Drugs & medicalClass II

Olympus Corporation of the Americas recalls Hemostasis Probe, Model: CD-B622LA

The BICOAG Hemostasis Probe with a coaxial plug was mispackaged into the package and labeled as a BICOAG Hemostasis Probe with a fixed pin connector.

Read the official notice →
Date
2024-04-17
Country
United States
Brand
Olympus Corporation of the Americas
Product
Hemostasis Probe, Model: CD-B622LA
Classification
Class II
Units
International distribution only.
Category
Drugs & medical
Source
FDA · Device

Copied automatically from the US Food and Drug Administration. Details may be shortened or out of date; the official notice is the authoritative source. Disclaimer

What to do

  1. Check the product name, strength and lot number against the official notice.
  2. Do not stop a prescribed medicine without talking to your doctor or pharmacist first.
  3. Ask the pharmacy or supplier about a replacement or refund.

General guidance from Recall Atlas. The official notice from the US Food and Drug Administration has the exact models, batch codes and remedy, and it takes priority.

More Olympus Corporation of the Americas recalls

  1. 2024-06-12
  2. US
    Drugs & medicalClass II

    Olympus Corporation of the Americas recalls OES 4000 Hysteroscope, Model/Catalog Number: A4674A

    The required leakage test was not performed following of a re-work for a non-conformity. This could potentially result in liquid ingress into the eyepiece and cause a foggy image.

    FDA · DeviceOlympus Corporation of the AmericasUS Nationwide distribution in the state of NY.Official notice →
  3. 2024-05-29
  4. US
    Drugs & medicalClass II

    Olympus Corporation of the Americas recalls InstaClear Lens Sheath. The InstaClear lens cleaner system and versatile sheath cleans and clears the lens of a rigid…

    Potential for distal end of the sheath to break off during a procedure.

    FDA · DeviceOlympus Corporation of the AmericasDomestic: Nationwide Distribution; Foreign: Australia, Canada, Germany, Japan.Official notice →
  5. 2024-05-08
  6. US
    Drugs & medicalClass II

    Olympus Corporation of the Americas recalls POWERSEAL 5 mm, 44 cm, Curved Jaw Sealer and Divider, Double Action is indicated for use in general surgery and such…

    Upon initial activation of the seal function, the POWERSEAL non-conformance results in an immediate Incomplete Seal Cycle tone with the accompanying message on the generator screen, and no energy would be delivered to…

    FDA · DeviceOlympus Corporation of the AmericasWorldwide distribution - US Nationwide and the countries of AU, CA, HK, KR.Official notice →
  7. 2024-04-03
  8. US
    Drugs & medicalClass II

    Olympus Corporation of the Americas recalls SOLTIVE Pro SuperPulsed Laser (TFL-SLS )

    A non-compliant power cord may have been supplied in regions that do not utilize the Type B or Type E/F plug configurations.

    FDA · DeviceOlympus Corporation of the AmericasDomestic: OUS Only; International distribution in the states of Australia,…Official notice →
  9. US
    Drugs & medicalClass II

    Olympus Corporation of the Americas recalls Insufflation unit, UHI-4.

    Olympus has become aware of an increased trend of both repairs and customer complaints reporting "UHI-4 stopping air supply due to the front panel LED turning off".

    FDA · DeviceOlympus Corporation of the AmericasDomestic: Nationwide Distribution.Official notice →
  10. US
    Drugs & medicalClass II

    Olympus Corporation of the Americas recalls InstaClear Lens Cleaner-indicated for use during routine diagnostic procedures and during endoscopic sinus surgery. Model…

    Japanese IFU version had slightly different language and illustrations from the English IFU, which could contribute to the tip and plate falling off. In order to maintain consistency globally, the Japanese IFU will be…

    FDA · DeviceOlympus Corporation of the AmericasInternational distribution to the country of Japan.Official notice →
All 114 Olympus Corporation of the Americas recalls →

Other drugs & medical recalls in the United States

  1. 2024-04-17
  2. US
    Drugs & medicalClass I

    Fresenius Kabi USA, LLC recalls LVP Software of the Ivenix Infusion System (IIS), Version 5.8.0.

    Software has anomalies that have the potential to cause underdose, overdose, or delay in therapy which could lead to serious patient harm or death.

    FDA · DeviceFresenius Kabi USA, LLCUS Distribution: CA, CO, ID, MI, NE, NJ, NV, OK, TX, UT, VA, WA, & WI.Official notice →
  3. US
    Drugs & medicalClass II

    Microbiologics Inc recalls KWIK-STIK Rhizopus stolonifer (+) derived from ATCC 6227b, a) REF 0209P and b) REF 0209K, Quality control kit for culture media

    The positive control material (Rhizopus stolonifer) within the KWIK-STIK assembly was contaminated with another fungal organism (Aspergillus brasiliensis). As a result, users may observe A. brasiliensis growth when…

    FDA · DeviceMicrobiologics IncWorldwide - US Nationwide distribution in the states of CA, MI, KY, MO, NM,…Official notice →
  4. US
    Drugs & medicalClass II

    RAYSEARCH LABORATORIES AB recalls RayStation Software Version Numbers version numbers 8B, 9A, 9B, 10A, 10B, 11A, 11B, 12A, 12B, 2023B, 2024A including some service packs.

    A use error that has occurred with RayStation/RayPlan, where material override using Silicon, Si, was incorrectly selected for a silicone gel polymer implant.

    FDA · DeviceRAYSEARCH LABORATORIES ABWorldwide distribution - US Nationwide and the countries of Australia,…Official notice →
  5. US
  6. US
    Drugs & medicalClass II

    Teva Pharmaceuticals USA, Inc recalls Isotretinoin Capsules, USP 40mg, 30-count 3x10 blister packs per carton, Rx Only, Manufactured For: Teva Pharmaceuticals USA, Inc.,…

    Superpotent Drug: The 3-month stability result for assay was found to be above specification limit

    FDA · DrugTeva Pharmaceuticals USA, IncNationwide in the USAOfficial notice →
  7. US