Olympus Corporation of the Americas recalls Olympus Gastrointestinal Videoscope, Model/Catalog Number: GIF-1100
It was discovered during device performance testing that the CCD imaging sensors were programmed with the incorrect color correction data and therefore, specifications are not met.
Read the official notice →- Date
- 2024-10-30
- Country
- United States
- Brand
- Olympus Corporation of the Americas
- Product
- Olympus Gastrointestinal Videoscope, Model/Catalog Number: GIF-1100
- Classification
- Class II
- Units
- US Nationwide distribution.
- Category
- Drugs & medical
- Source
- FDA · Device
Copied automatically from the US Food and Drug Administration. Details may be shortened or out of date; the official notice is the authoritative source. Disclaimer
What to do
- Check the product name, strength and lot number against the official notice.
- Do not stop a prescribed medicine without talking to your doctor or pharmacist first.
- Ask the pharmacy or supplier about a replacement or refund.
General guidance from Recall Atlas. The official notice from the US Food and Drug Administration has the exact models, batch codes and remedy, and it takes priority.
More Olympus Corporation of the Americas recalls
- 2024-11-06
US
Olympus Corporation of the Americas recalls Quest Spectrum 3 Fluorescent Imaging System Model/Catalog Number: ART-SYS-0029, ART-SYS-0030 The Quest Spectrum3 is a…
Does not meet the requirements for electromagnetic compatibility (EMC) of the standard IEC 60601-1-2. Electromagnetic interference occurring during a procedure may cause the device to malfunction requiring the…
US
Olympus Corporation of the Americas recalls Olympus Endoscopic Flushing Pump, Model/Catalog Number: OFP-2 (K10001141)
During product testing, Olympus identified an intermittent loss of function of the OFP-2 Flushing Pump caused by an internal component connection failure.
- 2024-10-30
US
Olympus Corporation of the Americas recalls EZDilate Wire Guided Balloon- Indicated for use in adult and adolescent (> 12 years) populations to endoscopically dilate…
Mislabeled Glo Cath Label (attached to the device) states: 11 mm, 12 mm, 13 mm however the balloon is 8.5mm, 9.5mm and 10.5 mm and may result in prolonged surgery. All other labeling is correct for the balloon,…
FDA · DeviceOlympus Corporation of the AmericasWorldwide Distribution: US distribution to states of: CA, FL, IL, NJ, NY, PA,…Official notice →US
Olympus Corporation of the Americas recalls Soltive Premium SuperPulsed Laser - Intended for incision, excision, resection, ablation, coagulation, hemostasis, and…
The term Bladder Stone was incorrectly translated in both Spanish and Portuguese to Kidney Stone(Clculo renal) on the systems Graphical User Interface (GUI), a potential to deliver an incorrect or unintended amount of…
FDA · DeviceOlympus Corporation of the AmericasWorldwide distribution - US Nationwide and the countries of AU, CA, CL, DE,…Official notice →- 2024-10-23
US
Olympus Corporation of the Americas recalls EVIS EXERA III Video System Center Model Number CV-190. Intended to be used with OLYMPUS camera heads, endoscopes, light…
Certain serial numbers of the CV-190 do not start up properly because parts that deviated from the specification were assembled into the power supply unit.
- 2024-09-18
US
Olympus Corporation of the Americas recalls Olympus Triangle Tip Electrosurgical Knives- to be used with Olympus endoscopes and electrosurgical units to cut tissue using…
Due to increase in complaints for the KD-640L and KD-645L Triangle Tip Electrosurgical Knives breaking off during use -issuing reminder to users to utilize the electrosurgical knife in accordance with the Instructions…
Other drugs & medical recalls in the United States
- 2024-10-30
US
Diasol, Inc recalls Diasol Acid Concentrate REF G100325-10Dex 100
Due to labeling issue. The box label is different that the gallon label in that gallon label shows a lower Calcium concentration than the box label.
US
Medtronic Neuromodulation recalls Medtronic Percept RC Implantable Neurostimulators (INS), Model B35300.
A limited number of Percept RC Implantable Neurostimulators (INS), Model B35300 will not be able to communicate with the A610 DBS Clinician Programmer Application upon use and an "Invalid Device" message will appear.
FDA · DeviceMedtronic NeuromodulationInternational distribution to the countries of Germany, UK, and Israel.Official notice →US
Smiths Medical ASD, Inc. recalls medex Arterial Kit 1/EA, Product Code REF MX7827
Due to a manufacturing defect, users may experience inaccurate pressure monitoring, an interruption in pressure monitoring, or an inability to zero the device due to either a leak within the TranStar Disposable…
US
MEDLINE INDUSTRIES, LP - Northfield recalls Medline brand medical procedure convenience kits, labeled as: a) LITHOTOMY PACK, REF DYNJ24362C; b) MAJOR GYNE LAPAROSCOPY…
It has been determined that certain Convenience kits that were packaged in breather pouch lot 323080002 may result in the potential of using non-sterile product if the vendor seal of the pouch is opened/unsealed.
FDA · DeviceMEDLINE INDUSTRIES, LP - NorthfieldWorldwide distribution - US Nationwide and the countries of Canada, Panama.Official notice →US
MEDLINE INDUSTRIES, LP - Northfield recalls Medline brand medical procedure convenience kits, labeled as: 1) THORACENTESIS TRAY, REF 00-400616J; 2) HH PACK SONICATION…
XXX
FDA · DeviceMEDLINE INDUSTRIES, LP - NorthfieldWorldwide distribution - US Nationwide and the countries of Canada, Panama.Official notice →US
Smiths Medical ASD Inc. recalls smiths medical portex, 'ORATOR' SPEAKING VALVE FOR TRACHEOSTOMY TUBE, REF 100/550/000
Uncertainty in the seal integrity of the sterile packaging.