Philips North America Llc recalls Philips Azurion Interventional Fluoroscopic X-ray System with Azurion R2.2 with LUC configuration start-up issue; Azurion R2.2…
(1)Start-up problem: Intermittently at start-up of the system, the communication of the control unit that manages the movement of the stand is not established; (2) Emergency stop recovery problem-warm restart is not…
Read the official notice →- Date
- 2022-01-19
- Country
- United States
- Brand
- Philips
- Product
- Philips Azurion Interventional Fluoroscopic X-ray System with Azurion R2.2 with LUC configuration start-up issue; Azurion R2.2 emergency…
- Classification
- Class II
- Units
- Worldwide distribution - US Nationwide distribution in the state of WA and the…
- Category
- Drugs & medical
- Source
- FDA · Device
Copied automatically from the US Food and Drug Administration. Details may be shortened or out of date; the official notice is the authoritative source. Disclaimer
What to do
- Check the product name, strength and lot number against the official notice.
- Do not stop a prescribed medicine without talking to your doctor or pharmacist first.
- Ask the pharmacy or supplier about a replacement or refund.
General guidance from Recall Atlas. The official notice from the US Food and Drug Administration has the exact models, batch codes and remedy, and it takes priority.
More Philips recalls
- 2022-02-09
US
Philips North America Llc recalls eCareManager version 4.2.1, eCareManager version 4.3.2, eCareManager version 4.3.3, eCareManager version 4.5.1, eCareManager version…
eCareManager (eCM) Sentry Score software not approved for use
FDA · DevicePhilips North America LlcWorldwide distribution - US Nationwide and the country of Australia.Official notice →- 2022-01-26
US
Philips North America Llc recalls Infa-Therm Transport Mattress, Reference Number 989805616831 1015
After the device was cleared by the FDA, Philips made changes to the labeling that do not fall within the existing FDA clearance. These devices with the modified labeling cannot be distributed or sold without new…
- 2022-01-19
US
Philips North America Llc recalls Wireless Footswtich used with the following Philips Allura Xper and Azurion Interventional Fluoroscopic X-ray System . Model Number:…
Wireless foot switch connection issues causing interruption of Fluoroscopy and exposure
FDA · DevicePhilips North America LlcWorldwide distribution - US Nationwide and the countries of Canada, Australia,…Official notice →- 2021-12-29
US
Philips North America Llc recalls Philips Allura Xper FD series with Software Version Number: 2.1.x- intended for Vascular, cardiovascular and neurovascular imaging…
Set screws securing the two gliding rods of the Extra Monitors (called 7th and 8th monitor) mounted on the Monitor Ceiling Suspension (MCS) can come loose, the Extra monitors might disengage from the Mounting Ceiling…
FDA · DevicePhilips North America LlcWorldwide distribution - US Nationwide distribution in the states of DC, GA,…Official notice →- 2021-12-15
US
Philips North America Llc recalls BV Endura with Software Release 2.3- A mobile, diagnostic X-ray image acquisition and viewing system. Model Number: 718075
Instructions for Use do not specify the maximum surface temperature of the X-ray tank (assembly housing the X-ray tube, beam filter & cooling oil) and the image intensifier/detector of the system as required an result…
FDA · DevicePhilips North America LlcWorldwide distribution - US Nationwide and the countries of Afghanistan,…Official notice →- 2021-12-01
US
Philips North America Llc recalls Continuous CT Fluoroscopy DVI Ceiling Mount - CCT Arm DVI Option - Gemini TF 64 System Number: 882471 Continuous CT (CCT) is a CT or…
Significant wear of the Monitor Bracket Assembly shaft may go unrecognized could lead to the Monitor Bracket Assembly detaching from the Suspension Arm, and resulting in injury to patient, operator, bystander, or…
FDA · DevicePhilips North America LlcUS (nationwide) distribution to states of: CA, ME, TX and OUS (Foreign) to…Official notice →
Other drugs & medical recalls in the United States
- 2022-01-19
US
Haag-Streit USA Inc recalls OCT-Camera 211 01 A3; Running software iOCT-Control 3 v1.4
Malfunction of the automatic laser beam shut-off, the OCT unit might not recognize whether the laser beam is safely switched off
US
Philips North America Llc recalls Wireless Footswtich used with the following Philips Allura Xper and Azurion Interventional Fluoroscopic X-ray System . Model Number:…
Wireless foot switch connection issues causing interruption of Fluoroscopy and exposure
FDA · DevicePhilips North America LlcWorldwide distribution - US Nationwide and the countries of Canada, Australia,…Official notice →US
Siemens Medical Solutions USA, Inc recalls SOMATOM Edge Plus-Computed tomography system Model 1026700
Software version VB20_SP5 my lead to a relevant degradation of head image quality, a potential risk to patients is misdiagnosis.
US
Reflexion Medical, Inc. recalls RefleXion Medical Radiotherapy System - System Label: "*** refleXion REF RXM1000 ***" Physics Guide/Treatment Delivery User…
Due to dose discrepancy when delivering a plan to a patient in a Non-HFS (Head First Supine) orientation specifically in Feet First Supine (FFS).
FDA · DeviceReflexion Medical, Inc.U.S. Nationwide distribution in the states of CA and TXOfficial notice →US
ROi CPS LLC recalls Regard, T AND A - Rogers, sterilized convenience kits
Kits lacks rubber latex warning label.
US
Getinge Usa Sales Inc recalls Vaporizer Sevoflurane Maquet Filling, Part No. 6886601 Component of the following systems: Product Code(s) UPN/UDI: Flow-c Anesthesia…
A potential chemical degradation of Sevoflurane by Lewis acids (metal oxides and metal halides) has been discovered that may result in inhalation and/or skin exposure of hydrogen fluoride.
FDA · DeviceGetinge Usa Sales IncWorldwide distribution - US Nationwide distribution in the states of VA and…Official notice →