Drugs & medicalClass II

ROi CPS LLC recalls Regard, T AND A - Rogers, sterilized convenience kits

Kits lacks rubber latex warning label.

Read the official notice →
Date
2022-01-19
Country
United States
Brand
ROi CPS
Product
Regard, T AND A - Rogers, sterilized convenience kits
Classification
Class II
Units
US Nationwide distribution in the state of MO.
Category
Drugs & medical
Source
FDA · Device

Copied automatically from the US Food and Drug Administration. Details may be shortened or out of date; the official notice is the authoritative source. Disclaimer

What to do

  1. Check the product name, strength and lot number against the official notice.
  2. Do not stop a prescribed medicine without talking to your doctor or pharmacist first.
  3. Ask the pharmacy or supplier about a replacement or refund.

General guidance from Recall Atlas. The official notice from the US Food and Drug Administration has the exact models, batch codes and remedy, and it takes priority.

More ROi CPS recalls

  1. 2022-05-18
  2. US
    Drugs & medicalClass II

    ROi CPS LLC recalls Regard brand CABG Pack A and B - Spartanburg

    Kits were packed into shipping boxes incorrectly.

    FDA · DeviceROi CPS LLCRecall impacted one domestic consignee in NCOfficial notice →
  3. 2022-05-04
  4. US
    Drugs & medicalClass II

    ROi CPS LLC recalls Convenience kits containing regard IV Start Kits: (1) regard Clinical Packaging Solutions, Item Number: 830015013, LD00200M - Vag Delivery…

    A drug kit component was recalled by its manufacturer due to the potential for inadequacy of the validation of test methods used in its manufacture.

    FDA · DeviceROi CPS LLCUS Nationwide distribution in the state of Missouri.Official notice →
  5. 2022-04-06
  6. US
    Drugs & medicalClass III

    ROi CPS LLC recalls Regard Custom Procedure Kits containing BD Luer Lok Access Devices

    Recalled devices were packaged in convenience kits.

    FDA · DeviceROi CPS LLCUS Nationwide distribution in the state of Missouri.Official notice →
  7. 2022-03-09
  8. US
    Drugs & medicalClass II

    ROi CPS LLC recalls Regard, TOTAL HIP - SPRINGFIELD, Item No. 8800297010, Case Quantity: 2, Sterile

    Products are mislabeled with the incorrect product identification on individual kit labeling.

    FDA · DeviceROi CPS LLCMissouriOfficial notice →
  9. 2021-12-29
  10. US
    Drugs & medicalClass II

    ROi CPS LLC recalls regard Clinical Packaging Solutions Item Number 830096005, OB01054E - Newborn Kit, containing 2 25GA x .625IN needles.

    The kit contains a non-sterile needle.

    FDA · DeviceROi CPS LLCDistribution was made to Missouri. There was no foreign/military/government…Official notice →
  11. 2021-11-24
  12. US
    Drugs & medicalClass II

    ROi CPS LLC recalls REGARD Vitrectomy Tray

    Povidone-Iodine swabstick manufactured by PDI, Inc. recalled due to Out of Specifications was used as a component in some of the ROI CPS, LLC products.

    FDA · DeviceROi CPS LLCUS Nationwide in the states of FL, LA and MO.Official notice →
All 20 ROi CPS recalls →

Other drugs & medical recalls in the United States

  1. 2022-01-19
  2. US
    Drugs & medicalClass II

    Siemens Medical Solutions USA, Inc recalls SOMATOM Edge Plus-Computed tomography system Model 1026700

    Software version VB20_SP5 my lead to a relevant degradation of head image quality, a potential risk to patients is misdiagnosis.

    FDA · DeviceSiemens Medical Solutions USA, IncUS Nationwide distribution.Official notice →
  3. US
    Drugs & medicalClass II

    Philips North America Llc recalls Philips Azurion Interventional Fluoroscopic X-ray System with Azurion R2.2 with LUC configuration start-up issue; Azurion R2.2…

    (1)Start-up problem: Intermittently at start-up of the system, the communication of the control unit that manages the movement of the stand is not established; (2) Emergency stop recovery problem-warm restart is not…

    FDA · DevicePhilips North America LlcWorldwide distribution - US Nationwide distribution in the state of WA and the…Official notice →
  4. US
    Drugs & medicalClass II

    Reflexion Medical, Inc. recalls RefleXion Medical Radiotherapy System - System Label: "*** refleXion REF RXM1000 ***" Physics Guide/Treatment Delivery User…

    Due to dose discrepancy when delivering a plan to a patient in a Non-HFS (Head First Supine) orientation specifically in Feet First Supine (FFS).

    FDA · DeviceReflexion Medical, Inc.U.S. Nationwide distribution in the states of CA and TXOfficial notice →
  5. US
    Drugs & medicalClass I

    Getinge Usa Sales Inc recalls Vaporizer Sevoflurane Maquet Filling, Part No. 6886601 Component of the following systems: Product Code(s) UPN/UDI: Flow-c Anesthesia…

    A potential chemical degradation of Sevoflurane by Lewis acids (metal oxides and metal halides) has been discovered that may result in inhalation and/or skin exposure of hydrogen fluoride.

    FDA · DeviceGetinge Usa Sales IncWorldwide distribution - US Nationwide distribution in the states of VA and…Official notice →
  6. US
    Drugs & medicalClass III

    Ki Mobility Llc recalls Ki Mobility, Little Wave Tyke (Weight Limit: 95 lbs), Specifically the IFU The Little Wave Tyke is a manually operated device that can be…

    The Instructions for Use provided with the device contained information intended for use of a mechanical wheelchair, not a chair with casters.

    FDA · DeviceKi Mobility LlcDomestic: IL, IN, KY, MI, MO, NC, SC, and TN.Official notice →
  7. US
    Drugs & medicalClass II

    Ion Beam Applications S.A. recalls Proteus 235; Version: PTS-8 versions before PTS-8.7.2

    Proton Therapy System (PTS) software can be started in the clinical environment (clinical user) while some processes are still running in the test environment (tcs user). Initiating the startup of the clinical…

    FDA · DeviceIon Beam Applications S.A.IN USA: VA, OK, NJ, TN OUS: Czech Republic.Official notice →