lead to a loss of device sterility and may lead to infection

    FDA · DeviceBeaver Visitec International, Inc.US Nationwide distribution.Official notice →
  4. US
    Drugs & medicalClass II

    Steris Corporation recalls VERIFY Steam Test Pack (20 Tests and 20 controls) Item Number: S3069 Intended for use in installation testing and routine monitoring of…

    Biological indicator (BI) used to promote growth of the G. stearothermophilus and B. atrophaeus spores was inconsistent in promoting growth throughout lot , some BIs within the lot performed as intended while others…

    FDA · DeviceSteris CorporationNationwide Foreign: AE AT CA CO ES FI GU IN IT JP MX SG TH TW ZAOfficial notice →
  5. US
    Drugs & medicalClass II

    Medtronic MiniMed recalls Guardian iOS app (MMT-8200) and Guardian Android app (MMT-8201), part of the Guardian 4 system

    An app, part of a continuous glucose monitoring system, for use with smartphone devices may automatically log out from CareLink, then the app is not able to upload data. When logged out, linked care partners will not…

    FDA · DeviceMedtronic MiniMedOUS: Austria, Belgium, Chile, Colombia, Czech Republic, Denmark, Finland,…Official notice →
  6. US
    Drugs & medicalClass II

    Siemens Medical Solutions USA, Inc recalls Programmable Diagnostic Computer

    The firm will be performing a software update to address a software error which affects the listed products. This correction addresses four potential software issues: 1) "PASSWORD STORE CORRUPTED" error message during…

    FDA · DeviceSiemens Medical Solutions USA, IncDomestic distribution nationwide. Worldwide foreign distribution to Albania…Official notice →
  7. US
    Drugs & medicalClass II

    ENDO Pharmaceuticals Solutions, Inc. recalls 9 Gauge Disposable Trocars with Blunt Pushers for Use with Subcutaneous Implants Part Number: 50134-01

    No expiration dates printed on the packaging

    FDA · DeviceENDO Pharmaceuticals Solutions, Inc.NationwideOfficial notice →