Olympus Corporation of the Americas recalls Single Use Repositionable Clip QUICKCLIP PRO 165 CM 10/BX- Intended for hemostatic metal clip for the GI Tract Catalog…
Complaints on clip deployment occurring during clinical procedures, leading to hazardous situations where clips fail to achieve their expected function or deploy incorrectly, potentially causing harms to patients. The…
Read the official notice →- Date
- 2023-06-28
- Country
- United States
- Brand
- Olympus Corporation of the Americas
- Product
- Single Use Repositionable Clip QUICKCLIP PRO 165 CM 10/BX- Intended for hemostatic metal clip for the GI Tract Catalog Number: HX-202LR.A
- Classification
- Class II
- Units
- US Nationwide distribution.
- Category
- Drugs & medical
- Source
- FDA · Device
Copied automatically from the US Food and Drug Administration. Details may be shortened or out of date; the official notice is the authoritative source. Disclaimer
What to do
- Check the product name, strength and lot number against the official notice.
- Do not stop a prescribed medicine without talking to your doctor or pharmacist first.
- Ask the pharmacy or supplier about a replacement or refund.
General guidance from Recall Atlas. The official notice from the US Food and Drug Administration has the exact models, batch codes and remedy, and it takes priority.
More Olympus Corporation of the Americas recalls
- 2023-08-16
US
Olympus Corporation of the Americas recalls Tracheal Intubation Fiberscope OLYMPUS LF-P- To be used with an Olympus Light Source, documentation equipment, display…
Revised OER-Pro, OER-Elite, OER-Mini endoscope reprocessor labeling- the LF-V and LF-P are no longer compatible endoscopes for reprocessing in the OERs. The addendum highlights the following changes: 1. Removal of LF-V…
- 2023-08-09
US
Olympus Corporation of the Americas recalls Always-On Tip Tracked Instruments (SPiN Drive instruments) Triple Needle Brush, 12mm L, 1.8mm OD-Intended for use in clinical…
Electromagnetic (EM) sensor tracking malfunction impacts the use of navigation functionality during endobronchial procedures and prompts the system to report a coil break warning message. This could result in a…
FDA · DeviceOlympus Corporation of the AmericasWorldwide Distribution: US (Nationwide) and OUS (foreign) to countries of: CN,…Official notice →- 2023-07-26
US
Olympus Corporation of the Americas recalls Olympus BRONCHOFIBERSCOPE BF Type, Model Numbers BF-PE2, BF-TE2
There have been complaints of endobronchial combustion during therapeutic laser procedures with the bronchoscope.
- 2023-05-31
US
Olympus Corporation of the Americas recalls ViziShot 2 Single Use Aspiration Needle, 21G- To be used with ultrasound endoscopes for ultrasound guided fine needle…
Product non-conformance pertaining to the angle of the needle does not meet manufacturing specifications and remained unbent, resulting to slight resistance experienced during needle advancement and retraction.
US
Olympus Corporation of the Americas recalls Olympus Single Use Distal Cover Model Number MAJ-2315 used with EVIS EXERA III Duodenovideoscope TJF-Q190V; Item code:…
The Single Use Distal Cover MAJ-2315 may cause mucosal injury and/or become detached from the endoscope during use. Consequences of a detached cover could include the risk of aspiration, inhalation, or obstruction that…
- 2023-04-26
US
Olympus Corporation of the Americas recalls Single Use Rotatable Clip Fixing Device Model HX 201LR 135 - Rotatable, 165cm long, 2.8 scope channel, standard clip Model…
There have been complaints that the clip did not come out of the tube sheath during the procedure.
Other drugs & medical recalls in the United States
- 2023-06-28
US
Heartware, Inc. recalls HeartWare Ventricular Assist Device System. a) Monitor 1521US Explore Tech, b) Monitor 1521GB Explore Tech, c) Monitor 1521IL Explore…
Medtronic has identified that the logfiles downloaded from the recently updated Model 1521 monitors (Serial Numbers: MON4xxxxxx for OUS and MON5xxxxxx for US), are unable to be processed by the Autologs web portal.
FDA · DeviceHeartware, Inc.Worldwide - US Nationwide distribution in the states of TX, MO, VA, and the…Official notice →US
Draeger Medical Systems, Inc. recalls Draeger Softbed Resuscitaire and/or Birthing Room Warmer with Scale [Model(s): RW82VHA-1, RW82VHA-1C, RW82VHA-1D, RW82-1, RW82-1C,…
The Resuscitaire Infant Radiant Warmer with the optional scale could display inaccurate weight values.
FDA · DeviceDraeger Medical Systems, Inc.US Nationwide distribution including in the states of AL, AR, AZ, CA, DE, FL,…Official notice →US
Karl Storz Endoscopy recalls Metal Tray, 178 x 135 x 37 mm, REF: 11580A; Metal Tray, 275 x 178 x 35 mm, REF: 11580B; Metal Tray, 373 x 178 x 35 mm, REF: 11580C; Metal…
Metal sterilization trays used to encase and protect medical devices during sterilization have a new 510(k) with instructions for use that is for the first time specific to the tray models, that also includes…
FDA · DeviceKarl Storz EndoscopyWorldwide - US Nationwide distribution including in the states of IN, MS, MN,…Official notice →US
Masimo Corporation recalls Masimo LNCS TC-I SpO2, Reusable Ear Sensor, REF: 2503
TC-I tip clip sensor may produce readings outside the accuracy specification.
FDA · DeviceMasimo CorporationUS: AK , AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY,…Official notice →US
MEDLINE INDUSTRIES, LP - Northfield recalls MEDLINE THERAHONEY GEL, 100% LEPTOSPERMUM SCOPARIUM HONEY, a) REF MNK0005, b) REF MNK0015 wound gel
Medline issued a recall on TheraHoney Wound Gel due to the manufacturer notifying Medline of a sterilization dose audit failure on the wound gel and not augmenting the dose after failure per ISO 11137.
FDA · DeviceMEDLINE INDUSTRIES, LP - NorthfieldWorldwide distribution - US Nationwide and the countries of SINGAPORE,…Official notice →US
Karl Storz Endoscopy recalls KARL STORZ Radel Sterilization Trays, Models: 27717A, 27717B, 39301A, 39301C, 39231XA, 39301C1, 39311A, 39317A
Plastic sterilization trays used to encase and protect medical devices during sterilization have a new 510(k) with instructions for use that is for the first time specific to the tray models, that also includes…
FDA · DeviceKarl Storz EndoscopyWorldwide - US Nationwide distribution including in the states of AK, AR, AZ,…Official notice →