Datascope Corp. recalls Cardiosave Rescue Intra-Aortic Balloon Pumps (IABP), Part Numbers 0998-00-0800-75, 0998-00-0800-83, 0998-00-0800-85
Users were experiencing a failure in the IAB Fiber Optic Sensor input on the IABP when inserting the Intra-Aortic Balloon Fiber Optic connector.
Read the official notice →- Date
- 2023-09-06
- Country
- United States
- Brand
- Datascope
- Product
- Cardiosave Rescue Intra-Aortic Balloon Pumps (IABP), Part Numbers 0998-00-0800-75, 0998-00-0800-83, 0998-00-0800-85
- Classification
- Class II
- Units
- US Nationwide. Global Distribution.
- Category
- Drugs & medical
- Source
- FDA · Device
Copied automatically from the US Food and Drug Administration. Details may be shortened or out of date; the official notice is the authoritative source. Disclaimer
What to do
- Check the product name, strength and lot number against the official notice.
- Do not stop a prescribed medicine without talking to your doctor or pharmacist first.
- Ask the pharmacy or supplier about a replacement or refund.
General guidance from Recall Atlas. The official notice from the US Food and Drug Administration has the exact models, batch codes and remedy, and it takes priority.
More Datascope recalls
- 2024-11-13
US
Datascope Corp. recalls MEGA 7.5Fr. 40cc Intra-Aortic Balloon with Accessories, Model Number 0684-00- 0295-01
There is discrepant labeling on the inner and outer packaging of the device.
FDA · DeviceDatascope Corp.International distribution to the countries of Germany, Spain , Finland, Great…Official notice →- 2024-10-02
US
Datascope Corp. recalls CARDIOSAVE RESCUE INTRA-AORTIC BALLOON PUMP (IABP). Electromechanical system used to inflate and deflate intra-aortic balloons.
Firm has developed a software correction (Revision D.01) for outside the US to mitigate product problems that pose a risk of hemodynamic instability and failure of transmission of clinical data in Cardiosave Hybrid and…
FDA · DeviceDatascope Corp.International distribution to the countries of UNITED ARAB EMIRATES, ARGENTINA…Official notice →US
Datascope Corp. recalls CARDIOSAVE HYBRID INTRA-AORTIC BALLOON PUMP (IABP). Electromechanical system used to inflate and deflate intra-aortic balloons.
Firm has developed a software correction (Revision D.00) outside the US to mitigate product problems that pose a risk of hemodynamic instability and failure of transmission of clinical data in Cardiosave Hybrid and…
FDA · DeviceDatascope Corp.International distribution to the countries of UNITED ARAB EMIRATES, ARGENTINA…Official notice →- 2023-09-06
US
Datascope Corp. recalls Cardiosave Rescue Intra-Aortic Balloon Pumps (IABP), Part Numbers 0998-00-0800-75, 0998-00-0800-83, 0998-00-0800-85
Users were identifying autofill failure conditions on the devices causing pump stops.
US
Datascope Corp. recalls Cardiosave Hybrid Intra-Aortic Balloon Pumps (IABP), Part Numbers 0998-00-0800-31, 0998-UC-0800-31, 0998-00-0800-32, 0998-00-0800-33,…
Users were reporting instances of "Gas Loss in IAB Circuit" and "Gas Gain in IAB Circuit" alarms while providing therapy. An internal investigation of the complaints determined that there is a potential trigger for…
US
Datascope Corp. recalls Cardiosave Hybrid Intra-Aortic Balloon Pumps (IABP), Part Numbers 0998-00-0800-31, 0998-UC-0800-31, 0998-00-0800-32, 0998-00-0800-33,…
Users reported "System Over Temperature" alarms associated with a loss of pumping and/or the Cardiosave system entering Standby mode.
Other drugs & medical recalls in the United States
- 2023-09-06
US
Olympus Corporation of the Americas recalls Olympus Bronchovideoscope, Models BF-3C40 & BF-N20.
Complaint of endoscope model becoming lodged in the endotracheal tube connector due to diameter being too large.
US
Accuray Incorporated recalls Upper roller bracket (part number 1088370) for the Gen2 gantry enclosure for the Radixact Treatment Delivery System.
It is possible the upper roller bracket in the Radixact System gantry enclosure may not provide enough clearance for service personnel to use a jib crane for moving components while servicing the system, resulting in…
FDA · DeviceAccuray IncorporatedWorldwide - US Nationwide distribution in the states of CA, GA, IL, LA, NJ,…Official notice →US
Beckman Coulter Inc. recalls Beckman Coulter, IVD, REF: OSR61171, IgA, 4 x 14 mL R1, 4 x 11 mL R2
Lipemic interference failed to meet the performance specification listed within the IFU.
FDA · DeviceBeckman Coulter Inc.US: AL AR AZ CA CO CT DC DE FL GA IA IL IN KS KY LA MA MD MI MN MO MS NC ND NE…Official notice →US
PHILIPS MEDICAL SYSTEMS NEDERLAND B.V. recalls Wireless Foot Switch distributed with Philips Allura Xper and Azurion Interventional Fluoroscopic X-Ray Systems
Loss of availability of the wireless foot switch during procedures.
FDA · DevicePHILIPS MEDICAL SYSTEMS NEDERLAND B.V.Worldwide distribution - US Nationwide and the countries of Australia,…Official notice →US
Oculus Optikgeraete GMBH recalls Oculus Pentacam AXL , Ref 70100, CE 0123
Their is a potential that optical devices with insufficient anti-reflective coating may lead to incorrect axial length measurements.
FDA · DeviceOculus Optikgeraete GMBHUS Nationwide distribution in the states of CA, FL, MA, MN, NY, NC, OH, PA,…Official notice →US
Waldemar Link GmbH & Co. KG (Mfg Site) recalls Link OptiStem Rasp Handle-Used with the OptiStem Rasp to prepare the femoral medullary canal for a knee prosthesis…
A crack in the weld of the rasp handle may lead to a fracture of the frontal sleeve when the rasp is further impacted or extracted. The rasp remains intact. The rasp would need to be pulled out by a clamp leading to a…