Drugs & medicalClass II

Datascope Corp. recalls Cardiosave Hybrid Intra-Aortic Balloon Pump (IABP), Model Nos. 0998-00-0800-31, 0998-00-0800-32, 0998-00-0800-33, 0998-00-0800-34,…

The Cardiosave IABP may unexpectedly shut down when the device is running on AC power, only one battery is installed in the IABP, and the battery is physically removed while the battery is being charged. This occurs…

Read the official notice →
Date
2021-10-06
Country
United States
Brand
Datascope
Product
Cardiosave Hybrid Intra-Aortic Balloon Pump (IABP), Model Nos. 0998-00-0800-31, 0998-00-0800-32, 0998-00-0800-33, 0998-00-0800-34,…
Classification
Class II
Units
Nationwide domestic distribution. Foreign distribution worldwide.
Category
Drugs & medical
Source
FDA · Device

Copied automatically from the US Food and Drug Administration. Details may be shortened or out of date; the official notice is the authoritative source. Disclaimer

What to do

  1. Check the product name, strength and lot number against the official notice.
  2. Do not stop a prescribed medicine without talking to your doctor or pharmacist first.
  3. Ask the pharmacy or supplier about a replacement or refund.

General guidance from Recall Atlas. The official notice from the US Food and Drug Administration has the exact models, batch codes and remedy, and it takes priority.

More Datascope recalls

  1. 2023-01-25
  2. US
    Drugs & medicalClass I

    Datascope Corp. recalls Cardiosave Hybrid Intra-Aortic Balloon Pump (IABP), Model Numbers 0998-00-0800-31, 0998-00-0800-32, 0998-00-0800-33, 0998-00-0800-34,…

    The Cardiosave IABP may shut down unexpectedly due to blood entering in the Cardiosave IABP when therapy is provided with a compromised intra-aortic balloon catheter.

    FDA · DeviceDatascope Corp.Worldwide Distribution - US Nationwide and the countries of Albania, Angola,…Official notice →
  3. 2022-11-02
  4. US
    Drugs & medicalClass II

    Datascope Corporation recalls LINEAR 7.5Fr. 40cc IAB with Accessories (APA), Model No. 0684-00-0480-02. intra-aortic balloon catheter.

    The balloon volume is incorrectly described as 34cc instead of 40cc on one of the three internal labels enclosed for hospital use. A labelling error may result in user confusion and/or delay in initiating therapy.…

    FDA · DeviceDatascope CorporationNo US distribution. Foreign distribution to Argentina, Brazil, Canada, Chile,…Official notice →
  5. 2022-09-14
  6. US
    Drugs & medicalClass II

    Datascope Corporation recalls MEGA 8Fr 50cc Intra-Aortic Balloon Catheter

    Certain lots containing undersized dilator.

    FDA · DeviceDatascope CorporationAwaiting domestic distribution. Foreign: Australia, Belgium, Canada, Czech…Official notice →
  7. 2021-12-22
  8. US
    Drugs & medicalClass I

    Datascope Corp. recalls Cardiosave Rescue IntraAortic Balloon Pump. Model Numbers: 0998-00-0800-75, 0998-UC-0800-75, 0998-00-0800-83, 0998-UC-0800-83,…

    Firm is initiating a correction due to the possibility of fluid ingress. Fluid entering the device may short various electronic components thus leading to system shutdown, which could result in unsafe hemodynamic…

    FDA · DeviceDatascope Corp.Worldwide distribution - US Nationwide.Official notice →
  9. 2021-10-27
  10. US
    Drugs & medicalClass I

    Datascope Corp. recalls Cardiosave Li-Ion Battery Pack, Part No. 0146-00-0097. Component of the Cardiosave Hybrid Intra-Aortic Balloon Pump (IABP) with the below…

    A Cardiosave Lithium-Ion Battery Pack used during protocol testing failed to meet the minimum runtime requirement per Getinge internal Product Specification. These nonconforming batteries were inadvertently released to…

