Datascope recalls
29 official recalls and safety alerts involving Datascope since 2021-10, from FDA · Device. Newest first. Check the official notice for exact models and batch codes.
This list only includes recalls Recall Atlas has collected since 2021-10. A product not listed here may still be affected; check with Datascope or the relevant agency.
- 2026-03-25
US
Datascope Corp. recalls Battery Charging Station; Model: 0998-00-0802;
The Battery Charging Station is an optional dual-bay charger used to charge the Cardiosave IABP Lithium- Ion batteries when they are not being used to operate the IABP. Getinge has identified that a protruding screw in…
FDA · DeviceDatascope Corp.Worldwide - US Nationwide distribution in the states of FL, DC, GA, IL, IN,…Official notice →- 2026-03-11
US
Datascope Corp. recalls CS300 IABP. Software Version CS300 IABP C.01.
The firm has identified that the battery runtime and cycle specifications contained in the devices Instructions for Use (IFU) need to be updated.
FDA · DeviceDatascope Corp.Worldwide distribution - US Nationwide and the countries of Afghanistan,…Official notice →- 2025-12-24
US
Datascope Corp. recalls Cardiosave Rescue
The IFU addendum updates the Vibration and Shock Table to reference the correct standards.
FDA · DeviceDatascope Corp.Domestic: Nationwide distribution. International: Albania, Algeria, Angola,…Official notice →- 2025-12-17
US
Datascope Corp. recalls Cardiosave Rescue. Intra-Aortic Balloon Pump system.
The IFU addendum revises the Preventative Maintenance schedule to align with the update introduced in the Service Manual as of June 2023.
FDA · DeviceDatascope Corp.Domestic: Nationwide Distribution; Foreign: Albania, Algeria, Angola,…Official notice →- 2024-11-13
US
Datascope Corp. recalls MEGA 7.5Fr. 40cc Intra-Aortic Balloon with Accessories, Model Number 0684-00- 0295-01
There is discrepant labeling on the inner and outer packaging of the device.
FDA · DeviceDatascope Corp.International distribution to the countries of Germany, Spain , Finland, Great…Official notice →- 2024-10-02
US
Datascope Corp. recalls CARDIOSAVE RESCUE INTRA-AORTIC BALLOON PUMP (IABP). Electromechanical system used to inflate and deflate intra-aortic balloons.
Firm has developed a software correction (Revision D.01) for outside the US to mitigate product problems that pose a risk of hemodynamic instability and failure of transmission of clinical data in Cardiosave Hybrid and…
FDA · DeviceDatascope Corp.International distribution to the countries of UNITED ARAB EMIRATES, ARGENTINA…Official notice →US
Datascope Corp. recalls CARDIOSAVE HYBRID INTRA-AORTIC BALLOON PUMP (IABP). Electromechanical system used to inflate and deflate intra-aortic balloons.
Firm has developed a software correction (Revision D.00) outside the US to mitigate product problems that pose a risk of hemodynamic instability and failure of transmission of clinical data in Cardiosave Hybrid and…
FDA · DeviceDatascope Corp.International distribution to the countries of UNITED ARAB EMIRATES, ARGENTINA…Official notice →- 2023-09-06
US
Datascope Corp. recalls Cardiosave Rescue Intra-Aortic Balloon Pumps (IABP), Part Numbers 0998-00-0800-75, 0998-00-0800-83, 0998-00-0800-85
Users were experiencing a failure in the IAB Fiber Optic Sensor input on the IABP when inserting the Intra-Aortic Balloon Fiber Optic connector.
US
Datascope Corp. recalls Cardiosave Rescue Intra-Aortic Balloon Pumps (IABP), Part Numbers 0998-00-0800-75, 0998-00-0800-83, 0998-00-0800-85
Users were identifying autofill failure conditions on the devices causing pump stops.
US
Datascope Corp. recalls Cardiosave Hybrid Intra-Aortic Balloon Pumps (IABP), Part Numbers 0998-00-0800-31, 0998-UC-0800-31, 0998-00-0800-32, 0998-00-0800-33,…
Users were reporting instances of "Gas Loss in IAB Circuit" and "Gas Gain in IAB Circuit" alarms while providing therapy. An internal investigation of the complaints determined that there is a potential trigger for…
US
Datascope Corp. recalls Cardiosave Hybrid Intra-Aortic Balloon Pumps (IABP), Part Numbers 0998-00-0800-31, 0998-UC-0800-31, 0998-00-0800-32, 0998-00-0800-33,…
Users reported "System Over Temperature" alarms associated with a loss of pumping and/or the Cardiosave system entering Standby mode.
