GE Medical Systems, LLC recalls GE Signa Artist, Nuclear Magnetic Resonance Imaging System
GE HealthCare has become aware that for certain MR systems (SIGNA Artist, SIGNA Voyager, Discovery MR450 1.5T, Optima MR450w 1.5T, SIGNA Architect, SIGNA Premier, SIGNA UHP, Discovery MR750 3.0T, Discovery MR750w 3.0T,…
Read the official notice →- Date
- 2025-08-20
- Country
- United States
- Brand
- GE Medical Systems
- Product
- GE Signa Artist, Nuclear Magnetic Resonance Imaging System
- Classification
- Class II
- Units
- Worldwide distribution.
- Category
- Drugs & medical
- Source
- FDA · Device
Copied automatically from the US Food and Drug Administration. Details may be shortened or out of date; the official notice is the authoritative source. Disclaimer
What to do
- Check the product name, strength and lot number against the official notice.
- Do not stop a prescribed medicine without talking to your doctor or pharmacist first.
- Ask the pharmacy or supplier about a replacement or refund.
General guidance from Recall Atlas. The official notice from the US Food and Drug Administration has the exact models, batch codes and remedy, and it takes priority.
More GE Medical Systems recalls
- 2025-10-29
US
GE Medical Systems, LLC recalls MAC VU360 Acquisition Trunk Cable and Module Holder, high performance, multichannel resting electrocardiograph, Model Number 2030360-001,…
If a user incorrectly places the Acquisition Module into the Acquisition Module Holder, it can result in excessive bending and wear of the Acquisition Trunk Cable. This is visible as a damaged sheath or exposed wires.…
FDA · DeviceGE Medical Systems, LLCWorldwide Distribution: US (nationwide) and OUS (foreign/Global) to countries…Official notice →- 2025-10-22
US
GE Medical Systems, LLC recalls LOGIQ P10 R4.5 HD ultrasound system with system software version R4.5.7., Model Number 5877535
The Ultrasound-Guided Attenuation Parameter (UGAP) measurement data may display inaccurate values representing liver steatosis. This could potentially lead to inappropriate clinical decisions impacting overall care.
- 2025-09-03
US
GE Medical Systems, LLC recalls GE HealthCare OEC 9900 Elite C-arm Systems.
GE HealthCare has become aware of a non-conformance to 21 CFR 1020.30 (k) in certain OEC 9800 and OEC 9900 Elite C-arm Systems. Specifically, impacted systems are missing the full layer of lead tape in the collimator…
- 2025-07-30
US
GE Medical Systems, LLC recalls SIGNA Architect
GE HealthCare has become aware that gradient coils for certain MR systems (see affected product list in this letter), under specific conditions, can produce elevated acoustic noise during scanning. As a result,…
FDA · DeviceGE Medical Systems, LLCWorldwide distribution - US Nationwide and the country of China.Official notice →- 2025-06-18
US
GE Medical Systems, LLC recalls OEC Elite: The OEC Elite mobile C-arm is intended to provide fluoroscopic and digital spot images of the patient anatomy, interventional…
OEC Elite and OEC 3D Mobile C-Arms Exceeding 4% Source to Image Distance in X-ray Field.
- 2025-04-23
US
GE Medical Systems, LLC recalls Venue Go Models: R2 (Model Number H45181VG), R3 (Model Number H45193VG), R4 (Model Number H45114VG), R5 (Model Numbers H45135VG,…
The battery for certain ultrasound systems can potentially develop an internal failure which could result in smoke or fire.
Other drugs & medical recalls in the United States
- 2025-08-20
US
Instrumentation Laboratory recalls HemosIL LMW Heparin Controls; Part Number: 0020300200;
Multiple complaints indicating lower than expected quality control (QC) results and in some instances QC recovery below the lower limit of the package insert acceptance range when testing is performed with HemosIL…
FDA · DeviceInstrumentation LaboratoryWorldwide - US Nationwide and the countries of Algeria, Andorra, Argentina,…Official notice →US
Beckman Coulter, Inc. recalls Access Thyroglobulin assay, a simultaneous one-step immunoenzymatic assay, Catalog Number 33860
Access Thyroglobulin reagent lot 439163 may generate erroneously high patient results. Falsely increased Access Thyroglobulin results may lead a physician to pursue unnecessary diagnostic imaging studies and/or…
FDA · DeviceBeckman Coulter, Inc.Worldwide - US Nationwide and the countries of Brazil, Canada, Georgia, Italy,…Official notice →US
CareFusion 303, Inc. recalls Description/REF: BD PYXIS MEDBANK MINI 1FH-1FM/169-137, BD PYXIS MEDBANK MINI 4HM/169-113, BD PYXIS MEDBANK MINI 4HH/169-112, BD PYXIS…
Reports of delayed access to medication in automated dispensing cabinets because of the override/Add Item workflow.
FDA · DeviceCareFusion 303, Inc.US Nationwide distribution including in the states of NY, MA, KY, VA, OR, OH,…Official notice →US
Insightra Medical Inc recalls Brand Name: Ultra IABP Catheter Kit Product Name: Ultra IABP Catheter Kit Model/Catalog Number: IMU7F-40, IMU7F-35, IMU7F-30, IMU7F-25,…
Device contains indications for use and device compatibility claims that have not be reviewed and approved for safety and effectiveness by the FDA. Instructions for use are not consistent with product training.
FDA · DeviceInsightra Medical IncWorldwide - US Nationwide distribution in the states of TX, MO, FL, MI and the…Official notice →US
PIE Medical Imaging B.V. recalls 3mensio Workstation (Vascular Fenestrated) software
When fenestrated analysis with clock measurements is started in diagnostic bioimaging software (intended to measure/visualize cardiovascular structures) and the 12h position of a single clock is changed, other clock…
FDA · DevicePIE Medical Imaging B.V.US Nationwide distribution in the states of MN, NJ, GA, NY, OH, KS, MA, CA,…Official notice →US
Merit Medical Systems, Inc. recalls Custom Manifold Kit: REF: K09-02239C, K09-02249UP, K09-09110B, K09-10294A, K09-11227A, K09-11456B, K09-12123B, K09-13342AP,…
High pressure tubing may relax over time causing it to partially separate from the connector resulting in a void that can trap air in the line. Use of the affected product may result in the trapped air being introduced…
FDA · DeviceMerit Medical Systems, Inc.US Nationwide distribution in the states of OH, CA, SD, VA, AZ, MT, TX, MI,…Official notice →