Cardiac Assist, Inc recalls LifeSPARC Pump (part number 5800-0000) contained within TandemLife LifeSPARC Priming Tray part number: 5800-1000. Note: The 5800-1000…
Epoxy used during manufacturing of the LifeSPARC Pump may have uncured epoxy. if undetectable curing may result in added risk of early pump failure,
Read the official notice →- Date
- 2023-09-06
- Country
- United States
- Brand
- Cardiac Assist
- Product
- LifeSPARC Pump (part number 5800-0000) contained within TandemLife LifeSPARC Priming Tray part number: 5800-1000. Note: The 5800-1000…
- Classification
- Class II
- Units
- US Nationwide distribution in the states of IL, FL.
- Category
- Drugs & medical
- Source
- FDA · Device
Copied automatically from the US Food and Drug Administration. Details may be shortened or out of date; the official notice is the authoritative source. Disclaimer
What to do
- Check the product name, strength and lot number against the official notice.
- Do not stop a prescribed medicine without talking to your doctor or pharmacist first.
- Ask the pharmacy or supplier about a replacement or refund.
General guidance from Recall Atlas. The official notice from the US Food and Drug Administration has the exact models, batch codes and remedy, and it takes priority.
More Cardiac Assist recalls
- 2023-01-25
US
Cardiac Assist, Inc recalls SHIP KIT, LIFESPARC, CONTROLLER AND ACCESSORIES - 5900-0000; SHIP KIT, LIFESPARC, CONTROLLER ONLY - 5900-0001. Component of LifeSPARC…
A software update (v1.1.5) has been developed to address the issue of Critical Failure which can occur on the LifeSPARC Controller. The Critical Failure was first addressed in the firm's recall initiated July 21, 2022.…
- 2022-10-05
US
Cardiac Assist, Inc recalls TANDEM LIFE LivaNova LIFESPARC Controller; Controller and Accessories Model 5900-0000; Controller Only Model 5900-0001. For circulatory or…
Critical Failure of the LifeSPARC controller when the software freezes or crashes and the screen does not display data.
- 2021-12-08
US
Cardiac Assist, Inc recalls SHIP KIT, PROCEDURE KIT, PROTEKDUO LS - DL31 for temporary Cardiopulmonary bypass. Including the following components: a) 31 Fr ProtekDuo…
Top label of the outer package of the collection kit Item 5820-3118 showed expiration date of 2022/05/01 (correct expiration date) while side label showed expiration date of 2021/05/01 (incorrect expiration date)
Other drugs & medical recalls in the United States
- 2023-09-06
US
LeMaitre Vascular, Inc. recalls The Chevalier Valvulotome is a device used to disrupt the valves of veins which are to be used as grafts in the in situ or non-reversed…
Burrs (molding flash) on the cutter may tear the harvested vien
FDA · DeviceLeMaitre Vascular, Inc.International distribution to the countries of Belgium, France, Germany, and…Official notice →US
Olympus Corporation of the Americas recalls EVIS EXERA III DUODENOVIDEOSCOPE OLYMPUS TJF-Q190V
Recent reports of patient infection.
US
Smiths Medical ASD Inc. recalls PORTEX 0.5u HYDROPHOBIC DISC FILTER, REF 002291 - Bacterial/Viral filter
Smiths Medical identified a labeling error in which the barcode (UDI) was not correctly reflecting the expiration date of the product
US
RAYSEARCH LABORATORIES AB recalls RayCare software, include version number 5A, 5B, 6A, including service packs-An oncology information system used to support workflows…
An issue was found in Software RayCare 5A, 5B, 6A, including service packs, where an allergy warning against medication substance (drug ingredient) will not be displayed as expected under certain circumstances
FDA · DeviceRAYSEARCH LABORATORIES ABUS Nationwide distribution in the state of TN.Official notice →US
Stryker Corporation recalls SurgiCount+ Software Application-indicated as an adjunctive technology for augmenting the manual process of counting, displaying and…
Potential for current software to miscount when scanning in multiple sponge-products from the same unique sponge-pack type
US
Datascope Corp. recalls Cardiosave Rescue Intra-Aortic Balloon Pumps (IABP), Part Numbers 0998-00-0800-75, 0998-00-0800-83, 0998-00-0800-85
Users were identifying autofill failure conditions on the devices causing pump stops.