Drugs & medicalClass I

Datascope Corp. recalls Cardiosave Hybrid Intra-Aortic Balloon Pumps (IABP), Part Numbers 0998-00-0800-31, 0998-UC-0800-31, 0998-00-0800-32, 0998-00-0800-33,…

Users were reporting instances of "Gas Loss in IAB Circuit" and "Gas Gain in IAB Circuit" alarms while providing therapy. An internal investigation of the complaints determined that there is a potential trigger for…

Read the official notice →
Date
2023-09-06
Country
United States
Brand
Datascope
Product
Cardiosave Hybrid Intra-Aortic Balloon Pumps (IABP), Part Numbers 0998-00-0800-31, 0998-UC-0800-31, 0998-00-0800-32, 0998-00-0800-33,…
Classification
Class I
Units
Worldwide - US Nationwide distribution.
Category
Drugs & medical
Source
FDA · Device

Copied automatically from the US Food and Drug Administration. Details may be shortened or out of date; the official notice is the authoritative source. Disclaimer

What to do

  1. Check the product name, strength and lot number against the official notice.
  2. Do not stop a prescribed medicine without talking to your doctor or pharmacist first.
  3. Ask the pharmacy or supplier about a replacement or refund.

General guidance from Recall Atlas. The official notice from the US Food and Drug Administration has the exact models, batch codes and remedy, and it takes priority.

More Datascope recalls

  1. 2024-10-02
  2. US
    Drugs & medicalClass I

    Datascope Corp. recalls CARDIOSAVE HYBRID INTRA-AORTIC BALLOON PUMP (IABP). Electromechanical system used to inflate and deflate intra-aortic balloons.

    Firm has developed a software correction (Revision D.00) outside the US to mitigate product problems that pose a risk of hemodynamic instability and failure of transmission of clinical data in Cardiosave Hybrid and…

    FDA · DeviceDatascope Corp.International distribution to the countries of UNITED ARAB EMIRATES, ARGENTINA…Official notice →
  3. 2023-09-06
  4. US
    Drugs & medicalClass II

    Datascope Corp. recalls Cardiosave Rescue Intra-Aortic Balloon Pumps (IABP), Part Numbers 0998-00-0800-75, 0998-00-0800-83, 0998-00-0800-85

    Users were experiencing a failure in the IAB Fiber Optic Sensor input on the IABP when inserting the Intra-Aortic Balloon Fiber Optic connector.

    FDA · DeviceDatascope Corp.US Nationwide. Global Distribution.Official notice →
  5. US
    Drugs & medicalClass I

    Datascope Corp. recalls Cardiosave Rescue Intra-Aortic Balloon Pumps (IABP), Part Numbers 0998-00-0800-75, 0998-00-0800-83, 0998-00-0800-85

    Users were identifying autofill failure conditions on the devices causing pump stops.

    FDA · DeviceDatascope Corp.Worldwide - US Nationwide distribution.Official notice →
  6. US
    Drugs & medicalClass I

    Datascope Corp. recalls Cardiosave Hybrid Intra-Aortic Balloon Pumps (IABP), Part Numbers 0998-00-0800-31, 0998-UC-0800-31, 0998-00-0800-32, 0998-00-0800-33,…

    Users reported "System Over Temperature" alarms associated with a loss of pumping and/or the Cardiosave system entering Standby mode.

    FDA · DeviceDatascope Corp.Worldwide - US Nationwide distribution.Official notice →
  7. US
    Drugs & medicalClass I

    Datascope Corp. recalls Cardiosave Hybrid Intra-Aortic Balloon Pumps (IABP), Part Numbers 0998-00-0800-31, 0998-UC-0800-31, 0998-00-0800-32, 0998-00-0800-33,…

    Users were reporting that the device was not charging as expected. It was discovered that users were unaware that the Cardiosave console was not completely inserted into the hospital cart. If the console is not fully…

    FDA · DeviceDatascope Corp.Worldwide - US Nationwide distribution.Official notice →
  8. 2023-08-02
  9. US
    Drugs & medicalClass II

    Datascope Corp. recalls LINEAR 7.5Fr.IAB, Intra-Aortic Balloon Catheter and Accessories, used to provide counter pulsation therapy in the aorta Product…

    During IAB catheter insertion, the introducer dilator may fracture at the hub, leaving the body of the introducer dilator housed within the sheath. Risks to health include damage to the femoral artery, descending…

    FDA · DeviceDatascope Corp.Domestic distribution nationwide. International distribution worldwide.Official notice →
All 29 Datascope recalls →

Other drugs & medical recalls in the United States

  1. 2023-09-06
  2. US
    Drugs & medicalClass II

    RAYSEARCH LABORATORIES AB recalls RayCare software, include version number 5A, 5B, 6A, including service packs-An oncology information system used to support workflows…

    An issue was found in Software RayCare 5A, 5B, 6A, including service packs, where an allergy warning against medication substance (drug ingredient) will not be displayed as expected under certain circumstances

    FDA · DeviceRAYSEARCH LABORATORIES ABUS Nationwide distribution in the state of TN.Official notice →
  3. US
    Drugs & medicalClass II

    Stryker Corporation recalls SurgiCount+ Software Application-indicated as an adjunctive technology for augmenting the manual process of counting, displaying and…

    Potential for current software to miscount when scanning in multiple sponge-products from the same unique sponge-pack type

    FDA · DeviceStryker CorporationUS Nationwide distribution.Official notice →
  4. US
    Drugs & medicalClass I

    Datascope Corp. recalls Cardiosave Rescue Intra-Aortic Balloon Pumps (IABP), Part Numbers 0998-00-0800-75, 0998-00-0800-83, 0998-00-0800-85

    Users were identifying autofill failure conditions on the devices causing pump stops.

    FDA · DeviceDatascope Corp.Worldwide - US Nationwide distribution.Official notice →
  5. US
    Drugs & medicalClass I

    Datascope Corp. recalls Cardiosave Hybrid Intra-Aortic Balloon Pumps (IABP), Part Numbers 0998-00-0800-31, 0998-UC-0800-31, 0998-00-0800-32, 0998-00-0800-33,…

    Users reported "System Over Temperature" alarms associated with a loss of pumping and/or the Cardiosave system entering Standby mode.

    FDA · DeviceDatascope Corp.Worldwide - US Nationwide distribution.Official notice →
  6. US
    Drugs & medicalClass I

    Datascope Corp. recalls Cardiosave Hybrid Intra-Aortic Balloon Pumps (IABP), Part Numbers 0998-00-0800-31, 0998-UC-0800-31, 0998-00-0800-32, 0998-00-0800-33,…

    Users were reporting that the device was not charging as expected. It was discovered that users were unaware that the Cardiosave console was not completely inserted into the hospital cart. If the console is not fully…

    FDA · DeviceDatascope Corp.Worldwide - US Nationwide distribution.Official notice →
  7. US
    Drugs & medicalClass II

    Siemens Healthcare Diagnostics, Inc. recalls Atellica CH Triglycerides (concentrated)-In vitro diagnostic use in the quantitative measurement of triglycerides in human…

    Reagent Carryover from LDL Cholesterol Direct (DLDL), Total Protein II (TP), Triglycerides (Trig), and Triglycerides_2 (Trig_2) Impacting Magnesium (Mg) Results

    FDA · DeviceSiemens Healthcare Diagnostics, Inc.Nationwide Foreign: Country Albania Algeria Argentina Australia Austria…Official notice →
See all drugs & medical recalls →