Drugs & medicalClass II

Olympus Corporation of the Americas recalls THUNDERBEAT, 5MM, 35CM, Front-actuated Grip Type S. Model: TB-0535FCS Indicated for open, laparoscopic (including…

Thunderbeats blue seal button may remain in the engaged position after the button is released and not immediately return to a neutral position may result in prolonged surgery

Read the official notice →
Date
2023-10-18
Country
United States
Brand
Olympus Corporation of the Americas
Product
THUNDERBEAT, 5MM, 35CM, Front-actuated Grip Type S. Model: TB-0535FCS Indicated for open, laparoscopic (including single-site surgery)…
Classification
Class II
Units
US Nationwide distribution.
Category
Drugs & medical
Source
FDA · Device

Copied automatically from the US Food and Drug Administration. Details may be shortened or out of date; the official notice is the authoritative source. Disclaimer

What to do

  1. Check the product name, strength and lot number against the official notice.
  2. Do not stop a prescribed medicine without talking to your doctor or pharmacist first.
  3. Ask the pharmacy or supplier about a replacement or refund.

General guidance from Recall Atlas. The official notice from the US Food and Drug Administration has the exact models, batch codes and remedy, and it takes priority.

More Olympus Corporation of the Americas recalls

  1. 2023-11-15
  2. US
    Drugs & medicalClass II

    Olympus Corporation of the Americas recalls EVIS EXERA III Gastrointestinal Videoscope

    Sterilization failures when devices are sterilized per the Reprocessing Manual due to lack of drying time for the endoscope channel.

    FDA · DeviceOlympus Corporation of the AmericasNationwide Distribution.Official notice →
  3. US
    Drugs & medicalClass I

    Olympus Corporation of the Americas recalls EVIS EXERA III BRONCHOVIDEOSCOPE OLYMPUS Model Number BF-P190

    There is a risk of endobronchial combustion if high-frequency cauterization is performed while supplying oxygen [and/or] the electrode section of the electrosurgical accessory is too close to the distal end of the…

    FDA · DeviceOlympus Corporation of the AmericasUS Nationwide distributionOfficial notice →
  4. 2023-10-25
  5. US
    Drugs & medicalClass I

    Olympus Corporation of the Americas recalls Olympus High-Flow Insufflation Unit, Model UHI-4. For laparoscopic surgery.

    There have been reports of patients suffering arrhythmias, reported as short cardiac arrests, during surgical procedures where UHI-4s were used. These events may have been due to an over insufflation of the abdominal…

    FDA · DeviceOlympus Corporation of the AmericasUS Nationwide distribution.Official notice →
  6. 2023-10-11
  7. US
    Drugs & medicalClass II

    Olympus Corporation of the Americas recalls Veran : Endobronchial Always on Track Models: INS-0392 AOTT 21ga Needle, 12mm L, 1.8mm OD, 5/BX INS-0382 AOTT 19ga Needle,…

    Non-conformances in Veran devices (ig4 Image Guided System, SPiN Thoracic Navigation System, SPiN Planning Laptop Workstation, and all associated navigation instruments) manufacturing to design requirements and…

    FDA · DeviceOlympus Corporation of the AmericasWorldwide distribution - US Nationwide and the countries of CN, DE, HK, IT,…Official notice →
  8. 2023-09-06
  9. US
    Drugs & medicalClass II

    Olympus Corporation of the Americas recalls Olympus Bronchovideoscope, Models BF-3C40 & BF-N20.

    Complaint of endoscope model becoming lodged in the endotracheal tube connector due to diameter being too large.

    FDA · DeviceOlympus Corporation of the AmericasUS Nationwide distribution.Official notice →
  10. US
    Drugs & medicalClass II

    Olympus Corporation of the Americas recalls Visera Hysterovideoscope Olympus HYF Type V

    IFU labeling is being updated to revise EtO gas mixture and elimination of reusable cleaning brushes.

    FDA · DeviceOlympus Corporation of the AmericasUS Nationwide distribution including in the states of AL, AZ, CA, CO, CT, FL,…Official notice →
All 114 Olympus Corporation of the Americas recalls →

Other drugs & medical recalls in the United States

  1. 2023-10-25
  2. US
    Drugs & medicalClass III

    SUN PHARMACEUTICAL INDUSTRIES INC recalls Bupropion Hydrochloride Extended-Release Tablets USP (SR), 150 mg, 60 tablets bottle, Rx Only, Distributed by: Sun…

    Failed Dissolution Specifications

    FDA · DrugSUN PHARMACEUTICAL INDUSTRIES INCNationwide in the USA.Official notice →
  3. US
    Drugs & medicalClass III

    Imprimis NJOF, LLC recalls MYDRIATIC-4: Tropicamide - Proparacaine - Phenylephrine - Ketorolac Sterile Ophthalmic Solution Drops 1% - 0.5% - 2.5% - 0.5%, 5mL bottle,…

    Subpotent: Out of Specification result observed for Ketorolac assay levels, below the 90.0-110.0% of label claim.

    FDA · DrugImprimis NJOF, LLCProduct was distributed to physician's offices nationwide.Official notice →
  4. 2023-10-18
  5. US
    Drugs & medicalClass II

    Becton Dickinson & Co. recalls BD BACTEC Plus Aerobic/F Culture Vials, Catalog Number 442023

    Certain vials of the BD BACTEC" Plus Aerobic/F Culture Vials have a labeling issue where duplicate barcode sequence numbers have been identified on more than one vial.

    FDA · DeviceBecton Dickinson & Co.Worldwide - US Nationwide distribution.Official notice →
  6. US
    Drugs & medicalClass II

    Beckman Coulter, Inc. recalls DxI 9000 Access Immunoassay Analyzer, REF C11137, Sofware Versions 1.15 and below; invitro diagnostic

    The firm has investigated and confirmed that when DxI 9000 is connected to Laboratory Information System (LIS, host), the sample barcode presented may be ignored and replaced with a sample ID that was previously in the…

    FDA · DeviceBeckman Coulter, Inc.Worldwide - US Nationwide distribution in the state of Florida, and the…Official notice →
  7. US
    Drugs & medicalClass II

    TECHNO-PATH MANUFACTURING LTD. recalls Multichem P, Part Code 08P90-10

    Technopath Manufacturing Ltd. has identified an issue with the glass vial commodity used to fill the Multichem S Plus Level 1, Multichem S Plus Level 2, Multichem S Plus Level 3, Multichem IA Plus and Multichem P…

    FDA · DeviceTECHNO-PATH MANUFACTURING LTD.US Nationwide distribution in the state of Illinois.Official notice →
  8. US
    Drugs & medicalClass II

    Encore Medical, LP recalls EMPOWR Acetabular System, Liner, 10¿ Hooded, HXe+, 40H, REF: 942-01-40H

    Knee insert and acetabular system parts were swapped within inner pack prior to the product being sealed within the tray, which could result in surgeons not having the correct product for the surgery due to different…

    FDA · DeviceEncore Medical, LPUS Nationwide distribution in the states of FL, CA, IN, HI, LA, MN, AZ, TX,…Official notice →
See all drugs & medical recalls →