Datascope Corp. recalls Cardiosave Rescue Intra-Aortic Balloon Pumps (IABP), Part Numbers 0998-00-0800-75, 0998-00-0800-83, 0998-00-0800-85
Users were identifying autofill failure conditions on the devices causing pump stops.
Read the official notice →- Date
- 2023-09-06
- Country
- United States
- Brand
- Datascope
- Product
- Cardiosave Rescue Intra-Aortic Balloon Pumps (IABP), Part Numbers 0998-00-0800-75, 0998-00-0800-83, 0998-00-0800-85
- Classification
- Class I
- Units
- Worldwide - US Nationwide distribution.
- Category
- Drugs & medical
- Source
- FDA · Device
Copied automatically from the US Food and Drug Administration. Details may be shortened or out of date; the official notice is the authoritative source. Disclaimer
What to do
- Check the product name, strength and lot number against the official notice.
- Do not stop a prescribed medicine without talking to your doctor or pharmacist first.
- Ask the pharmacy or supplier about a replacement or refund.
General guidance from Recall Atlas. The official notice from the US Food and Drug Administration has the exact models, batch codes and remedy, and it takes priority.
More Datascope recalls
- 2024-10-02
US
Datascope Corp. recalls CARDIOSAVE RESCUE INTRA-AORTIC BALLOON PUMP (IABP). Electromechanical system used to inflate and deflate intra-aortic balloons.
Firm has developed a software correction (Revision D.01) for outside the US to mitigate product problems that pose a risk of hemodynamic instability and failure of transmission of clinical data in Cardiosave Hybrid and…
FDA · DeviceDatascope Corp.International distribution to the countries of UNITED ARAB EMIRATES, ARGENTINA…Official notice →US
Datascope Corp. recalls CARDIOSAVE HYBRID INTRA-AORTIC BALLOON PUMP (IABP). Electromechanical system used to inflate and deflate intra-aortic balloons.
Firm has developed a software correction (Revision D.00) outside the US to mitigate product problems that pose a risk of hemodynamic instability and failure of transmission of clinical data in Cardiosave Hybrid and…
FDA · DeviceDatascope Corp.International distribution to the countries of UNITED ARAB EMIRATES, ARGENTINA…Official notice →- 2023-09-06
US
Datascope Corp. recalls Cardiosave Rescue Intra-Aortic Balloon Pumps (IABP), Part Numbers 0998-00-0800-75, 0998-00-0800-83, 0998-00-0800-85
Users were experiencing a failure in the IAB Fiber Optic Sensor input on the IABP when inserting the Intra-Aortic Balloon Fiber Optic connector.
US
Datascope Corp. recalls Cardiosave Hybrid Intra-Aortic Balloon Pumps (IABP), Part Numbers 0998-00-0800-31, 0998-UC-0800-31, 0998-00-0800-32, 0998-00-0800-33,…
Users were reporting instances of "Gas Loss in IAB Circuit" and "Gas Gain in IAB Circuit" alarms while providing therapy. An internal investigation of the complaints determined that there is a potential trigger for…
US
Datascope Corp. recalls Cardiosave Hybrid Intra-Aortic Balloon Pumps (IABP), Part Numbers 0998-00-0800-31, 0998-UC-0800-31, 0998-00-0800-32, 0998-00-0800-33,…
Users reported "System Over Temperature" alarms associated with a loss of pumping and/or the Cardiosave system entering Standby mode.
US
Datascope Corp. recalls Cardiosave Hybrid Intra-Aortic Balloon Pumps (IABP), Part Numbers 0998-00-0800-31, 0998-UC-0800-31, 0998-00-0800-32, 0998-00-0800-33,…
Users were reporting that the device was not charging as expected. It was discovered that users were unaware that the Cardiosave console was not completely inserted into the hospital cart. If the console is not fully…
Other drugs & medical recalls in the United States
- 2023-09-06
US
Cardiac Assist, Inc recalls LifeSPARC Pump (part number 5800-0000) contained within TandemLife LifeSPARC Priming Tray part number: 5800-1000. Note: The 5800-1000…
Epoxy used during manufacturing of the LifeSPARC Pump may have uncured epoxy. if undetectable curing may result in added risk of early pump failure,
US
RAYSEARCH LABORATORIES AB recalls RayCare software, include version number 5A, 5B, 6A, including service packs-An oncology information system used to support workflows…
An issue was found in Software RayCare 5A, 5B, 6A, including service packs, where an allergy warning against medication substance (drug ingredient) will not be displayed as expected under certain circumstances
FDA · DeviceRAYSEARCH LABORATORIES ABUS Nationwide distribution in the state of TN.Official notice →US
Stryker Corporation recalls SurgiCount+ Software Application-indicated as an adjunctive technology for augmenting the manual process of counting, displaying and…
Potential for current software to miscount when scanning in multiple sponge-products from the same unique sponge-pack type
US
Datascope Corp. recalls Cardiosave Hybrid Intra-Aortic Balloon Pumps (IABP), Part Numbers 0998-00-0800-31, 0998-UC-0800-31, 0998-00-0800-32, 0998-00-0800-33,…
Users were reporting instances of "Gas Loss in IAB Circuit" and "Gas Gain in IAB Circuit" alarms while providing therapy. An internal investigation of the complaints determined that there is a potential trigger for…
US
Datascope Corp. recalls Cardiosave Hybrid Intra-Aortic Balloon Pumps (IABP), Part Numbers 0998-00-0800-31, 0998-UC-0800-31, 0998-00-0800-32, 0998-00-0800-33,…
Users reported "System Over Temperature" alarms associated with a loss of pumping and/or the Cardiosave system entering Standby mode.
US
Datascope Corp. recalls Cardiosave Hybrid Intra-Aortic Balloon Pumps (IABP), Part Numbers 0998-00-0800-31, 0998-UC-0800-31, 0998-00-0800-32, 0998-00-0800-33,…
Users were reporting that the device was not charging as expected. It was discovered that users were unaware that the Cardiosave console was not completely inserted into the hospital cart. If the console is not fully…