Drugs & medicalClass II

Olympus Corporation of the Americas recalls Spiration Valve System (SVS) Valve in Cartridge -A one-way endobronchial valves indicated for adult patients with shortness…

Mislabeled Patient Charge Label indicating model number SVS-V7-00 (7mm) instead of SVS-V9-00 (9mm)

Read the official notice →
Date
2022-02-09
Country
United States
Brand
Olympus Corporation of the Americas
Product
Spiration Valve System (SVS) Valve in Cartridge -A one-way endobronchial valves indicated for adult patients with shortness of breath and…
Classification
Class II
Units
US Nationwide distribution in the states of AZ, CA, MA, NJ, PA, SC, TX, WI.
Category
Drugs & medical
Source
FDA · Device

Copied automatically from the US Food and Drug Administration. Details may be shortened or out of date; the official notice is the authoritative source. Disclaimer

What to do

  1. Check the product name, strength and lot number against the official notice.
  2. Do not stop a prescribed medicine without talking to your doctor or pharmacist first.
  3. Ask the pharmacy or supplier about a replacement or refund.

General guidance from Recall Atlas. The official notice from the US Food and Drug Administration has the exact models, batch codes and remedy, and it takes priority.

More Olympus Corporation of the Americas recalls

  1. 2022-05-18
  2. US
    Drugs & medicalClass II

    Olympus Corporation of the Americas recalls BF-3C160: EVIS EXERA Bronchovideoscope

    Bronchofiberscope Updated instructions on the use of certain chemicals to be used for reprocessing, conditions for ETO gas sterilization, rinsing steps after high-level disinfection, and a recommendation on using…

    FDA · DeviceOlympus Corporation of the AmericasUS Nationwide distribution.Official notice →
  3. 2022-05-11
  4. US
    Drugs & medicalClass II

    Olympus Corporation of the Americas recalls Ultrasonic Gastrovideoscope, GF-UC140P-AL5. Intended for endoscopic real-time ultrasound imaging.

    GF-UC140P-AL5 is no longer a compatible endoscope for reprocessing in the OER-Pro and presents an infection control risk to patients

    FDA · DeviceOlympus Corporation of the AmericasNationwideOfficial notice →
  5. 2022-02-23
  6. US
    Drugs & medicalClass II

    Olympus Corporation of the Americas recalls OER-Elite Endoscope Reprocessor-for use in cleaning and high-level disinfection of heat sensitive Olympus flexible…

    The detergent and alcohol tank tubing may have been incorrectly assembled in the OER-Elite. If the tubing is incorrectly assembled, alcohol will be dispensed instead of detergent for the cleaning cycle and detergent…

    FDA · DeviceOlympus Corporation of the AmericasNationwideOfficial notice →
  7. 2021-12-22
  8. US
    Drugs & medicalClass II

    Olympus Corporation of the Americas recalls EVIS EXERA II DUODENOVIDEOSCOPE OLYMPUS TJF-Q180V Model Number: TJF-Q180V

    Use of a TJF-Q180V with adhesive deterioration or other damages may pose a risk of endoscope contamination due to ineffective reprocessing or fluid invasion; contaminated endoscope can present an infection risk to…

    FDA · DeviceOlympus Corporation of the AmericasUS Nationwide distribution.Official notice →
  9. 2021-11-03
  10. US
    Drugs & medicalClass II

    Olympus Corporation of the Americas recalls ASC PneumoLiner device, Part No. WA90500US (US market only) The Pneumoliner device is intended for use as a multiple…

    Due to manufacturing error, the Pneumoliner Bag Distal Tab that exits the Introducer shaft is in the wrong orientation. This results in the user deploying the bag upside down, which will make tissue encapsulation and…

    FDA · DeviceOlympus Corporation of the AmericasUS Nationwide distribution to states of: AL, CA, CT, FL, IL, MD, MI, MN, MO,…Official notice →
  11. 2021-10-06
  12. US
    Drugs & medicalClass II

    Olympus Corporation of the Americas recalls BF-XP60 OES Bronchofiberscope, Model No. BF-XP60

    The firm is issuing validated, new reprocessing instructions according to the FDA 2015 guidance, "Reprocessing Medical Devices in Health Care Settings: Validation Methods and Labeling."

    FDA · DeviceOlympus Corporation of the AmericasDomestic distribution nationwide. Product also distributed globally.Official notice →
All 114 Olympus Corporation of the Americas recalls →

Other drugs & medical recalls in the United States

  1. 2022-02-09
  2. US
    Drugs & medicalClass II

    Young Dental Mfg Co I LLC recalls YOUNG DH Dry Heat Indicators, PRESSURE SENSITIVE LABELS with COLOR INDICATOR STRIP, REF 116312 Intended for use by a health care…

    The Dry Heat Indicator Labels packaging may contain TTS Indicator Labels instead of Dry Heat Indicator Labels.

    FDA · DeviceYoung Dental Mfg Co I LLCUS DistributionOfficial notice →
  3. US
    Drugs & medicalClass II

    MEDLINE INDUSTRIES, LP - Northfield recalls HARDY DIAGNOSTICS CatScreen, 25dsk/vial

    Product stored incorrectly in temperature controlled setting instead of refrigeration.

    FDA · DeviceMEDLINE INDUSTRIES, LP - NorthfieldUS Nationwide distribution in the state of California.Official notice →
  4. US
    Drugs & medicalClass II

    Luminex Corporation recalls ARIES HSV 1&2 Assay, Catalog/REF# 50-10017

    Impacted lot may leak inside the ARIES instrument. Testing confirmed a leak rate of 0.9%. Leaking cassettes could potentially lead to incorrect results or a delay in reporting results due to an Invalid.

    FDA · DeviceLuminex CorporationWorldwide distribution - US Nationwide distribution in the states of New…Official notice →
  5. US
    Drugs & medicalClass I

    Abiomed, Inc. recalls OXY-1 System, Model Number 001-0500-001. Used during cardiopulmonary bypass. OXY-1 System consists of Console, Pump Drive, E Drive, and…

    The firm has received complaints of the OXY-1 System Console experiencing a power interruption while in use, which disrupts therapy delivered to a patient on support. Disruption of therapy could lead to prolonged…

    FDA · DeviceAbiomed, Inc.US distribution to FL, GA, IN, MD, NC, PA, TX.Official notice →
  6. US
    Drugs & medicalClass II

    Philips North America Llc recalls eCareManager version 4.2.1, eCareManager version 4.3.2, eCareManager version 4.3.3, eCareManager version 4.5.1, eCareManager version…

    eCareManager (eCM) Sentry Score software not approved for use

    FDA · DevicePhilips North America LlcWorldwide distribution - US Nationwide and the country of Australia.Official notice →
  7. US
    Drugs & medicalClass II

    Smith & Nephew, Inc. recalls INTEGRA SURFIX ALPHA Screwdriver Torx 10, REF 219 135 ND

    The affected screwdrivers are out of specification and may not mate with the Surfix Alpha Lock screw

    FDA · DeviceSmith & Nephew, Inc.US Nationwide distribution in the states of CA and OH.Official notice →
See all drugs & medical recalls →