Drugs & medicalClass II

Beckman Coulter, Inc. recalls Access Thyroglobulin assay, a simultaneous one-step immunoenzymatic assay, Catalog Number 33860

Access Thyroglobulin reagent lot 439163 may generate erroneously high patient results. Falsely increased Access Thyroglobulin results may lead a physician to pursue unnecessary diagnostic imaging studies and/or…

Read the official notice →
Date
2025-08-20
Country
United States
Brand
Beckman Coulter
Product
Access Thyroglobulin assay, a simultaneous one-step immunoenzymatic assay, Catalog Number 33860
Classification
Class II
Units
Worldwide - US Nationwide and the countries of Brazil, Canada, Georgia, Italy,…
Category
Drugs & medical
Source
FDA · Device

Copied automatically from the US Food and Drug Administration. Details may be shortened or out of date; the official notice is the authoritative source. Disclaimer

What to do

  1. Check the product name, strength and lot number against the official notice.
  2. Do not stop a prescribed medicine without talking to your doctor or pharmacist first.
  3. Ask the pharmacy or supplier about a replacement or refund.

General guidance from Recall Atlas. The official notice from the US Food and Drug Administration has the exact models, batch codes and remedy, and it takes priority.

More Beckman Coulter recalls

  1. 2025-10-15
  2. US
    Drugs & medicalClass II

    Beckman Coulter Inc. recalls SYNCHRON Systems Phosphorus (PHOSm) Reagent REF: 467868 For In Vitro Diagnostic Use, Rx Only.

    Specific SYNCHRON Systems Modular Chemistry Phosphorus (PHOSm REF 467868) reagent lots may fail calibration before the lot expiration date which causes a delay in obtaining patient results.

    FDA · DeviceBeckman Coulter Inc.US: AL KY LA MA MD MS NC NE NJ NY OK PA TX WA WV OUS: Canada Mexico SingaporeOfficial notice →
  3. 2025-10-08
  4. US
    Drugs & medicalClass II

    Beckman Coulter Inc. recalls DxC 500 AU Module w/ISE, DxC 500i, With ISE, REF: C63522 and DxC 500 AU Module, DxC 500i, Without ISE, REF: C6352; part of the DxC 500i…

    On a clinical analyzer, when ordering a Clinical Chemistry (CC) combination test simultaneously with an Immunoassay (IA) test, if the IA test is processed between the CC tests, CC sample status will remain In Progress…

    FDA · DeviceBeckman Coulter Inc.Internationally distribution to the countries of Poland, India, Dominican…Official notice →
  5. 2025-09-24
  6. US
    Drugs & medicalClass II

    Beckman Coulter Inc. recalls UniCel DxH 800 COULTER Cellular Analysis System, configured with the Led HGB Photometer REFs:B24802, 629029, B63322

    Hematology analyzers configured with HGB photometers may result in erroneously high hemoglobin (HGB) results on samples with elevated white blood cell (WBC) counts.

    FDA · DeviceBeckman Coulter Inc.Worldwide - US Nationwide distribution including in the states of AL, AZ, CA,…Official notice →
  7. 2025-08-13
  8. US
    Drugs & medicalClass II

    Beckman Coulter, Inc. recalls DxI 800 Access Immunoassay Analyzer W/Dual Gantry, Part Number A71457

    Beckman Coulter has identified that the UniCel DxI 600/800 systems with linear slide PnP gantries running software versions 5.7 and up faced incorrect PnP X-motor current settings where the PnP X-motor current…

    FDA · DeviceBeckman Coulter, Inc.Worldwide distribution - US Nationwide.Official notice →
  9. US
    Drugs & medicalClass II

    Beckman Coulter Inc. recalls UIBC (Unsaturated Iron Binding Capacity), REF: OSR61205,

    Beckman Coulter identified that the Unsaturated Iron Binding Capacity (UIBC) assay is not meeting their labeled hemoglobin interference claims of no significant interference (i.e., less than 10%) up to 200 mg/dL…

    FDA · DeviceBeckman Coulter Inc.US: AK, AL, AR, CO, CT, DC, DE, ID, UT OUS: Algeria, Argentina, Australia,…Official notice →
  10. 2025-07-30
  11. US
    Drugs & medicalClass II

    Beckman Coulter Inc. recalls Estrone RIA, REF: DSL8700. Estrone RIA is an IVD used for the quantitative measurement of estrone in human serum and plasma.

    A specific lot of bovine serum albumin (BSA) used for manufacturing of coated tubes (CT) that are included in the Estrone RIA kit lots caused different affinity of manufactured coated tubes to patient samples, causing…

    FDA · DeviceBeckman Coulter Inc.US: North Carolina OUS: Argentina, Austria, Belgium, Canada, France, Germany,…Official notice →
All 84 Beckman Coulter recalls →

Other drugs & medical recalls in the United States

  1. 2025-08-27
  2. US
  3. 2025-08-20
  4. US
    Drugs & medicalClass III

    Diversatek Healthcare recalls Zvu Anorectal Manometry Procedure Kit, REF: HRAM-PROC-10

    Anorectal manometry procedure kit outer kit label is mislabeled with an incorrect expiration date; the manufacturing date was used in place of the expiration date. All inner labels on kit components display the correct…

    FDA · DeviceDiversatek HealthcareWorldwide - US Nationwide distribution in the states of NM, MA, CO, OH and the…Official notice →
  5. US
    Drugs & medicalClass II

    Instrumentation Laboratory recalls HemosIL LMW Heparin Controls; Part Number: 0020300200;

    Multiple complaints indicating lower than expected quality control (QC) results and in some instances QC recovery below the lower limit of the package insert acceptance range when testing is performed with HemosIL…

    FDA · DeviceInstrumentation LaboratoryWorldwide - US Nationwide and the countries of Algeria, Andorra, Argentina,…Official notice →
  6. US
    Drugs & medicalClass III

    CareFusion 303, Inc. recalls Description/REF: BD PYXIS MEDBANK MINI 1FH-1FM/169-137, BD PYXIS MEDBANK MINI 4HM/169-113, BD PYXIS MEDBANK MINI 4HH/169-112, BD PYXIS…

    Reports of delayed access to medication in automated dispensing cabinets because of the override/Add Item workflow.

    FDA · DeviceCareFusion 303, Inc.US Nationwide distribution including in the states of NY, MA, KY, VA, OR, OH,…Official notice →
  7. US
    Drugs & medicalClass II

    GE Medical Systems, LLC recalls GE Signa Artist, Nuclear Magnetic Resonance Imaging System

    GE HealthCare has become aware that for certain MR systems (SIGNA Artist, SIGNA Voyager, Discovery MR450 1.5T, Optima MR450w 1.5T, SIGNA Architect, SIGNA Premier, SIGNA UHP, Discovery MR750 3.0T, Discovery MR750w 3.0T,…

    FDA · DeviceGE Medical Systems, LLCWorldwide distribution.Official notice →
  8. US
    Drugs & medicalClass II

    Insightra Medical Inc recalls Brand Name: Ultra IABP Catheter Kit Product Name: Ultra IABP Catheter Kit Model/Catalog Number: IMU7F-40, IMU7F-35, IMU7F-30, IMU7F-25,…

    Device contains indications for use and device compatibility claims that have not be reviewed and approved for safety and effectiveness by the FDA. Instructions for use are not consistent with product training.

    FDA · DeviceInsightra Medical IncWorldwide - US Nationwide distribution in the states of TX, MO, FL, MI and the…Official notice →
See all drugs & medical recalls →