Smiths Medical ASD, Inc. recalls medex Arterial Kit 1/EA, Product Code REF MX7827
Due to a manufacturing defect, users may experience inaccurate pressure monitoring, an interruption in pressure monitoring, or an inability to zero the device due to either a leak within the TranStar Disposable…
Read the official notice →- Date
- 2024-10-30
- Country
- United States
- Brand
- Smiths Medical ASD
- Product
- medex Arterial Kit 1/EA, Product Code REF MX7827
- Classification
- Class II
- Units
- Worldwide distribution.
- Category
- Drugs & medical
- Source
- FDA · Device
Copied automatically from the US Food and Drug Administration. Details may be shortened or out of date; the official notice is the authoritative source. Disclaimer
What to do
- Check the product name, strength and lot number against the official notice.
- Do not stop a prescribed medicine without talking to your doctor or pharmacist first.
- Ask the pharmacy or supplier about a replacement or refund.
General guidance from Recall Atlas. The official notice from the US Food and Drug Administration has the exact models, batch codes and remedy, and it takes priority.
More Smiths Medical ASD recalls
- 2024-12-04
US
Smiths Medical ASD, Inc. recalls smiths medical PORTEX Tracheal Tube Exchange Guide, 15Ch x 70cm, REF 14-504-68 and REF 14-504-68JP
Smiths Medical has identified a potential for ingress of fluid into the device during reprocessing. The ingress takes place at the rear of the device between the end cap and the tube. This could lead to staining of the…
- 2024-11-27
US
Smiths Medical ASD, Inc. recalls smiths medical CADD-Solis Li-ion Rechargeable Battery Packs AC adapter, Model Number REF 21-0270-25, for use with the CADD-Solis…
The adapter input plug can become damaged or broken. If the input plug is damaged, the metal contacts to the body of the AC adapter may be exposed or one or more of the AC mains prongs may separate from the input plug.
- 2024-11-06
US
Smiths Medical ASD Inc. recalls CADD-Solis Li-ion Rechargeable Battery Packs, REF 21-2160-XX, which are also sold as a component of the following CADD Solis Ambulatory…
Damage to the battery pack may have caused a short to a capacitor within the battery pack. While the battery encasement is designed to be flame retardant, a short to the capacitor could lead to melting of the battery…
- 2024-10-30
US
Smiths Medical ASD Inc. recalls smiths medical portex, 'ORATOR' SPEAKING VALVE FOR TRACHEOSTOMY TUBE, REF 100/550/000
Uncertainty in the seal integrity of the sterile packaging.
US
Smiths Medical ASD Inc. recalls CADD Medication Cassette Reservoir, 100mL, FS, Product Code 21-7302-24JP
Certain CADD Medication Cassette Reservoirs may exhibit a weakened weld joint between the medication bag and tubing due to a production equipment malfunction. This could result in a medication leakage.
- 2024-10-02
US
Smiths Medical ASD Inc. recalls paraPAC plus 310 ventilator kit with internal PEEP and CPAP, Item Number P310N
Smiths Medical became aware of an issue related to a potential for the patient outlet connector to loosen/detach from the paraPAC Plus P300 and P310 ventilators impacting the active ventilation function.
Other drugs & medical recalls in the United States
- 2024-10-30
US
Blue Ortho recalls Predict+ is a data-driven, clinical decision support tool that uses machine learning to provide predictions of individual patient outcomes…
Predict+ is being recalled for lack of pre-market clearance/approval.
US
Diasol, Inc recalls Diasol Acid Concentrate REF G100325-10Dex 100
Due to labeling issue. The box label is different that the gallon label in that gallon label shows a lower Calcium concentration than the box label.
US
Medtronic Neuromodulation recalls Medtronic Percept RC Implantable Neurostimulators (INS), Model B35300.
A limited number of Percept RC Implantable Neurostimulators (INS), Model B35300 will not be able to communicate with the A610 DBS Clinician Programmer Application upon use and an "Invalid Device" message will appear.
FDA · DeviceMedtronic NeuromodulationInternational distribution to the countries of Germany, UK, and Israel.Official notice →US
Olympus Corporation of the Americas recalls Olympus Gastrointestinal Videoscope, Model/Catalog Number: GIF-1100
It was discovered during device performance testing that the CCD imaging sensors were programmed with the incorrect color correction data and therefore, specifications are not met.
US
MEDLINE INDUSTRIES, LP - Northfield recalls Medline brand medical procedure convenience kits, labeled as: a) LITHOTOMY PACK, REF DYNJ24362C; b) MAJOR GYNE LAPAROSCOPY…
It has been determined that certain Convenience kits that were packaged in breather pouch lot 323080002 may result in the potential of using non-sterile product if the vendor seal of the pouch is opened/unsealed.
FDA · DeviceMEDLINE INDUSTRIES, LP - NorthfieldWorldwide distribution - US Nationwide and the countries of Canada, Panama.Official notice →US
MEDLINE INDUSTRIES, LP - Northfield recalls Medline brand medical procedure convenience kits, labeled as: 1) THORACENTESIS TRAY, REF 00-400616J; 2) HH PACK SONICATION…
XXX
FDA · DeviceMEDLINE INDUSTRIES, LP - NorthfieldWorldwide distribution - US Nationwide and the countries of Canada, Panama.Official notice →