Beckman Coulter, Inc. recalls DxI 9000 Access Immunoassay Analyzer C11137
Analyzer has calibration issues where curves switched from passed to failed due to system errors during aspiration. This anomaly causes erroneous results and delays if unnoticed. RLU values may be misinterpreted as…
Read the official notice →- Date
- 2025-07-02
- Country
- United States
- Brand
- Beckman Coulter
- Product
- DxI 9000 Access Immunoassay Analyzer C11137
- Classification
- Class II
- Units
- Worldwide - US Nationwide distribution in the states of AR, CA, FL, GA, HI,…
- Category
- Drugs & medical
- Source
- FDA · Device
Copied automatically from the US Food and Drug Administration. Details may be shortened or out of date; the official notice is the authoritative source. Disclaimer
What to do
- Check the product name, strength and lot number against the official notice.
- Do not stop a prescribed medicine without talking to your doctor or pharmacist first.
- Ask the pharmacy or supplier about a replacement or refund.
General guidance from Recall Atlas. The official notice from the US Food and Drug Administration has the exact models, batch codes and remedy, and it takes priority.
More Beckman Coulter recalls
- 2025-07-30
US
Beckman Coulter Inc. recalls Estrone RIA, REF: DSL8700. Estrone RIA is an IVD used for the quantitative measurement of estrone in human serum and plasma.
A specific lot of bovine serum albumin (BSA) used for manufacturing of coated tubes (CT) that are included in the Estrone RIA kit lots caused different affinity of manufactured coated tubes to patient samples, causing…
FDA · DeviceBeckman Coulter Inc.US: North Carolina OUS: Argentina, Austria, Belgium, Canada, France, Germany,…Official notice →- 2025-07-23
US
Beckman Coulter, Inc. recalls Access Vitamin B12 Reagent, Catalog Number 33000, Lot Number 439850
Beckman Coulter has received complaints of erroneous Vitamin B12 results when using Access Vitamin B12 Reagent lot 439850. The affected reagent lot may cause erroneously low or high Vitamin B12 results, leading to a…
FDA · DeviceBeckman Coulter, Inc.International distribution to the country of Andorra, Bulgaria, Denmark,…Official notice →- 2025-07-02
US
Beckman Coulter Inc. recalls REMISOL Advance UDI-DI code: 13700962601874 All Software Versions: B92487; B92488; C24317; C28652; C37500; D04164; C44703; C57017; C69412;…
Due to a software issue, the software may incorrectly process the date of birth (DoB) received from the Laboratory Information System (LIS) and when the DoB is set to 01/01/1900 or left empty, REMISOL Advance may…
FDA · DeviceBeckman Coulter Inc.Worldwide - U.S. Nationwide distribution including in the states of AK, AL,AR,…Official notice →- 2025-06-18
US
Beckman Coulter, Inc. recalls DxI 9000 Access Immunoassay Analyzer, Catalog Number C11137, running software versions 1.20
A new motor/encoder and cable was introduced into instruments manufactured starting with serial number 300141 incorrectly. This incorrect installation is introducing error codes for the Dxl 9000 which prevents the…
FDA · DeviceBeckman Coulter, Inc.Worldwide distribution: US (Nationwide) and OUS Global (Foreign) to countries…Official notice →- 2025-06-11
US
Beckman Coulter, Inc. recalls DxI 9000 Access Immunoassay Analyzer, Catalog Number C11137, running software version 1.20
A new software update (v1.20) is introducing error codes for the Dxl 9000 which prevents the assay from being performed and a delay in results from being generated, and may require field service to resolve the delay.
FDA · DeviceBeckman Coulter, Inc.Worldwide Distribution: US (Nationwide) and OUS (Global) to countries of:…Official notice →US
Beckman Coulter, Inc. recalls Access Testosterone assay, Catalog Number 33560
Inadequate preventative maintenance (PM) performed on the scales of the RPL5 fill line, led to splashing (cross-contamination) between the wells of testosterone reagent packs lot 538081, which could potentially result…
Other drugs & medical recalls in the United States
- 2025-07-02
US
Becton Dickinson Infusion Therapy Systems, Inc. recalls REF: 381434 BD Insyte Autoguard Shielded IV Catheter, 20 GA x 1.16 in (1.1 x 30 mm) 60 mL/min, STERILEEO, RxOnly
Due to customer complaints, there is the possibility that catheter needle may be slow to retract or fail to retract.
FDA · DeviceBecton Dickinson Infusion Therapy Systems, Inc.Worldwide - US Nationwide distribution including in the states of AL, AR, AZ,…Official notice →US
Jolife AB recalls LUCAS 2, 3 and 3.1 LUCAS chest compression System is to be used for performing external cardiac compressions on adult patients who have…
Due to demonstration units of chest compression system being provided to a customer which may potentially be utilized for clinical use.
FDA · DeviceJolife ABU.S. Nationwide distribution in the states of IN, MI, OH and TX.Official notice →US
BIGFOOT BIOMEDICAL recalls Bigfoot Unity Diabetes Management System SW-300689-01 Bigfoot Inject v1 Pen Cap LCAP Firmware 1.3.0 Release Configuration, Rev H…
Due to a software/firmware issue diabetes management system may provide a device error and cause the insulin pen to become unresponsive and unable to function. This error may pose a potential health risk for people…
FDA · DeviceBIGFOOT BIOMEDICALU.S. Nationwide distribution in the states of AL, FL, GA, IA, IL, NY, PA, TX,…Official notice →US
MEDLINE INDUSTRIES, LP - Northfield recalls Medline UMBILICAL TRAY W/3.5&5FR CATH, REF UVT1250
There is a possible packaging defect impacting the outer Tyvec pouch that protects the sterile product.
FDA · DeviceMEDLINE INDUSTRIES, LP - NorthfieldUS Nationwide distribution in the state of VA.Official notice →US
Barco N.V. recalls Model: Description/REF: MNA-420: MNA-420 ENC -H/K9303320, MNA-420 DEC -H V2/K9303331; MNA-440: MNA-440 ENC -H/K9303300, MNA-440 DEC -H…
Cathode ray tube display system encoders and decoders have same MAC address; new system setup with two or more affected units can lead to systems not functioning reliably:audio, video and/or control signals might be…
US
Integra LifeSciences Corp. (NeuroSciences) recalls MicroMyst Applicator, Box of 5. Consisting of a multilumen applicator and filtered line for use with the Integra Flow…
Lack of sterility assurance.
FDA · DeviceIntegra LifeSciences Corp. (NeuroSciences)Worldwide distribution - US Nationwide and the countries of Taiwan (Province…Official notice →