Drugs & medicalClass I

Datascope Corp. recalls CARDIOSAVE RESCUE INTRA-AORTIC BALLOON PUMP (IABP). Electromechanical system used to inflate and deflate intra-aortic balloons.

Firm has developed a software correction (Revision D.01) for outside the US to mitigate product problems that pose a risk of hemodynamic instability and failure of transmission of clinical data in Cardiosave Hybrid and…

Read the official notice →
Date
2024-10-02
Country
United States
Brand
Datascope
Product
CARDIOSAVE RESCUE INTRA-AORTIC BALLOON PUMP (IABP). Electromechanical system used to inflate and deflate intra-aortic balloons.
Classification
Class I
Units
International distribution to the countries of UNITED ARAB EMIRATES, ARGENTINA…
Category
Drugs & medical
Source
FDA · Device

Copied automatically from the US Food and Drug Administration. Details may be shortened or out of date; the official notice is the authoritative source. Disclaimer

What to do

  1. Check the product name, strength and lot number against the official notice.
  2. Do not stop a prescribed medicine without talking to your doctor or pharmacist first.
  3. Ask the pharmacy or supplier about a replacement or refund.

General guidance from Recall Atlas. The official notice from the US Food and Drug Administration has the exact models, batch codes and remedy, and it takes priority.

More Datascope recalls

  1. 2025-12-24
  2. US
    Drugs & medicalClass III

    Datascope Corp. recalls Cardiosave Rescue

    The IFU addendum updates the Vibration and Shock Table to reference the correct standards.

    FDA · DeviceDatascope Corp.Domestic: Nationwide distribution. International: Albania, Algeria, Angola,…Official notice →
  3. 2025-12-17
  4. US
    Drugs & medicalClass II

    Datascope Corp. recalls Cardiosave Rescue. Intra-Aortic Balloon Pump system.

    The IFU addendum revises the Preventative Maintenance schedule to align with the update introduced in the Service Manual as of June 2023.

    FDA · DeviceDatascope Corp.Domestic: Nationwide Distribution; Foreign: Albania, Algeria, Angola,…Official notice →
  5. 2024-11-13
  6. US
    Drugs & medicalClass II

    Datascope Corp. recalls MEGA 7.5Fr. 40cc Intra-Aortic Balloon with Accessories, Model Number 0684-00- 0295-01

    There is discrepant labeling on the inner and outer packaging of the device.

    FDA · DeviceDatascope Corp.International distribution to the countries of Germany, Spain , Finland, Great…Official notice →
  7. 2024-10-02
  8. US
    Drugs & medicalClass I

    Datascope Corp. recalls CARDIOSAVE HYBRID INTRA-AORTIC BALLOON PUMP (IABP). Electromechanical system used to inflate and deflate intra-aortic balloons.

    Firm has developed a software correction (Revision D.00) outside the US to mitigate product problems that pose a risk of hemodynamic instability and failure of transmission of clinical data in Cardiosave Hybrid and…

    FDA · DeviceDatascope Corp.International distribution to the countries of UNITED ARAB EMIRATES, ARGENTINA…Official notice →
  9. 2023-09-06
  10. US
    Drugs & medicalClass II

    Datascope Corp. recalls Cardiosave Rescue Intra-Aortic Balloon Pumps (IABP), Part Numbers 0998-00-0800-75, 0998-00-0800-83, 0998-00-0800-85

    Users were experiencing a failure in the IAB Fiber Optic Sensor input on the IABP when inserting the Intra-Aortic Balloon Fiber Optic connector.

    FDA · DeviceDatascope Corp.US Nationwide. Global Distribution.Official notice →
  11. US
    Drugs & medicalClass I

    Datascope Corp. recalls Cardiosave Rescue Intra-Aortic Balloon Pumps (IABP), Part Numbers 0998-00-0800-75, 0998-00-0800-83, 0998-00-0800-85

    Users were identifying autofill failure conditions on the devices causing pump stops.

    FDA · DeviceDatascope Corp.Worldwide - US Nationwide distribution.Official notice →
All 29 Datascope recalls →

Other drugs & medical recalls in the United States

  1. 2024-10-02
  2. US
    Drugs & medicalClass II

    Roche Diagnostics Operations, Inc. recalls Creatine Kinase (CK) used on cobas c 311/501/502 and COBAS INTEGRA 400 plus-In vitro test for the quantitative determination…

    Creatine Kinase (CK) Assay Abnormal calibrations on the cobas¿ c 311/501/502 and COBAS INTEGRA 400 plus Analyzers may lead to erroneous results and misdiagnosis.

    FDA · DeviceRoche Diagnostics Operations, Inc.US Nationwide distribution.Official notice →
  3. US
    Drugs & medicalClass II

    FUJIFILM Healthcare Americas Corporation recalls The device is a mobile x-ray system designed to work with Fujifilm's GOS and CsI scintillator FDR D-EVO2 (DR-ID)…

    When the key switch is turned on, the exposure conditions are set to the default values (85kV, 1.6 mAs, approx. 0.13mGy@ SID 1m) intended for adult chest imaging. In addition, if exposure conditions are set from the…

    FDA · DeviceFUJIFILM Healthcare Americas CorporationUS Nationwide and WorldwideOfficial notice →
  4. US
    Drugs & medicalClass I

    Fresenius Kabi USA, LLC recalls Ivenix Infusion System (IIS), LVP Software; LVP-SW-0004, Software Version 5.9.1 and prior. Infusion pump software for infusion management…

    The software has anomalies that have the potential to cause alarms, nonfunctioning pump, or unresponsive screen while continuing therapy. These could cause serious patient harm or death.

    FDA · DeviceFresenius Kabi USA, LLCUS Nationwide.Official notice →
  5. US
    Drugs & medicalClass I

    Datascope Corp. recalls CARDIOSAVE HYBRID INTRA-AORTIC BALLOON PUMP (IABP). Electromechanical system used to inflate and deflate intra-aortic balloons.

    Firm has developed a software correction (Revision D.00) outside the US to mitigate product problems that pose a risk of hemodynamic instability and failure of transmission of clinical data in Cardiosave Hybrid and…

    FDA · DeviceDatascope Corp.International distribution to the countries of UNITED ARAB EMIRATES, ARGENTINA…Official notice →
  6. US
    Drugs & medicalClass I

    Smiths Medical ASD Inc. recalls paraPAC plus 310 ventilator kit with internal PEEP and CPAP, Item Number P310N

    Smiths Medical became aware of an issue related to a potential for the patient outlet connector to loosen/detach from the paraPAC Plus P300 and P310 ventilators impacting the active ventilation function.

    FDA · DeviceSmiths Medical ASD Inc.Worldwide distribution.Official notice →
  7. US
    Drugs & medicalClass II

    Wipro GE Healthcare Private Ltd. recalls GE Panda iRES Warmer, Model/Catalog Numbers: 1) M1112198-01045518; 2) M1112198-01045579; 3) M1112198-01047148; 4)…

    GE HealthCare has become aware that complete electrical safety testing was not conducted during manufacturing of certain Giraffe and Panda iRes Warmers. There are multiple electrical safety protections included in the…

    FDA · DeviceWipro GE Healthcare Private Ltd.US and Australia, Denmark, France, Germany, Ireland, ITALY, New Zealand,…Official notice →
See all drugs & medical recalls →