US
Datascope Corp. recalls Cardiosave Hybrid Intra-Aortic Balloon Pumps (IABP), Part Numbers 0998-00-0800-31, 0998-UC-0800-31, 0998-00-0800-32, 0998-00-0800-33,…
Users were reporting that the device was not charging as expected. It was discovered that users were unaware that the Cardiosave console was not completely inserted into the hospital cart. If the console is not fully…
- 2023-08-02
US
Datascope Corp. recalls LINEAR 7.5Fr.IAB, Intra-Aortic Balloon Catheter and Accessories, used to provide counter pulsation therapy in the aorta Product…
During IAB catheter insertion, the introducer dilator may fracture at the hub, leaving the body of the introducer dilator housed within the sheath. Risks to health include damage to the femoral artery, descending…
FDA · DeviceDatascope Corp.Domestic distribution nationwide. International distribution worldwide.Official notice →- 2023-07-26
US
Datascope Corp. recalls Cardiosave Hybrid Intra-Aortic Balloon Pump (IABP), Model numbers 0998-00-0800-31 0998-00-0800-32 0998-00-0800-33 0998-00-0800-34…
An unexpected shutdown of the IABP may occur due to failures of tantalum capacitors in the Power Management Board and/or Solenoid Board. This issue may lead to an unexpected interruption of therapy.
FDA · DeviceDatascope Corp.Worldwide Distribution- US (Nationwide) and OUS (foreign) countries of:…Official notice →US
Datascope Corp. recalls Cardiosave Hybrid Intra-Aortic Balloon Pump (IABP), Model numbers 0998-00-0800-31 0998-00-0800-32 0998-00-0800-33 0998-00-0800-34…
IABP may lose the ability to charge batteries in one or both bay slots. Therapy may be interrupted if batteries fail to charge and the device is disconnected from AC power. Low battery alarms may alert the User to the…
FDA · DeviceDatascope Corp.Worldwide Distribution: US (Nationwide) and OUS (foreign) to countries of:…Official notice →- 2023-06-21
US
Datascope Corp. recalls Cardiosave Hybrid Intra-Aortic Balloon Pumps (IABP), Model Number 0998-00-0800-55, Part Number 0998-00-0800-XX
The digital IFUs provided with the Cardiosave IABP during the software update to version D.00 incorrectly annotated compliance to standard 60601-1-12:2014 within the updated IFU.
- 2023-05-03
US
Datascope Corp. recalls MAQUET CARDIOSAVE hybrid Intra-Aortic Balloon Pump, Model Number 0998-XX-0800-32
The power cord supplied (Type J plug end) for the Cardiosave Hybrid IABP to Brazil is incompatible with the Brazilian Type N power receptacle.
- 2023-03-29
US
Datascope Corp. recalls Cardiosave Hybrid Intra-Aortic Balloon Pump (IABP), Model Numbers Model Numbers 0998-00-0800-75, 0998-00-0800-83, 0998-00-0800-85
There is an issue related to the Safety Disk not meeting a performance requirement that may impact Intra-Aortic Balloon catheter (IAB) displacement volume. This potential failure of an impacted Safety Disk may lead to…
- 2023-03-15
US
Datascope Corp. recalls Cardiosave Hybrid Intra-Aortic Balloon Pump. Model Numbers: D998-00-0800-31, D998-00-0800-32, D998-00-0800-33, D998-00-0800-35,…
Firm has received reports of damaged, worn, or torn O-rings on the Cardiosave Pump Console quick disconnect fitting, which results in helium tank leaks. If the helium supply is depleted, therapy will be interrupted.
