Drugs & medicalClass II

GE Medical Systems Information Technologies Inc recalls GE HealthCare CRITIKON ONE-CUF blood pressure cuffs, Model numbers: 1. ONE-A1-1B, ONE-CUF, SMALL ADULT, 1 TB…

Use of the blood pressure cuffs could result in inaccurate non-invasive blood pressure measurements.

Read the official notice →
Date
2024-06-12
Country
United States
Brand
GE Medical Systems Information Technologies
Product
GE HealthCare CRITIKON ONE-CUF blood pressure cuffs, Model numbers: 1. ONE-A1-1B, ONE-CUF, SMALL ADULT, 1 TB BAYONET, 17 - 25 CM, 20/BOX;…
Classification
Class II
Units
US Nationwide distribution in the states of AL, AR, AZ, CA, FL, GA, IA, IL,…
Category
Drugs & medical
Source
FDA · Device

Copied automatically from the US Food and Drug Administration. Details may be shortened or out of date; the official notice is the authoritative source. Disclaimer

What to do

  1. Check the product name, strength and lot number against the official notice.
  2. Do not stop a prescribed medicine without talking to your doctor or pharmacist first.
  3. Ask the pharmacy or supplier about a replacement or refund.

General guidance from Recall Atlas. The official notice from the US Food and Drug Administration has the exact models, batch codes and remedy, and it takes priority.

More GE Medical Systems Information Technologies recalls

  1. 2026-08-12
  2. US
    Drugs & medicalClass II

    GE Medical Systems Information Technologies Inc recalls GE Healthcare MAC VU360, REF# part number 2030361-023; Electrocardiograph

    Software issue issue with the MAC VU360 v3.0 system. After selecting the order refresh icon, an open preview or review screen might fail to close automatically after 2 minutes of inactivity. As a result, data from a…

    FDA · DeviceGE Medical Systems Information Technologies IncWorldwide distribution - US Nationwide and the countries of Belgium, Hong…Official notice →
  3. 2026-06-10
  4. US
    Drugs & medicalClass II

    GE Medical Systems Information Technologies Inc recalls CASE v7.0 Cardiac Testing System Model no. 3031234-001 Product Code SU3

    Electrical sparking may occur at the printer head. If the Electrostatic Discharge (ESD) printer brush becomes dislodged, electrical sparking at the printer head can result in charring or discoloration of thermal…

    FDA · DeviceGE Medical Systems Information Technologies IncWorldwide distribution - US Nationwide and the countries of Algeria,…Official notice →
  5. 2026-06-03
  6. US
    Drugs & medicalClass II

    GE Medical Systems Information Technologies Inc recalls GE HealthCare TRADE IN CARESCAPE TELEMETRY SERVER ARK-2250L - NORTH AMERICA, REF #5697603-902, Product Code KSA;…

    GE HealthCare has become aware of an unlikely situation where potential loss of ECG and SpO2 monitoring involving ApexPro CARESCAPE Telemetry Server (CTS) v5.0 and v6.0, can result from a persistent NO COMM or OFF…

    FDA · DeviceGE Medical Systems Information Technologies IncWorldwide distribution.Official notice →
  7. 2025-06-18
  8. US
    Drugs & medicalClass II

    GE Medical Systems Information Technologies Inc recalls MUSE 5 Systems, Model Numbers 2021274-001, 2023032-005, 2027478-001, Large capacity client-server-based computer…

    The MUSE Administrator Account active directory or local credential could potentially be used from your local network to log into a GE HealthCare MUSE 5 system and potentially manipulate recorded patient data.

    FDA · DeviceGE Medical Systems Information Technologies IncWorldwide - US Nationwide distribution.Official notice →
  9. 2024-07-10
  10. US
    Drugs & medicalClass II

    GE Medical Systems Information Technologies Inc recalls Prucka 3 Amplifier Model #P1801PA used with CardioLab AltiX electrophysiology recording system and ComboLab…

    Capacitors in certain Prucka 3 Amplifiers used with CardioLab / ComboLab systems could fail resulting in transient oscillations on the display and inability to view surface and intracardiac ECG waveforms.

