GE Medical Systems Information Technologies recalls

10 official recalls and safety alerts involving GE Medical Systems Information Technologies since 2023-09, from FDA · Device. Newest first. Check the official notice for exact models and batch codes.

This list only includes recalls Recall Atlas has collected since 2021-10. A product not listed here may still be affected; check with GE Medical Systems Information Technologies or the relevant agency.

  1. 2026-09-23
  2. US
    Drugs & medicalClass II

    GE Medical Systems Information Technologies Inc recalls Carescape Central Station CSCS V3 MAS800 Desktop Part Number 5867474-04, 5867474-13 Product code STU CSCS V3…

    Device may not resume patient monitoring for at least 3 minutes following system restart. Issue can lead to potential delay in recognition of clinically significant events.

    FDA · DeviceGE Medical Systems Information Technologies IncWorldwide - US Nationwide distribution.Official notice →
  3. 2026-09-02
  4. US
    Drugs & medicalClass II

    GE Medical Systems Information Technologies Inc recalls PORTRAIT CORE SERVICES SOFTWARE V1.0.5, Ref# 5697353

    Software issue may result in a loss of patient monitoring at the Hub and Central Viewer. In addition, all alarm settings could be reset to factory defaults following the update and will not follow the alarm settings…

    FDA · DeviceGE Medical Systems Information Technologies IncUS: MI, NY and Australia, Czech Republic, Finland, Germany, Italy,…Official notice →
  5. 2026-08-12
  6. US
    Drugs & medicalClass II

    GE Medical Systems Information Technologies Inc recalls GE Healthcare Portrait Core Services SW V 1.2.0, REF 5700592. patient monitoring system

    A software vulnerability may allow unauthorized persons to gain privileged access to the device, which could result in loss of patient monitoring, missed patient alarms, and/or access to Protected Health Information…

    FDA · DeviceGE Medical Systems Information Technologies IncWorldwide - US Nationwide distribution in the state of MI, and the countries…Official notice →
  7. US
    Drugs & medicalClass II

    GE Medical Systems Information Technologies Inc recalls GE Healthcare MAC VU360, REF# part number 2030361-023; Electrocardiograph

    Software issue issue with the MAC VU360 v3.0 system. After selecting the order refresh icon, an open preview or review screen might fail to close automatically after 2 minutes of inactivity. As a result, data from a…

    FDA · DeviceGE Medical Systems Information Technologies IncWorldwide distribution - US Nationwide and the countries of Belgium, Hong…Official notice →
  8. 2026-06-10
  9. US
    Drugs & medicalClass II

    GE Medical Systems Information Technologies Inc recalls CASE v7.0 Cardiac Testing System Model no. 3031234-001 Product Code SU3

    Electrical sparking may occur at the printer head. If the Electrostatic Discharge (ESD) printer brush becomes dislodged, electrical sparking at the printer head can result in charring or discoloration of thermal…

    FDA · DeviceGE Medical Systems Information Technologies IncWorldwide distribution - US Nationwide and the countries of Algeria,…Official notice →
  10. 2026-06-03
  11. US
    Drugs & medicalClass II

    GE Medical Systems Information Technologies Inc recalls GE HealthCare TRADE IN CARESCAPE TELEMETRY SERVER ARK-2250L - NORTH AMERICA, REF #5697603-902, Product Code KSA;…

    GE HealthCare has become aware of an unlikely situation where potential loss of ECG and SpO2 monitoring involving ApexPro CARESCAPE Telemetry Server (CTS) v5.0 and v6.0, can result from a persistent NO COMM or OFF…

    FDA · DeviceGE Medical Systems Information Technologies IncWorldwide distribution.Official notice →
  12. 2025-06-18
  13. US
    Drugs & medicalClass II

    GE Medical Systems Information Technologies Inc recalls MUSE 5 Systems, Model Numbers 2021274-001, 2023032-005, 2027478-001, Large capacity client-server-based computer…

    The MUSE Administrator Account active directory or local credential could potentially be used from your local network to log into a GE HealthCare MUSE 5 system and potentially manipulate recorded patient data.

    FDA · DeviceGE Medical Systems Information Technologies IncWorldwide - US Nationwide distribution.Official notice →
  14. 2024-07-10
  15. US
    Drugs & medicalClass II

    GE Medical Systems Information Technologies Inc recalls Prucka 3 Amplifier Model #P1801PA used with CardioLab AltiX electrophysiology recording system and ComboLab…

    Capacitors in certain Prucka 3 Amplifiers used with CardioLab / ComboLab systems could fail resulting in transient oscillations on the display and inability to view surface and intracardiac ECG waveforms.

    FDA · DeviceGE Medical Systems Information Technologies IncWorldwide - US Nationwide distribution in the states of AK, AZ, CA, CO, FL,…Official notice →
  16. 2024-06-12
  17. US
    Drugs & medicalClass II

    GE Medical Systems Information Technologies Inc recalls GE HealthCare CRITIKON ONE-CUF blood pressure cuffs, Model numbers: 1. ONE-A1-1B, ONE-CUF, SMALL ADULT, 1 TB…

    Use of the blood pressure cuffs could result in inaccurate non-invasive blood pressure measurements.

    FDA · DeviceGE Medical Systems Information Technologies IncUS Nationwide distribution in the states of AL, AR, AZ, CA, FL, GA, IA, IL,…Official notice →
  18. 2023-09-27
  19. US
    Drugs & medicalClass II

    GE Medical Systems Information Technologies Inc recalls Prucka 3 Amplifiers, Model P1801PA, and Field Replaceable Units (FRUs), Model 5875569, used with CardioLab AltiX…

    A diode on the power supply of the Prucka 3 Amplifier used with CardioLab/ComboLab systems could reach elevated temperatures and fail, then power off and become inoperable.

    FDA · DeviceGE Medical Systems Information Technologies IncDistribution was made to CA, CO, FL, GA, IL, IN, KY, MA, NC, NJ, OH, OR, PA,…Official notice →