Olympus Corporation of the Americas recalls OES 4000 Hysteroscope, Model/Catalog Number: A4674A
The required leakage test was not performed following of a re-work for a non-conformity. This could potentially result in liquid ingress into the eyepiece and cause a foggy image.
Read the official notice →- Date
- 2024-06-12
- Country
- United States
- Brand
- Olympus Corporation of the Americas
- Product
- OES 4000 Hysteroscope, Model/Catalog Number: A4674A
- Classification
- Class II
- Units
- US Nationwide distribution in the state of NY.
- Category
- Drugs & medical
- Source
- FDA · Device
Copied automatically from the US Food and Drug Administration. Details may be shortened or out of date; the official notice is the authoritative source. Disclaimer
What to do
- Check the product name, strength and lot number against the official notice.
- Do not stop a prescribed medicine without talking to your doctor or pharmacist first.
- Ask the pharmacy or supplier about a replacement or refund.
General guidance from Recall Atlas. The official notice from the US Food and Drug Administration has the exact models, batch codes and remedy, and it takes priority.
More Olympus Corporation of the Americas recalls
- 2024-08-14
US
Olympus Corporation of the Americas recalls Soltive Premium SuperPulsed Laser System, Model TFL-PLS
Difficulties in pairing the wireless footswitch with the Soltive Laser, potentially delaying the surgical procedure.
FDA · DeviceOlympus Corporation of the AmericasWorldwide distribution - US Nationwide and the countries of Australia, Canada,…Official notice →- 2024-07-31
US
Olympus Corporation of the Americas recalls ESG-410 Electrosurgical Generator- In conjunction with electrosurgical accessories intended for cutting and coagulation of…
Electrosurgical Generators ESG-410 does not power on upon use, may result in delays in initiating treatment or prolonged surgery
- 2024-07-10
US
Olympus Corporation of the Americas recalls EasySuite 4K Integrated Operating Room System Model/Catalog Number: EasySuite 4K- The EasySuite 4K system includes a touch…
A subset of the ES4K systems do not have complete records for earth leakage testing. The earth leakage test is a series of measurements intended to confirm that the amount of electrical current flowing to the earth is…
FDA · DeviceOlympus Corporation of the AmericasWorldwide distribution - US Nationwide and the countries of Germany, Hong Kong.Official notice →- 2024-05-29
US
Olympus Corporation of the Americas recalls InstaClear Lens Sheath. The InstaClear lens cleaner system and versatile sheath cleans and clears the lens of a rigid…
Potential for distal end of the sheath to break off during a procedure.
FDA · DeviceOlympus Corporation of the AmericasDomestic: Nationwide Distribution; Foreign: Australia, Canada, Germany, Japan.Official notice →- 2024-05-08
US
Olympus Corporation of the Americas recalls POWERSEAL 5 mm, 44 cm, Curved Jaw Sealer and Divider, Double Action is indicated for use in general surgery and such…
Upon initial activation of the seal function, the POWERSEAL non-conformance results in an immediate Incomplete Seal Cycle tone with the accompanying message on the generator screen, and no energy would be delivered to…
FDA · DeviceOlympus Corporation of the AmericasWorldwide distribution - US Nationwide and the countries of AU, CA, HK, KR.Official notice →- 2024-04-17
US
Olympus Corporation of the Americas recalls Hemostasis Probe, Model: CD-B622LA
The BICOAG Hemostasis Probe with a coaxial plug was mispackaged into the package and labeled as a BICOAG Hemostasis Probe with a fixed pin connector.
Other drugs & medical recalls in the United States
- 2024-06-19
US
US
PACIRA PHARMACEUTICALS INC recalls Drug Vial Label: Zilretta¿ (triamcinolone acetonide extended-release injectable suspension), 32 mg per vial, Rx only, Manufactured…
Failed Stability Specifications - at 12 months 2-8 degrees C followed by 6 weeks at 25 degrees C.
US
Winder Laboratories, LLC recalls Phenazopyridine HCl Tablets USP, 100 mg, 100-count bottles, Rx Only, Manufactured by: Winder Laboratories, LLC. 716 Patrick Industrial…
Product Mix Up. A bottle of Phenazopyridine HCl tablets USP 100 mg contained Phenobarbital tablets 16.2 mg.
- 2024-06-12
US
Remote Diagnostic Technologies Ltd. recalls Philips Tempus LS-Manual Defibrillator, Part Number: 00-3020 when using Tempus LS-Manual User/Operator Manual
Simultaneous ECG measurements with multiple devices, including a manual defibrillator, on the same patient may cause electrical noise or interference, which may lead to distortions of the ECG signal; distortions may…
FDA · DeviceRemote Diagnostic Technologies Ltd.US Nationwide distribution including in the states of VA, DC, MA, WY, AZ, FL,…Official notice →US
GE Medical Systems Information Technologies Inc recalls GE HealthCare CRITIKON ONE-CUF blood pressure cuffs, Model numbers: 1. ONE-A1-1B, ONE-CUF, SMALL ADULT, 1 TB…
Use of the blood pressure cuffs could result in inaccurate non-invasive blood pressure measurements.
FDA · DeviceGE Medical Systems Information Technologies IncUS Nationwide distribution in the states of AL, AR, AZ, CA, FL, GA, IA, IL,…Official notice →US
Medtronic Navigation, Inc.-Littleton recalls O-arm O2 Imaging System-Mobile x-ray system designed for 2D fluoroscopic and 3D imaging for adult and pediatric patients.…
Potential for an electrical component of the O-arm" O2 Imaging System to intermittently shut off,failure can cause the potential for surgical delay, additional imaging, cancellation and rescheduling of surgery, or…
FDA · DeviceMedtronic Navigation, Inc.-LittletonWorldwide - US Nationwide distribution in the states of AZ, CA, CO, CT, FL,…Official notice →