    FDA · DeviceDatascope Corp.Worldwide distribution. US nationwide, Australia, Brazil, Bahrain, Belgium,…Official notice →
  11. 2021-10-13
  12. US
    Drugs & medicalClass II

    Datascope Corp. recalls Cardiosave Hybrid Intra-Aortic Balloon Pump (IABP), Model Nos. 0998-00-0800-31, 0998-00-0800-32, 0998-00-0800-33, 0998-00-0800-34,…

    The Helium Indicator on the Cardiosave display may over represent the amount of helium in unit in the presence of Electromagnetic Interference.

    FDA · DeviceDatascope Corp.Worldwide distribution - US Nationwide distribution.Official notice →
All 29 Datascope recalls →

Other drugs & medical recalls in the United States

  1. 2021-10-13
  2. US
    Drugs & medicalClass II

    AMIVAS (US), LLC recalls Artesunate for Injection, 110 mg/vial, packaged in a) 2x2 pack containing 2 Single-dose vials artesunate (73607-001-01) and 2 vials…

    Lack of Assurance of Sterility: Air filter failed post integrity test leading to a lack of sterility assurance.

    FDA · DrugAMIVAS (US), LLCProduct was distributed to 3 major distributors who may have further…Official notice →
  3. US
    Drugs & medicalClass I

    Alpha-Tek LLC recalls ALPHA MALE+ Male Enhancer, fruit chew strips, 1 strip per foil pouch, Distributed by Umbrella Labs, Made in the USA, UPC 8 60003 03770 6.…

    Marketed Without an Approved NDA/ANDA: FDA analysis determined the presence of tadalafil

    FDA · DrugAlpha-Tek LLCNationwide in the USOfficial notice →
  4. 2021-10-06
  5. US
    Drugs & medicalClass II

    Angiodynamics Inc. (Navilyst Medical Inc.) recalls Xcela PICC with PASV Valve Technology/Kit 5 F-55cm; (Maximum Flow Rate 3.5 mL/Sec)-indicated for short or long-term…

    Mislabeled: The product is labeled as 5F (55cm; Maximum Flow Rate 3.5 mL/Sec) and the product is actually 4F (55cm; Maximum Flow Rate 3.5 mL/Sec).

    FDA · DeviceAngiodynamics Inc. (Navilyst Medical Inc.)CanadaOfficial notice →
  6. US
    Drugs & medicalClass II

    ConvaTec, Inc recalls DuoDERM Control Gel Formula Border Dressings 15x15cm (1x5pk)

    Primary packaging labeling may be incorrect.

    FDA · DeviceConvaTec, IncDistribution in US and Canada US distribution in CA, FL, GA, IL, MD, MN, NC.…Official notice →
  7. US
    Drugs & medicalClass II

    Corin Ltd recalls Corin Unity Knee CS Insert, Size 5 / 9.0mm, Fixed Condylar Stabilised - UHMWPE, REF 1125182, Sterile, and Corin Unity Knee CS Insert, Size…

    The packaging system on the devices have potential physical and water damage resulting in damage to the devices, loss of sterility, or contamination of the devices.

    FDA · DeviceCorin LtdUS Nationwide distribution in the states of AZ, CA, FL, IL, OK, PA, TX, UT,…Official notice →
  8. US
    Drugs & medicalClass II

    Beckman Coulter Inc. recalls DxA 5000 (DxA Automation System, DxA 5000 fit), "For in Vitro Diagnostic Use", Reference Numbers: B87352 with B57634, UDI: 15099590707118

    There is a potential that sample carriers (w or w/o samples) within automated sampling system may transport samples to the outer lane which does not contain a physical barrier to restrain samples. Samples are dropped…

    FDA · DeviceBeckman Coulter Inc.US Nationwide distribution in the states of AL, CA, IL, In, KY, MN, MI, OH,…Official notice →
See all drugs & medical recalls →