Cardiac Assist, Inc recalls SHIP KIT, PROCEDURE KIT, PROTEKDUO LS - DL31 for temporary Cardiopulmonary bypass. Including the following components: a) 31 Fr ProtekDuo…
Top label of the outer package of the collection kit Item 5820-3118 showed expiration date of 2022/05/01 (correct expiration date) while side label showed expiration date of 2021/05/01 (incorrect expiration date)
Read the official notice →- Date
- 2021-12-08
- Country
- United States
- Brand
- Cardiac Assist
- Product
- SHIP KIT, PROCEDURE KIT, PROTEKDUO LS - DL31 for temporary Cardiopulmonary bypass. Including the following components: a) 31 Fr ProtekDuo…
- Classification
- Class III
- Units
- US Distribution to states of: DC, MA, SC, and VA.
- Category
- Drugs & medical
- Source
- FDA · Device
Copied automatically from the US Food and Drug Administration. Details may be shortened or out of date; the official notice is the authoritative source. Disclaimer
What to do
- Check the product name, strength and lot number against the official notice.
- Do not stop a prescribed medicine without talking to your doctor or pharmacist first.
- Ask the pharmacy or supplier about a replacement or refund.
General guidance from Recall Atlas. The official notice from the US Food and Drug Administration has the exact models, batch codes and remedy, and it takes priority.
More Cardiac Assist recalls
- 2023-09-06
US
Cardiac Assist, Inc recalls LifeSPARC Pump (part number 5800-0000) contained within TandemLife LifeSPARC Priming Tray part number: 5800-1000. Note: The 5800-1000…
Epoxy used during manufacturing of the LifeSPARC Pump may have uncured epoxy. if undetectable curing may result in added risk of early pump failure,
- 2023-01-25
US
Cardiac Assist, Inc recalls SHIP KIT, LIFESPARC, CONTROLLER AND ACCESSORIES - 5900-0000; SHIP KIT, LIFESPARC, CONTROLLER ONLY - 5900-0001. Component of LifeSPARC…
A software update (v1.1.5) has been developed to address the issue of Critical Failure which can occur on the LifeSPARC Controller. The Critical Failure was first addressed in the firm's recall initiated July 21, 2022.…
- 2022-10-05
US
Cardiac Assist, Inc recalls TANDEM LIFE LivaNova LIFESPARC Controller; Controller and Accessories Model 5900-0000; Controller Only Model 5900-0001. For circulatory or…
Critical Failure of the LifeSPARC controller when the software freezes or crashes and the screen does not display data.
Other drugs & medical recalls in the United States
- 2021-12-08
US
Howmedica Osteonics Corp. recalls 28MM -4 LFIT V40 HEAD, Catalog Number 6260-9-122
There is a potential for the outer white Tyvek lid to debond from the sealed package.
US
US
BioFire Diagnostics, LLC recalls BioFire Blood Culture Identification 2 (BCID2) Panel Part No: RFIT-ASY-0147 UDI: 00815381020338 Blood Culture Catalog No./Description:…
Due to an increased risk of false positive Acinetobacter calcoaceticus- baumannii complex results. The cause may be as a due to non-viable contamination in the blood culture vials.
FDA · DeviceBioFire Diagnostics, LLCWorldwide distribution - U.S. Nationwide distribution in the states of AL, AR,…Official notice →US
DRG International, Inc. recalls Total Thyroxine (T4) ELISA; REF EIA-1781
Firm has identified two lots of product that do not meet product shelf life, resulting in failed Controls. The clinical impact may be a delay in diagnosis and a consumption of patient samples.
FDA · DeviceDRG International, Inc.Domestic: NY; International: TZ, VN, BD, CO, TT, YE, BO, DE, SV, GE, PY, TG,…Official notice →US
Beeken Biomedical, LLC recalls NuStat 8" x12" Trauma Pad XR Hemafiber, 5-pack. For temporary control of internal organ space bleeding REF: XR-0812-5
Endotoxin levels exceeded allowable limits.
US
Thoratec Corp. recalls Abbott HeartMate Touch Communication System, REF: HMT1100 (US), HMT1100-R (Rental US), HMT1150 (EU)
If LVAS communication system is trying to establish Bluetooth connection and another Bluetooth-enabled device is nearby and advertising for Bluetooth connection, the Bluetooth connectivity interference may cause the…
FDA · DeviceThoratec Corp.Worldwide distribution - US Nationwide distribution in the states of FL, IL,…Official notice →