Drugs & medicalClass II

Olympus Corporation of the Americas recalls Olympus Triangle Tip Electrosurgical Knives- to be used with Olympus endoscopes and electrosurgical units to cut tissue using…

Due to increase in complaints for the KD-640L and KD-645L Triangle Tip Electrosurgical Knives breaking off during use -issuing reminder to users to utilize the electrosurgical knife in accordance with the Instructions…

Read the official notice →
Date
2024-09-18
Country
United States
Brand
Olympus Corporation of the Americas
Product
Olympus Triangle Tip Electrosurgical Knives- to be used with Olympus endoscopes and electrosurgical units to cut tissue using…
Classification
Class II
Units
Nationwide
Category
Drugs & medical
Source
FDA · Device

Copied automatically from the US Food and Drug Administration. Details may be shortened or out of date; the official notice is the authoritative source. Disclaimer

What to do

  1. Check the product name, strength and lot number against the official notice.
  2. Do not stop a prescribed medicine without talking to your doctor or pharmacist first.
  3. Ask the pharmacy or supplier about a replacement or refund.

General guidance from Recall Atlas. The official notice from the US Food and Drug Administration has the exact models, batch codes and remedy, and it takes priority.

More Olympus Corporation of the Americas recalls

  1. 2024-10-30
  2. US
    Drugs & medicalClass II

    Olympus Corporation of the Americas recalls Olympus Gastrointestinal Videoscope, Model/Catalog Number: GIF-1100

    It was discovered during device performance testing that the CCD imaging sensors were programmed with the incorrect color correction data and therefore, specifications are not met.

    FDA · DeviceOlympus Corporation of the AmericasUS Nationwide distribution.Official notice →
  3. US
    Drugs & medicalClass II

    Olympus Corporation of the Americas recalls Soltive Premium SuperPulsed Laser - Intended for incision, excision, resection, ablation, coagulation, hemostasis, and…

    The term Bladder Stone was incorrectly translated in both Spanish and Portuguese to Kidney Stone(Clculo renal) on the systems Graphical User Interface (GUI), a potential to deliver an incorrect or unintended amount of…

    FDA · DeviceOlympus Corporation of the AmericasWorldwide distribution - US Nationwide and the countries of AU, CA, CL, DE,…Official notice →
  4. 2024-10-23
  5. US
    Drugs & medicalClass II

    Olympus Corporation of the Americas recalls EVIS EXERA III Video System Center Model Number CV-190. Intended to be used with OLYMPUS camera heads, endoscopes, light…

    Certain serial numbers of the CV-190 do not start up properly because parts that deviated from the specification were assembled into the power supply unit.

    FDA · DeviceOlympus Corporation of the AmericasUS Nationwide distribution.Official notice →
  6. 2024-08-14
  7. US
    Drugs & medicalClass II

    Olympus Corporation of the Americas recalls The Everest Bipolar Cutting Forceps are intended to be passed through a 5-mm or 10-mm cannula, depending on the model/outer…

    Fractures and breakages in packaging trays and Tyvek covers, which may result in a sterility breach

    FDA · DeviceOlympus Corporation of the AmericasWorldwide distribution - US Nationwide and the countries of CA, DE, AU, SG, KR.Official notice →
  8. US
    Drugs & medicalClass II

    Olympus Corporation of the Americas recalls Soltive Premium SuperPulsed Laser System, Model TFL-PLS

    Difficulties in pairing the wireless footswitch with the Soltive Laser, potentially delaying the surgical procedure.

    FDA · DeviceOlympus Corporation of the AmericasWorldwide distribution - US Nationwide and the countries of Australia, Canada,…Official notice →
  9. 2024-07-31
  10. US
    Drugs & medicalClass II

    Olympus Corporation of the Americas recalls ESG-410 Electrosurgical Generator- In conjunction with electrosurgical accessories intended for cutting and coagulation of…

    Electrosurgical Generators ESG-410 does not power on upon use, may result in delays in initiating treatment or prolonged surgery

    FDA · DeviceOlympus Corporation of the AmericasUS Nationwide distribution.Official notice →
All 114 Olympus Corporation of the Americas recalls →

Other drugs & medical recalls in the United States

  1. 2024-09-18
  2. US
    Drugs & medicalClass II

    PACIFIC LASERTECH, LLC recalls 25-LHP-928, HeNe Laser System, 25 mW

    These laser products fail to comply with 21 CFR 1040.10(f)(4) Key control which requires, Each laser system classified as a Class IIIb or IV laser product shall incorporate a key-actuated master control. The key shall…

    FDA · DevicePACIFIC LASERTECH, LLCUS Nationwide DistributionOfficial notice →
  3. US
    Drugs & medicalClass I

    Smiths Medical ASD Inc. recalls Pneupac paraPAC plus 310 ventilator kit with PEEP and CPAP, REF P310NXX

    There is a potential for inadvertent tidal volume knob movement from the original setting when set at high (1000-1500 mL) and low (70-150 mL) settings.

    FDA · DeviceSmiths Medical ASD Inc.WorldwideOfficial notice →
  4. US
    Drugs & medicalClass I

    B Braun Medical Inc recalls Infusomat Space Infusion System/ Large Volume Pump, Product Code 8713050U

    Extension of previous recall RES 92978. Pumps have the potential for upstream occlusion sensor to cease proper function due to buildup of electrostatic charge during operation triggering false downstream and upstream…

    FDA · DeviceB Braun Medical IncUS distribution to AL, CA, GA, IL, MI, and TX.Official notice →
  5. US
    Drugs & medicalClass III

    Elevate Oral Care recalls Just Right 5000, 1.1% Sodium Fluoride, Candy Apple Flavor, Net Wt. 3.4oz (97g), Rx Only, Manufactured for Elevate Oral Care, LLC in the…

    LABELING: LABEL MIX-UP.

    FDA · DrugElevate Oral Care120 direct customers, Nationwide US. No foreign, DOD/VA.Official notice →
  6. US
    Drugs & medicalClass II

    PREQUEL SKIN recalls BARRIER THERAPY SKIN PROTECTANT CREAM (1% colloidal oatmeal), 10 FL OZ/296 ML- tube, Distributed By: Prequel, LOS ANGELES, CA 90069. UPC 8…

    Microbial contamination of non-sterile Products -

    FDA · DrugPREQUEL SKINDistributed nationwide in USA and on lineOfficial notice →
  7. 2024-09-11
  8. US
    Drugs & medicalClass II

    Beckman Coulter Biomedical GmbH recalls PE Centrifuge Temperature Controlled, REF: B36366, a component of the Power Express.

    The adhesive joint between the plastic socket and metal component of the Power Express Sample Holder can fail, causing the metal part to detach and enter the centrifuge. This detachment can result in damage to the…

    FDA · DeviceBeckman Coulter Biomedical GmbHWorldwide distribution - US Nationwide and the countries of Australia,…Official notice →
See all drugs & medical recalls →