Drugs & medicalClass II

Olympus Corporation of the Americas recalls Olympus Endoscopic Flushing Pump, Model/Catalog Number: OFP-2 (K10001141)

During product testing, Olympus identified an intermittent loss of function of the OFP-2 Flushing Pump caused by an internal component connection failure.

Read the official notice →
Date
2024-11-06
Country
United States
Brand
Olympus Corporation of the Americas
Product
Olympus Endoscopic Flushing Pump, Model/Catalog Number: OFP-2 (K10001141)
Classification
Class II
Units
US Nationwide distribution.
Category
Drugs & medical
Source
FDA · Device

Copied automatically from the US Food and Drug Administration. Details may be shortened or out of date; the official notice is the authoritative source. Disclaimer

What to do

  1. Check the product name, strength and lot number against the official notice.
  2. Do not stop a prescribed medicine without talking to your doctor or pharmacist first.
  3. Ask the pharmacy or supplier about a replacement or refund.

General guidance from Recall Atlas. The official notice from the US Food and Drug Administration has the exact models, batch codes and remedy, and it takes priority.

More Olympus Corporation of the Americas recalls

  1. 2024-11-20
  2. US
    Drugs & medicalClass II

    Olympus Corporation of the Americas recalls Brand Name: Olympus Product Name: Cystoscope Outer Sheath Model/Catalog Number: WA22810A Software Version: N/A Product…

    Olympus is removing the statement of compatibility with a GreenLight Laser for BPH therapy from the Instruction for Use (IFU) due to the potential for a damaged tip during use of a laser probe.

    FDA · DeviceOlympus Corporation of the AmericasUS Nationwide distribution including in the states of AK, AL, AR, AZ, CA, CO,…Official notice →
  3. 2024-11-06
  4. US
    Drugs & medicalClass II

    Olympus Corporation of the Americas recalls Olympus Electrosurgical Generator, Model Number ESG-410 (SEP) Version WA91327U

    There was an increase in complaints indicating an occurrence of system error E1226 "Could not write to Audio Buffer" during procedures using the ESG-410 (SEP).

    FDA · DeviceOlympus Corporation of the AmericasUS Nationwide.Official notice →
  5. US
    Drugs & medicalClass II

    Olympus Corporation of the Americas recalls Quest Spectrum 3 Fluorescent Imaging System Model/Catalog Number: ART-SYS-0029, ART-SYS-0030 The Quest Spectrum3 is a…

    Does not meet the requirements for electromagnetic compatibility (EMC) of the standard IEC 60601-1-2. Electromagnetic interference occurring during a procedure may cause the device to malfunction requiring the…

    FDA · DeviceOlympus Corporation of the AmericasOHOfficial notice →
  6. 2024-10-30
  7. US
    Drugs & medicalClass II

    Olympus Corporation of the Americas recalls EZDilate Wire Guided Balloon- Indicated for use in adult and adolescent (> 12 years) populations to endoscopically dilate…

    Mislabeled Glo Cath Label (attached to the device) states: 11 mm, 12 mm, 13 mm however the balloon is 8.5mm, 9.5mm and 10.5 mm and may result in prolonged surgery. All other labeling is correct for the balloon,…

    FDA · DeviceOlympus Corporation of the AmericasWorldwide Distribution: US distribution to states of: CA, FL, IL, NJ, NY, PA,…Official notice →
  8. US
    Drugs & medicalClass II

    Olympus Corporation of the Americas recalls Olympus Gastrointestinal Videoscope, Model/Catalog Number: GIF-1100

    It was discovered during device performance testing that the CCD imaging sensors were programmed with the incorrect color correction data and therefore, specifications are not met.

    FDA · DeviceOlympus Corporation of the AmericasUS Nationwide distribution.Official notice →
  9. US
    Drugs & medicalClass II

    Olympus Corporation of the Americas recalls Soltive Premium SuperPulsed Laser - Intended for incision, excision, resection, ablation, coagulation, hemostasis, and…

    The term Bladder Stone was incorrectly translated in both Spanish and Portuguese to Kidney Stone(Clculo renal) on the systems Graphical User Interface (GUI), a potential to deliver an incorrect or unintended amount of…

    FDA · DeviceOlympus Corporation of the AmericasWorldwide distribution - US Nationwide and the countries of AU, CA, CL, DE,…Official notice →
All 114 Olympus Corporation of the Americas recalls →

Other drugs & medical recalls in the United States

  1. 2024-11-06
  2. US
    Drugs & medicalClass II

    Olympus Corporation of the Americas recalls Quest Spectrum 3 Fluorescent Imaging System Model/Catalog Number: ART-SYS-0029, ART-SYS-0030 The Quest Spectrum3 is a…

    Does not meet the requirements for electromagnetic compatibility (EMC) of the standard IEC 60601-1-2. Electromagnetic interference occurring during a procedure may cause the device to malfunction requiring the…

    FDA · DeviceOlympus Corporation of the AmericasOHOfficial notice →
  3. US
    Drugs & medicalClass II

    ImaCor Inc. recalls ClariTEE Probe- Miniature transesophageal echo ultrasound probe for use with ImaCor ultrasound diagnostic platforms. Model/Catalog Number:…

    Epoxy used to seal the edge of the transducer into its paddle exceeded specification. This variability could attenuate the ultrasound signal with a potential impact on the quality of images obtained from the probes.

    FDA · DeviceImaCor Inc.FL ,GA, NJ, TNOfficial notice →
  4. US
    Drugs & medicalClass II

    Physio-Control, Inc. recalls LIFEPAK 15 (LP15) Model/Catalog Numbers: o 97577-000101 o 99577-000046 o 99577-000055 o 99577-000061 o 99577-000076 o 99577-001250 o…

    Due to an out of tolerance tool being used on monitor/defibrillator systems.

    FDA · DevicePhysio-Control, Inc.U.S.: CT, GA, MA, NY, OH, WA, and WV. O.U.S.: N/AOfficial notice →
  5. US
    Drugs & medicalClass II

    Baxter Healthcare Corporation recalls a. Hillrom Centrella Smart+ Medical-Surgical Hospital Bed, Product codes: P7900B000001, P7900B100010, P7900B100024, P7900B100208,…

    The beds have a potential for an interface issue with the Rauland Responder 5 Nurse Call System where the bed exit system will alert at the bed, but may fail to send a remote alert through the nurse call system.

    FDA · DeviceBaxter Healthcare CorporationWorldwide distribution - US Nationwide. There was government and military…Official notice →
  6. US
    Drugs & medicalClass I

    Smiths Medical ASD Inc. recalls CADD-Solis Li-ion Rechargeable Battery Packs, REF 21-2160-XX, which are also sold as a component of the following CADD Solis Ambulatory…

    Damage to the battery pack may have caused a short to a capacitor within the battery pack. While the battery encasement is designed to be flame retardant, a short to the capacitor could lead to melting of the battery…

    FDA · DeviceSmiths Medical ASD Inc.Worldwide DistributionOfficial notice →
  7. US
    Drugs & medicalClass I

    Percussionaire Corporation recalls Phasitron5 Breathing Circuit (P5-10, P5-HC, P5-UC-10). component of Intrapulmonary Percussive Ventilation, used to provide airway…

    As a result of product complaints, breathing circuits are experiencing discolored fluid which are formed when therapy was being given with specific medication through the nebulizer cup. Discoloration is caused by a…

    FDA · DevicePercussionaire CorporationU.S.: PA, AR, MI, CA, TX, OH, NY, FL, IL, NE, OR, GA, WI, MO, OK, AL, WV, MT,…Official notice →
See all drugs & medical recalls →