Philips North America Llc recalls eCareManager version 4.2.1, eCareManager version 4.3.2, eCareManager version 4.3.3, eCareManager version 4.5.1, eCareManager version…
eCareManager (eCM) Sentry Score software not approved for use
Read the official notice →- Date
- 2022-02-09
- Country
- United States
- Brand
- Philips
- Product
- eCareManager version 4.2.1, eCareManager version 4.3.2, eCareManager version 4.3.3, eCareManager version 4.5.1, eCareManager version…
- Classification
- Class II
- Units
- Worldwide distribution - US Nationwide and the country of Australia.
- Category
- Drugs & medical
- Source
- FDA · Device
Copied automatically from the US Food and Drug Administration. Details may be shortened or out of date; the official notice is the authoritative source. Disclaimer
What to do
- Check the product name, strength and lot number against the official notice.
- Do not stop a prescribed medicine without talking to your doctor or pharmacist first.
- Ask the pharmacy or supplier about a replacement or refund.
General guidance from Recall Atlas. The official notice from the US Food and Drug Administration has the exact models, batch codes and remedy, and it takes priority.
More Philips recalls
- 2022-04-13
US
Philips North America LLC recalls Philips, REF M5071A, Adult. Smart PADS CARTRIDGE, 8+ years FOR USE WITH HEARSTART HS1 DEFIBRILLATORS
There is a potential for AED pads to experience gel separation from the foam/tin backing when peeled from the plastic liner. The gel may fold onto itself resulting in reduced surface area or gel on the pad, or it may…
FDA · DevicePhilips North America LLCWorldwide distribution. US Nationwide including Puerto Rico and Guam, Germany,…Official notice →US
Philips North America Llc recalls Philips Allura (Model Numbers 722010, 722012, 722026, 722027, 722028, 722029, 722035) and Philips Azurion (Model Numbers 722063,…
Due to a software defect, the Philips StentBoost Live R2.0 might not process X-ray images of the current run and instead show an image processed in a previous run. The image shown could be from the same patient or from…
- 2022-03-16
US
Philips North America Llc recalls Philips Allura Xper, UNIQ(Allura Xper system-China) Interventional Fluoroscopic X-ray System products with a Laird chiller WITHOUT an…
Due to a leak in the detector cooling system, cooling liquid may leak outside the drip tray of the chiller. If this occurs the liquid may drip onto electrical components in the R cabinet located in the technical room,…
FDA · DevicePhilips North America LlcUS Nationwide Foreign: Argentina Austria Bangladesh Brazil Chile China Denmark…Official notice →- 2022-01-26
US
Philips North America Llc recalls Infa-Therm Transport Mattress, Reference Number 989805616831 1015
After the device was cleared by the FDA, Philips made changes to the labeling that do not fall within the existing FDA clearance. These devices with the modified labeling cannot be distributed or sold without new…
- 2022-01-19
US
Philips North America Llc recalls Wireless Footswtich used with the following Philips Allura Xper and Azurion Interventional Fluoroscopic X-ray System . Model Number:…
Wireless foot switch connection issues causing interruption of Fluoroscopy and exposure
FDA · DevicePhilips North America LlcWorldwide distribution - US Nationwide and the countries of Canada, Australia,…Official notice →US
Philips North America Llc recalls Philips Azurion Interventional Fluoroscopic X-ray System with Azurion R2.2 with LUC configuration start-up issue; Azurion R2.2…
(1)Start-up problem: Intermittently at start-up of the system, the communication of the control unit that manages the movement of the stand is not established; (2) Emergency stop recovery problem-warm restart is not…
FDA · DevicePhilips North America LlcWorldwide distribution - US Nationwide distribution in the state of WA and the…Official notice →
Other drugs & medical recalls in the United States
- 2022-02-09
US
Luminex Corporation recalls ARIES HSV 1&2 Assay, Catalog/REF# 50-10017
Impacted lot may leak inside the ARIES instrument. Testing confirmed a leak rate of 0.9%. Leaking cassettes could potentially lead to incorrect results or a delay in reporting results due to an Invalid.
FDA · DeviceLuminex CorporationWorldwide distribution - US Nationwide distribution in the states of New…Official notice →US
Olympus Corporation of the Americas recalls Spiration Valve System (SVS) Valve in Cartridge -A one-way endobronchial valves indicated for adult patients with shortness…
Mislabeled Patient Charge Label indicating model number SVS-V7-00 (7mm) instead of SVS-V9-00 (9mm)
FDA · DeviceOlympus Corporation of the AmericasUS Nationwide distribution in the states of AZ, CA, MA, NJ, PA, SC, TX, WI.Official notice →US
Abiomed, Inc. recalls OXY-1 System, Model Number 001-0500-001. Used during cardiopulmonary bypass. OXY-1 System consists of Console, Pump Drive, E Drive, and…
The firm has received complaints of the OXY-1 System Console experiencing a power interruption while in use, which disrupts therapy delivered to a patient on support. Disruption of therapy could lead to prolonged…
US
Smith & Nephew, Inc. recalls INTEGRA SURFIX ALPHA Screwdriver Torx 10, REF 219 135 ND
The affected screwdrivers are out of specification and may not mate with the Surfix Alpha Lock screw
FDA · DeviceSmith & Nephew, Inc.US Nationwide distribution in the states of CA and OH.Official notice →US
Siemens Medical Solutions USA, Inc. recalls ACUSON Sequoia Ultrasound Imaging System with Software Version VA25A/B
Due to a calculation error in the measurement when using 2D trace (manual trace) tool. The trace circumference value is overestimated and may potentially result in misdiagnosis of a patient's condition.
FDA · DeviceSiemens Medical Solutions USA, Inc.Worldwide distribution - U.S. Nationwide distribution in the states of AL, AK,…Official notice →US
Microbiologics Inc recalls KWIK-STI(TM), GBS QC Set (Live Culture), REF 0243U. Specifically affected component: KWIK-STIK(TM), Lactobacillus acidophilus derived from…
The expiration date on the label of one component (negative control 0243U) in the set lot is incorrectly labeled. The KWIK-STIK negative control 0243U was labeled with an expiration date of 7/31/2020 when it should…
FDA · DeviceMicrobiologics IncUS Nationwide distribution in the states of AL, CA, CO, IL, MD, MO, MS, NM,…Official notice →