Smith & Nephew, Inc. recalls INTEGRA SURFIX ALPHA Screwdriver Torx 10, REF 219 135 ND
The affected screwdrivers are out of specification and may not mate with the Surfix Alpha Lock screw
Read the official notice →- Date
- 2022-02-09
- Country
- United States
- Brand
- Smith & Nephew
- Product
- INTEGRA SURFIX ALPHA Screwdriver Torx 10, REF 219 135 ND
- Classification
- Class II
- Units
- US Nationwide distribution in the states of CA and OH.
- Category
- Drugs & medical
- Source
- FDA · Device
Copied automatically from the US Food and Drug Administration. Details may be shortened or out of date; the official notice is the authoritative source. Disclaimer
What to do
- Check the product name, strength and lot number against the official notice.
- Do not stop a prescribed medicine without talking to your doctor or pharmacist first.
- Ask the pharmacy or supplier about a replacement or refund.
General guidance from Recall Atlas. The official notice from the US Food and Drug Administration has the exact models, batch codes and remedy, and it takes priority.
More Smith & Nephew recalls
- 2022-07-27
US
Smith & Nephew, Inc. recalls (1) Genesis II Non-Porous Tibial Baseplate Size 5 Right, Product #71420186. (2) Genesis II Non-Porous Tibial Baseplate Size 4 Left,…
The product was mispackaged (swapped). The devices within the inner tray matches the inner tray but does not match the outer box label and the chart sticks.
FDA · DeviceSmith & Nephew, Inc.US Nationwide distribution in the states of AZ, CA, GA, KS, LA, MN, MI, MO,…Official notice →- 2022-05-25
US
Smith & Nephew Inc recalls LEGION NARROW PS OXIN SZ 6N RT and JRNY II BCS FEMORAL OXIN LT SZ 5
It was reported that a JOURNEY II BCS Femoral OXIN LT SZ5 was in a box labelled as a LEGION NARROW PS OXIN SZ 6N RT.
FDA · DeviceSmith & Nephew IncWorldwide distribution - US Nationwide distribution in the states of CT, NC,…Official notice →- 2022-02-16
US
Smith & Nephew, Inc. recalls Acufex Access Advanced Positioning Kit - indicated for use to position and distract the extremity of patient during open and arthroscopic…
Products were shipped to the Global Distribution Center instead of the sterilizer and therefore product did not undergo the sterilization process
FDA · DeviceSmith & Nephew, Inc.Worldwide distribution - US Nationwide distribution in the states of CO, CT,…Official notice →- 2021-12-22
US
Smith & Nephew, Inc. recalls smith&Nephew EVOS MEDIAL DISTAL FEMUR PLATE, L, 5H, 90MM, REF 72573101
Testing vendor notified Smith & Nephew that previous MRI scanning conditions provided in Exponents reports were erroneous. Erroneous data from the reports was incorporated in the released IFU 81110294 EVOS Large…
FDA · DeviceSmith & Nephew, Inc.US Nationwide distribution in the states of CO, FL, GA, MA, MN, MO, NC, NJ,…Official notice →- 2021-11-24
US
Smith & Nephew, Inc. recalls Smith and Nephew EVOS Self-Tapping Locking Screw SST: (1)REF 72413575, 3.5MM x 75MM, and (2) REF 72413570, 3.5MM X 70MM. Sterile.
The labels on the packages of screws were switched.
FDA · DeviceSmith & Nephew, Inc.Distribution was made to AL, AZ, CA, CO, CT, FL, GA, IL, IN, KY, LA, MA, ME,…Official notice →- 2021-10-27
US
Smith & Nephew, Inc. recalls 6.5MMX165MM CANNULATED SCREW
The product label incorrectly indicated that the screws are fully threaded, but the screws contained in the package were partially threaded.
FDA · DeviceSmith & Nephew, Inc.Worldwide distribution - US Nationwide distribution in the states of OK, TN…Official notice →
Other drugs & medical recalls in the United States
- 2022-02-09
US
Olympus Corporation of the Americas recalls Spiration Valve System (SVS) Valve in Cartridge -A one-way endobronchial valves indicated for adult patients with shortness…
Mislabeled Patient Charge Label indicating model number SVS-V7-00 (7mm) instead of SVS-V9-00 (9mm)
FDA · DeviceOlympus Corporation of the AmericasUS Nationwide distribution in the states of AZ, CA, MA, NJ, PA, SC, TX, WI.Official notice →US
Abiomed, Inc. recalls OXY-1 System, Model Number 001-0500-001. Used during cardiopulmonary bypass. OXY-1 System consists of Console, Pump Drive, E Drive, and…
The firm has received complaints of the OXY-1 System Console experiencing a power interruption while in use, which disrupts therapy delivered to a patient on support. Disruption of therapy could lead to prolonged…
US
Philips North America Llc recalls eCareManager version 4.2.1, eCareManager version 4.3.2, eCareManager version 4.3.3, eCareManager version 4.5.1, eCareManager version…
eCareManager (eCM) Sentry Score software not approved for use
FDA · DevicePhilips North America LlcWorldwide distribution - US Nationwide and the country of Australia.Official notice →US
Siemens Medical Solutions USA, Inc. recalls ACUSON Sequoia Ultrasound Imaging System with Software Version VA25A/B
Due to a calculation error in the measurement when using 2D trace (manual trace) tool. The trace circumference value is overestimated and may potentially result in misdiagnosis of a patient's condition.
FDA · DeviceSiemens Medical Solutions USA, Inc.Worldwide distribution - U.S. Nationwide distribution in the states of AL, AK,…Official notice →US
Microbiologics Inc recalls KWIK-STI(TM), GBS QC Set (Live Culture), REF 0243U. Specifically affected component: KWIK-STIK(TM), Lactobacillus acidophilus derived from…
The expiration date on the label of one component (negative control 0243U) in the set lot is incorrectly labeled. The KWIK-STIK negative control 0243U was labeled with an expiration date of 7/31/2020 when it should…
FDA · DeviceMicrobiologics IncUS Nationwide distribution in the states of AL, CA, CO, IL, MD, MO, MS, NM,…Official notice →US
Immuno-Mycologics, Inc recalls CRYPTOCOCCAL ANTIGEN LATERAL FLOW ASSAY, IMMY, Lateral Flow Assay for the Detection of Cryptococcal Antigen in serum and CSF, REF CR2003,…
The firm found immunoassay products used to detect cryptococcal antigen in serum and cerebrospinal fluid to have reduced specificity (90% now versus 99% before). As a result, a small number of samples with positive…
FDA · DeviceImmuno-Mycologics, IncUS Nationwide distribution in the states of AL, AR, AZ, CA, CO, CT, FL, GA,…Official notice →