Beckman Coulter, Inc. recalls Access Total T4, REF 33800, IVD, thyroxine reagent assay.
Four lots of Access Total T4 reagent lots demonstrated imprecision outside of the 10% specification which may result in erroneous or delayed results.
Read the official notice →- Date
- 2024-03-27
- Country
- United States
- Brand
- Beckman Coulter
- Product
- Access Total T4, REF 33800, IVD, thyroxine reagent assay.
- Classification
- Class II
- Units
- Worldwide distribution - US Nationwide, including Puerto Rico and the…
- Category
- Drugs & medical
- Source
- FDA · Device
Copied automatically from the US Food and Drug Administration. Details may be shortened or out of date; the official notice is the authoritative source. Disclaimer
What to do
- Check the product name, strength and lot number against the official notice.
- Do not stop a prescribed medicine without talking to your doctor or pharmacist first.
- Ask the pharmacy or supplier about a replacement or refund.
General guidance from Recall Atlas. The official notice from the US Food and Drug Administration has the exact models, batch codes and remedy, and it takes priority.
More Beckman Coulter recalls
- 2024-07-10
US
Beckman Coulter, Inc. recalls DxI 9000 Access Immunoassay Analyzer, C11137; in-vitro diagnostic
Beckman Coulter is recalling their DxI 9000 Access Immunoassay Analyzers, an in vitro diagnostic device used for the quantitative, semi-quantitative, or qualitative determination of various analyte concentrations found…
US
Beckman Coulter, Inc. recalls Beckman Coulter DxI 9000 Access Immunoassay Analyzer is a discrete photometric chemistry analyzer for clinical use. It is an in-vitro…
The reason for the recall is the pneumatic tubing used in the analytical pick-and-place (PnP) assembly cable may fully or partially disconnect from the flex cable assembly fittings during normal operation. The DxI 9000…
FDA · DeviceBeckman Coulter, Inc.Worldwide - US Nationwide distribution in the state of Indiana and the…Official notice →- 2024-04-03
US
Beckman Coulter, Inc. recalls Access Substrate, REF 81906, For use with the Access Immunoassay Systems
On 07 February 2024 Beckman Coulter made the decision to conduct a field action on Access Substrate (81906) lots: 234600, 337984, 338040, 338214, 338229, 338289, 338322, 338333, 338451, 338587, and 338538. Beckman…
FDA · DeviceBeckman Coulter, Inc.Worldwide Distribution: US (nationwide) and OUS (foreign) to countries of:…Official notice →- 2024-02-14
US
Beckman Coulter, Inc. recalls DxI 9000 Access Immunoassay Analyzer running software version 1.16 or prior,
There are potential performance issues found in the DxI 9000 Access Immunoassay Analyzer, including: 1) Access Ultrasensitive Insulin users who configure SI units (pmol/L) applies an incorrect conversion factor; 2)…
FDA · DeviceBeckman Coulter, Inc.Worldwide distribution - US Nationwide and the countries of Australia,…Official notice →US
Beckman Coulter, Inc. recalls Access Ultrasensitive Insulin only used on the DxI 9000 Access Immunoassay Analyzer, Catalog Number 33410
There is an issue with the Access Ultrasensitive Insulin assay protocol file (APF) for use with the DxI 9000 Access Immunoassay Analyzer (APF 182 test definition version 6.7 and below). The Access Ultrasensitive…
FDA · DeviceBeckman Coulter, Inc.Worldwide distribution - US Nationwide and the countries of Australia,…Official notice →- 2024-01-31
US
Beckman Coulter, Inc. recalls Beckman Coulter DxU 850m/840m Microscopy Analyzers, Part Numbers: C49513, C76947
Beckman Coulter has become aware that the cannula, which is found in the Flowcell and the iQClear Rinse Waste Well Assembly, may rust. Rust from the cannula has the potential to generate erroneous results.
FDA · DeviceBeckman Coulter, Inc.US and Australia, Brazil, Canada, China, France, Germany, Ghana, India,…Official notice →
Other drugs & medical recalls in the United States
- 2024-03-27
US
Sarnova HC, Llc recalls Curaplex Nasal Intubation Kit w/ 7mm Endotracheal Tube-To perform a nasal intubation to establish an airway in an emergency. Item Number:…
Endotracheal Tube Holder included in certain kits is not compatible with a nasal intubation procedure, leaving practitioners without the necessary means to secure the inserted endotracheal tube.
US
Natus Neurology Inc recalls TENSIVE Conductive Adhesive Gel, REF 016-401600
Natus has become aware that its third-party supplier Parker Laboratories Inc. is conducting a voluntary recall related to their Tensive Conductive Adhesive Gel, Ref. 22-60 Lot Number A0623009. They have received five…
FDA · DeviceNatus Neurology IncWorldwide distribution - US Nationwide and the country of Canada.Official notice →US
Siemens Healthcare Diagnostics, Inc. recalls DCA Systems HbA1c Reagent Kit
The reason for the recall is that the Siemens DCA HbA1c Reagent Kit was confirmed to display a positive bias without alerting the user. The potential impact of the defect is that erroneous test results could be…
FDA · DeviceSiemens Healthcare Diagnostics, Inc.Worldwide distribution - US Nationwide and the country of Taiwan.Official notice →US
Thoratec Corp. recalls HeartMate 3 Left Ventricular Assist System (LVAS): 1. HeartMate 3 Sealed Outflow Graft with Bend Relief, Investigational Device Exemption…
Observed outflow graft deformation known as Extrinsic Outflow Graft Obstruction (EOGO) associated with Left Ventricular Assist System.
FDA · DeviceThoratec Corp.Worldwide distribution: U.S.(nationwide): AL,AR, AZ, CA, CO, CT, DC, DE, FL,…Official notice →US
Smiths Medical ASD Inc. recalls smiths medical Pneupac paraPAC plus 310 ventilator kit with PEEP and CPAP, REF P310NUS
When a paraPAC plus ventilator is switched to the operating mode of VENTILATE, the ventilator may intermittently provide continuous positive gas flow instead of the intended cycling like a human breath. This…
FDA · DeviceSmiths Medical ASD Inc.Worldwide distribution: US (nationwide) and OUS countries of: Algeria,…Official notice →US
ARROW INTERNATIONAL Inc. recalls ARROW QuickFlash Radial Artery Catherization (wire, guide, catheter), Part Numbers: a) REF AK-04220; b) REF ASK-04220-FMH1; c) REF…
Teleflex is initiating this voluntary recall for the above-mentioned products due to reports received indicating a potentially defective component. The complaints received relate to resistance of the guidewire…
FDA · DeviceARROW INTERNATIONAL Inc.Worldwide distribution: US (nationwide) and OUS (foreign) to countries of: AE,…Official notice →