Datascope Corp. recalls Cardiosave Rescue IntraAortic Balloon Pump. Model Numbers: 0998-00-0800-75, 0998-UC-0800-75, 0998-00-0800-83, 0998-UC-0800-83,…
Firm is initiating a correction due to the possibility of fluid ingress. Fluid entering the device may short various electronic components thus leading to system shutdown, which could result in unsafe hemodynamic…
Read the official notice →- Date
- 2021-12-22
- Country
- United States
- Brand
- Datascope
- Product
- Cardiosave Rescue IntraAortic Balloon Pump. Model Numbers: 0998-00-0800-75, 0998-UC-0800-75, 0998-00-0800-83, 0998-UC-0800-83,…
- Classification
- Class I
- Units
- Worldwide distribution - US Nationwide.
- Category
- Drugs & medical
- Source
- FDA · Device
Copied automatically from the US Food and Drug Administration. Details may be shortened or out of date; the official notice is the authoritative source. Disclaimer
What to do
- Check the product name, strength and lot number against the official notice.
- Do not stop a prescribed medicine without talking to your doctor or pharmacist first.
- Ask the pharmacy or supplier about a replacement or refund.
General guidance from Recall Atlas. The official notice from the US Food and Drug Administration has the exact models, batch codes and remedy, and it takes priority.
More Datascope recalls
- 2023-01-25
US
Datascope Corp. recalls Cardiosave Hybrid Intra-Aortic Balloon Pump (IABP), Model Numbers 0998-00-0800-31, 0998-00-0800-32, 0998-00-0800-33, 0998-00-0800-34,…
The Cardiosave IABP may shut down unexpectedly due to blood entering in the Cardiosave IABP when therapy is provided with a compromised intra-aortic balloon catheter.
FDA · DeviceDatascope Corp.Worldwide Distribution - US Nationwide and the countries of Albania, Angola,…Official notice →- 2022-11-02
US
Datascope Corporation recalls LINEAR 7.5Fr. 40cc IAB with Accessories (APA), Model No. 0684-00-0480-02. intra-aortic balloon catheter.
The balloon volume is incorrectly described as 34cc instead of 40cc on one of the three internal labels enclosed for hospital use. A labelling error may result in user confusion and/or delay in initiating therapy.…
FDA · DeviceDatascope CorporationNo US distribution. Foreign distribution to Argentina, Brazil, Canada, Chile,…Official notice →- 2022-09-14
US
Datascope Corporation recalls MEGA 8Fr 50cc Intra-Aortic Balloon Catheter
Certain lots containing undersized dilator.
FDA · DeviceDatascope CorporationAwaiting domestic distribution. Foreign: Australia, Belgium, Canada, Czech…Official notice →- 2021-10-27
US
Datascope Corp. recalls Cardiosave Li-Ion Battery Pack, Part No. 0146-00-0097. Component of the Cardiosave Hybrid Intra-Aortic Balloon Pump (IABP) with the below…
A Cardiosave Lithium-Ion Battery Pack used during protocol testing failed to meet the minimum runtime requirement per Getinge internal Product Specification. These nonconforming batteries were inadvertently released to…
FDA · DeviceDatascope Corp.Worldwide distribution. US nationwide, Australia, Brazil, Bahrain, Belgium,…Official notice →- 2021-10-13
US
Datascope Corp. recalls Cardiosave Hybrid Intra-Aortic Balloon Pump (IABP), Model Nos. 0998-00-0800-31, 0998-00-0800-32, 0998-00-0800-33, 0998-00-0800-34,…
The Helium Indicator on the Cardiosave display may over represent the amount of helium in unit in the presence of Electromagnetic Interference.
- 2021-10-06
US
Datascope Corp. recalls Cardiosave Hybrid Intra-Aortic Balloon Pump (IABP), Model Nos. 0998-00-0800-31, 0998-00-0800-32, 0998-00-0800-33, 0998-00-0800-34,…
The Cardiosave IABP may unexpectedly shut down when the device is running on AC power, only one battery is installed in the IABP, and the battery is physically removed while the battery is being charged. This occurs…
FDA · DeviceDatascope Corp.Nationwide domestic distribution. Foreign distribution worldwide.Official notice →
Other drugs & medical recalls in the United States
- 2021-12-22
US
Mentor Texas, LP. recalls MENTOR Saline-Filled Breast Implants are constructed from room temperature vulcanized silicone elastomer, made of polydimethylsiloxane.…
Due to a manufacturing issue, Implant may have a potentially weakened area on the base patch portion of the implant which may cause leakage/deflation of the affected device.
FDA · DeviceMentor Texas, LP.Worldwide distribution - U.S. Nationwide distribution in the states of AK, AL,…Official notice →US
Smith & Nephew, Inc. recalls smith&Nephew EVOS MEDIAL DISTAL FEMUR PLATE, L, 5H, 90MM, REF 72573101
Testing vendor notified Smith & Nephew that previous MRI scanning conditions provided in Exponents reports were erroneous. Erroneous data from the reports was incorporated in the released IFU 81110294 EVOS Large…
FDA · DeviceSmith & Nephew, Inc.US Nationwide distribution in the states of CO, FL, GA, MA, MN, MO, NC, NJ,…Official notice →US
Medline Industries Inc recalls EVLP Convenience Pack/Kit
NaCl 250 mL Bag Inj expired 04/2020 while the kit expiration date was after 04/2020.
US
US
US