FDA · DeviceDatascope Corp.Nationwide US distribution. Worldwide international distribution.Official notice →US
Datascope Corp. recalls Cardiosave Rescue Intra-Aortic Balloon Pump (IABP), Model Numbers 0998-00-0800-75, 0998-00-0800-83, and 0998-00-0800-85
There have been reported failures of the high pressure helium regulator, which may cause a helium leak in the Cardiosave Hospital Cart. In instance of helium regulator failure, a Pump Console s internal reservoir of…
US
Datascope Corp. recalls Cardiosave Hybrid Model Nos. 0998-00-0800-31 0998-00-0800-32 0998-00-0800-33 0998-00-0800-34 0998-00-0800-35 0998-00-0800-45…
An unexpected shutdown of the IABP may occur due to loss of communication between the Executive Processor PCBA and the Video Generator PCBA. This may result in an interruption to therapy which may threaten the…
FDA · DeviceDatascope Corp.Worldwide Distribution: US (nationwide) including states of: AK, AL, AR, AZ,…Official notice →US
Datascope Corp. recalls Cardiosave Rescue Intra-Aortic Balloon Pumps (IABP), Containing Coiled Cord cable (part number 0012-00-1801, Model Numbers…
An unexpected shutdown of the IABP may occur due to a failure of the connection between the Coiled Cord cable and the Cable Assembly backplane to the Coiled Cord cable which provides the communication between the…
FDA · DeviceDatascope Corp.Worldwide Distribution: US (nationwide) and OUS (Foreign) ARGENTINA,…Official notice →- 2023-01-25
US
Datascope Corp. recalls Cardiosave Hybrid Intra-Aortic Balloon Pump (IABP), Model Numbers 0998-00-0800-31, 0998-00-0800-32, 0998-00-0800-33, 0998-00-0800-34,…
The Cardiosave IABP may shut down unexpectedly due to blood entering in the Cardiosave IABP when therapy is provided with a compromised intra-aortic balloon catheter.
FDA · DeviceDatascope Corp.Worldwide Distribution - US Nationwide and the countries of Albania, Angola,…Official notice →- 2022-11-02
US
Datascope Corporation recalls LINEAR 7.5Fr. 40cc IAB with Accessories (APA), Model No. 0684-00-0480-02. intra-aortic balloon catheter.
The balloon volume is incorrectly described as 34cc instead of 40cc on one of the three internal labels enclosed for hospital use. A labelling error may result in user confusion and/or delay in initiating therapy.…
FDA · DeviceDatascope CorporationNo US distribution. Foreign distribution to Argentina, Brazil, Canada, Chile,…Official notice →- 2022-09-14
US
Datascope Corporation recalls MEGA 8Fr 50cc Intra-Aortic Balloon Catheter
Certain lots containing undersized dilator.
FDA · DeviceDatascope CorporationAwaiting domestic distribution. Foreign: Australia, Belgium, Canada, Czech…Official notice →- 2021-12-22
US
Datascope Corp. recalls Cardiosave Rescue IntraAortic Balloon Pump. Model Numbers: 0998-00-0800-75, 0998-UC-0800-75, 0998-00-0800-83, 0998-UC-0800-83,…
Firm is initiating a correction due to the possibility of fluid ingress. Fluid entering the device may short various electronic components thus leading to system shutdown, which could result in unsafe hemodynamic…
- 2021-10-27
US
Datascope Corp. recalls Cardiosave Li-Ion Battery Pack, Part No. 0146-00-0097. Component of the Cardiosave Hybrid Intra-Aortic Balloon Pump (IABP) with the below…
A Cardiosave Lithium-Ion Battery Pack used during protocol testing failed to meet the minimum runtime requirement per Getinge internal Product Specification. These nonconforming batteries were inadvertently released to…
FDA · DeviceDatascope Corp.Worldwide distribution. US nationwide, Australia, Brazil, Bahrain, Belgium,…Official notice →- 2021-10-13
US
Datascope Corp. recalls Cardiosave Hybrid Intra-Aortic Balloon Pump (IABP), Model Nos. 0998-00-0800-31, 0998-00-0800-32, 0998-00-0800-33, 0998-00-0800-34,…
The Helium Indicator on the Cardiosave display may over represent the amount of helium in unit in the presence of Electromagnetic Interference.
- 2021-10-06
US
Datascope Corp. recalls Cardiosave Hybrid Intra-Aortic Balloon Pump (IABP), Model Nos. 0998-00-0800-31, 0998-00-0800-32, 0998-00-0800-33, 0998-00-0800-34,…
The Cardiosave IABP may unexpectedly shut down when the device is running on AC power, only one battery is installed in the IABP, and the battery is physically removed while the battery is being charged. This occurs…
FDA · DeviceDatascope Corp.Nationwide domestic distribution. Foreign distribution worldwide.Official notice →