    FDA · DeviceGE Medical Systems Information Technologies IncWorldwide - US Nationwide distribution in the states of AK, AZ, CA, CO, FL,…Official notice →
  11. 2023-09-27
  12. US
    Drugs & medicalClass II

    GE Medical Systems Information Technologies Inc recalls Prucka 3 Amplifiers, Model P1801PA, and Field Replaceable Units (FRUs), Model 5875569, used with CardioLab AltiX…

    A diode on the power supply of the Prucka 3 Amplifier used with CardioLab/ComboLab systems could reach elevated temperatures and fail, then power off and become inoperable.

    FDA · DeviceGE Medical Systems Information Technologies IncDistribution was made to CA, CO, FL, GA, IL, IN, KY, MA, NC, NJ, OH, OR, PA,…Official notice →
All 10 GE Medical Systems Information Technologies recalls →

Other drugs & medical recalls in the United States

  1. 2024-06-19
  2. US
    Drugs & medicalClass II

    Winder Laboratories, LLC recalls Phenazopyridine HCl Tablets USP, 100 mg, 100-count bottles, Rx Only, Manufactured by: Winder Laboratories, LLC. 716 Patrick Industrial…

    Product Mix Up. A bottle of Phenazopyridine HCl tablets USP 100 mg contained Phenobarbital tablets 16.2 mg.

    FDA · DrugWinder Laboratories, LLCNationwide within the United StatesOfficial notice →
  3. 2024-06-12
  4. US
    Drugs & medicalClass II

    Olympus Corporation of the Americas recalls OES 4000 Hysteroscope, Model/Catalog Number: A4674A

    The required leakage test was not performed following of a re-work for a non-conformity. This could potentially result in liquid ingress into the eyepiece and cause a foggy image.

    FDA · DeviceOlympus Corporation of the AmericasUS Nationwide distribution in the state of NY.Official notice →
  5. US
    Drugs & medicalClass II

    Remote Diagnostic Technologies Ltd. recalls Philips Tempus LS-Manual Defibrillator, Part Number: 00-3020 when using Tempus LS-Manual User/Operator Manual

    Simultaneous ECG measurements with multiple devices, including a manual defibrillator, on the same patient may cause electrical noise or interference, which may lead to distortions of the ECG signal; distortions may…

    FDA · DeviceRemote Diagnostic Technologies Ltd.US Nationwide distribution including in the states of VA, DC, MA, WY, AZ, FL,…Official notice →
  6. US
    Drugs & medicalClass II

    Medtronic Navigation, Inc.-Littleton recalls O-arm O2 Imaging System-Mobile x-ray system designed for 2D fluoroscopic and 3D imaging for adult and pediatric patients.…

    Potential for an electrical component of the O-arm" O2 Imaging System to intermittently shut off,failure can cause the potential for surgical delay, additional imaging, cancellation and rescheduling of surgery, or…

    FDA · DeviceMedtronic Navigation, Inc.-LittletonWorldwide - US Nationwide distribution in the states of AZ, CA, CO, CT, FL,…Official notice →
  7. US
    Drugs & medicalClass II

    Nihon Kohden America Inc recalls BSM-3000 Series Bedside Monitor REF BSM-3572A The Life Scope BSM-3000 Series Bedside Monitoring System is intended to monitor, display…

    Due to software issue the device may give false "SpO2 Probe Failure" alarm

    FDA · DeviceNihon Kohden America IncU.S. Nationwide distribution in the state of IA.Official notice →
  8. US
    Drugs & medicalClass II

    MEDLINE INDUSTRIES, LP - Northfield recalls MEDLINE STERILE IRIS SCISSORS CVD/STD REF DYNJ04049 (case) and DYNJ04049H (individual unit)

    There is the potential of the tip protector to fall off 4.5" sterile, curved, Iris Scissors

    FDA · DeviceMEDLINE INDUSTRIES, LP - NorthfieldWorldwide distribution - US Nationwide and the countries of Singapore, Canada.Official notice →
See all drugs & medical recalls →