Smith & Nephew, Inc. recalls smith&Nephew EVOS MEDIAL DISTAL FEMUR PLATE, L, 5H, 90MM, REF 72573101
Testing vendor notified Smith & Nephew that previous MRI scanning conditions provided in Exponents reports were erroneous. Erroneous data from the reports was incorporated in the released IFU 81110294 EVOS Large…
Read the official notice →- Date
- 2021-12-22
- Country
- United States
- Brand
- Smith & Nephew
- Product
- smith&Nephew EVOS MEDIAL DISTAL FEMUR PLATE, L, 5H, 90MM, REF 72573101
- Classification
- Class II
- Units
- US Nationwide distribution in the states of CO, FL, GA, MA, MN, MO, NC, NJ,…
- Category
- Drugs & medical
- Source
- FDA · Device
Copied automatically from the US Food and Drug Administration. Details may be shortened or out of date; the official notice is the authoritative source. Disclaimer
What to do
- Check the product name, strength and lot number against the official notice.
- Do not stop a prescribed medicine without talking to your doctor or pharmacist first.
- Ask the pharmacy or supplier about a replacement or refund.
General guidance from Recall Atlas. The official notice from the US Food and Drug Administration has the exact models, batch codes and remedy, and it takes priority.
More Smith & Nephew recalls
- 2022-07-27
US
Smith & Nephew, Inc. recalls (1) Genesis II Non-Porous Tibial Baseplate Size 5 Right, Product #71420186. (2) Genesis II Non-Porous Tibial Baseplate Size 4 Left,…
The product was mispackaged (swapped). The devices within the inner tray matches the inner tray but does not match the outer box label and the chart sticks.
FDA · DeviceSmith & Nephew, Inc.US Nationwide distribution in the states of AZ, CA, GA, KS, LA, MN, MI, MO,…Official notice →- 2022-05-25
US
Smith & Nephew Inc recalls LEGION NARROW PS OXIN SZ 6N RT and JRNY II BCS FEMORAL OXIN LT SZ 5
It was reported that a JOURNEY II BCS Femoral OXIN LT SZ5 was in a box labelled as a LEGION NARROW PS OXIN SZ 6N RT.
FDA · DeviceSmith & Nephew IncWorldwide distribution - US Nationwide distribution in the states of CT, NC,…Official notice →- 2022-02-16
US
Smith & Nephew, Inc. recalls Acufex Access Advanced Positioning Kit - indicated for use to position and distract the extremity of patient during open and arthroscopic…
Products were shipped to the Global Distribution Center instead of the sterilizer and therefore product did not undergo the sterilization process
FDA · DeviceSmith & Nephew, Inc.Worldwide distribution - US Nationwide distribution in the states of CO, CT,…Official notice →- 2022-02-09
US
Smith & Nephew, Inc. recalls INTEGRA SURFIX ALPHA Screwdriver Torx 10, REF 219 135 ND
The affected screwdrivers are out of specification and may not mate with the Surfix Alpha Lock screw
FDA · DeviceSmith & Nephew, Inc.US Nationwide distribution in the states of CA and OH.Official notice →- 2021-11-24
US
Smith & Nephew, Inc. recalls Smith and Nephew EVOS Self-Tapping Locking Screw SST: (1)REF 72413575, 3.5MM x 75MM, and (2) REF 72413570, 3.5MM X 70MM. Sterile.
The labels on the packages of screws were switched.
FDA · DeviceSmith & Nephew, Inc.Distribution was made to AL, AZ, CA, CO, CT, FL, GA, IL, IN, KY, LA, MA, ME,…Official notice →- 2021-10-27
US
Smith & Nephew, Inc. recalls 6.5MMX165MM CANNULATED SCREW
The product label incorrectly indicated that the screws are fully threaded, but the screws contained in the package were partially threaded.
FDA · DeviceSmith & Nephew, Inc.Worldwide distribution - US Nationwide distribution in the states of OK, TN…Official notice →
Other drugs & medical recalls in the United States
- 2021-12-22
US
Olympus Corporation of the Americas recalls EVIS EXERA II DUODENOVIDEOSCOPE OLYMPUS TJF-Q180V Model Number: TJF-Q180V
Use of a TJF-Q180V with adhesive deterioration or other damages may pose a risk of endoscope contamination due to ineffective reprocessing or fluid invasion; contaminated endoscope can present an infection risk to…
US
Cardinal Health recalls RoyalSilk Non-Reinforced Surgical Gowns X-large SKU: 9548
Potential for open packaging seals compromising the sterility of the surgical gowns
FDA · DeviceCardinal HealthWorldwide distribution - US Nationwide and the countries of Chile, Thailand,…Official notice →US
Mentor Texas, LP. recalls MENTOR Saline-Filled Breast Implants are constructed from room temperature vulcanized silicone elastomer, made of polydimethylsiloxane.…
Due to a manufacturing issue, Implant may have a potentially weakened area on the base patch portion of the implant which may cause leakage/deflation of the affected device.
FDA · DeviceMentor Texas, LP.Worldwide distribution - U.S. Nationwide distribution in the states of AK, AL,…Official notice →US
Medline Industries Inc recalls EVLP Convenience Pack/Kit
NaCl 250 mL Bag Inj expired 04/2020 while the kit expiration date was after 04/2020.
US
Datascope Corp. recalls Cardiosave Rescue IntraAortic Balloon Pump. Model Numbers: 0998-00-0800-75, 0998-UC-0800-75, 0998-00-0800-83, 0998-UC-0800-83,…
Firm is initiating a correction due to the possibility of fluid ingress. Fluid entering the device may short various electronic components thus leading to system shutdown, which could result in unsafe hemodynamic…
US