Drugs & medicalClass II

Reflexion Medical, Inc. recalls RefleXion Medical Radiotherapy System - System Label: "*** refleXion REF RXM1000 ***" Physics Guide/Treatment Delivery User…

Due to dose discrepancy when delivering a plan to a patient in a Non-HFS (Head First Supine) orientation specifically in Feet First Supine (FFS).

Read the official notice →
Date
2022-01-19
Country
United States
Brand
Reflexion Medical
Product
RefleXion Medical Radiotherapy System - System Label: "*** refleXion REF RXM1000 ***" Physics Guide/Treatment Delivery User…
Classification
Class II
Units
U.S. Nationwide distribution in the states of CA and TX
Category
Drugs & medical
Source
FDA · Device

Copied automatically from the US Food and Drug Administration. Details may be shortened or out of date; the official notice is the authoritative source. Disclaimer

What to do

  1. Check the product name, strength and lot number against the official notice.
  2. Do not stop a prescribed medicine without talking to your doctor or pharmacist first.
  3. Ask the pharmacy or supplier about a replacement or refund.

General guidance from Recall Atlas. The official notice from the US Food and Drug Administration has the exact models, batch codes and remedy, and it takes priority.

More Reflexion Medical recalls

  1. 2026-04-08
  2. US
    Drugs & medicalClass II

    Reflexion Medical, Inc. recalls RefleXion X1 Radiotherapy System Model RXM1000, Part #800-00006-00

    Due to a radiotherapy medical system software defect, when performing angular roll corrections followed by a repeat localization, the roll corrections applied in the initial localization scan are not carried through to…

    FDA · DeviceReflexion Medical, Inc.US: CA, TX, PA, CT, NJ, OR, LA, OHOfficial notice →
  3. 2025-09-10
  4. US
    Drugs & medicalClass II

    Reflexion Medical, Inc. recalls RefleXion X1 Model RXM1000 UDI Codes: (01)00860003983812(11)230504(21)X11001 (01)00860003983812(11)230518(21)X11002…

    Due to incorrect software configuration that potentially allows more than two (2) fractions within a 12-hour period (fractionLimits) and could potentially lead to adverse events (toxicity).

    FDA · DeviceReflexion Medical, Inc.U.S. Nationwide distribution in the states of CA, CT, LA, NJ, OR, PA, and TX.Official notice →
  5. 2024-06-05
  6. US
    Drugs & medicalClass II

    Reflexion Medical, Inc. recalls RefleXion X1 Radiotherapy System Model RXM1000 Software version 2.1.19-3 Radiation therapy system is a device that combines the…

    A potential dose error exists for patients treated with an out of session SCINTIX partial fraction.

    FDA · DeviceReflexion Medical, Inc.U.S.: CA, CT, NJ, OR, PA and TX No O.U.S.Official notice →
All 4 Reflexion Medical recalls →

Other drugs & medical recalls in the United States

  1. 2022-01-19
  2. US
    Drugs & medicalClass II

    Philips North America Llc recalls Wireless Footswtich used with the following Philips Allura Xper and Azurion Interventional Fluoroscopic X-ray System . Model Number:…

    Wireless foot switch connection issues causing interruption of Fluoroscopy and exposure

    FDA · DevicePhilips North America LlcWorldwide distribution - US Nationwide and the countries of Canada, Australia,…Official notice →
  3. US
    Drugs & medicalClass II

    Siemens Medical Solutions USA, Inc recalls SOMATOM Edge Plus-Computed tomography system Model 1026700

    Software version VB20_SP5 my lead to a relevant degradation of head image quality, a potential risk to patients is misdiagnosis.

    FDA · DeviceSiemens Medical Solutions USA, IncUS Nationwide distribution.Official notice →
  4. US
    Drugs & medicalClass II

    Philips North America Llc recalls Philips Azurion Interventional Fluoroscopic X-ray System with Azurion R2.2 with LUC configuration start-up issue; Azurion R2.2…

    (1)Start-up problem: Intermittently at start-up of the system, the communication of the control unit that manages the movement of the stand is not established; (2) Emergency stop recovery problem-warm restart is not…

    FDA · DevicePhilips North America LlcWorldwide distribution - US Nationwide distribution in the state of WA and the…Official notice →
  5. US
    Drugs & medicalClass II

    ROi CPS LLC recalls Regard, T AND A - Rogers, sterilized convenience kits

    Kits lacks rubber latex warning label.

    FDA · DeviceROi CPS LLCUS Nationwide distribution in the state of MO.Official notice →
  6. US
    Drugs & medicalClass I

    Getinge Usa Sales Inc recalls Vaporizer Sevoflurane Maquet Filling, Part No. 6886601 Component of the following systems: Product Code(s) UPN/UDI: Flow-c Anesthesia…

    A potential chemical degradation of Sevoflurane by Lewis acids (metal oxides and metal halides) has been discovered that may result in inhalation and/or skin exposure of hydrogen fluoride.

    FDA · DeviceGetinge Usa Sales IncWorldwide distribution - US Nationwide distribution in the states of VA and…Official notice →
  7. US
    Drugs & medicalClass III

    Ki Mobility Llc recalls Ki Mobility, Little Wave Tyke (Weight Limit: 95 lbs), Specifically the IFU The Little Wave Tyke is a manually operated device that can be…

    The Instructions for Use provided with the device contained information intended for use of a mechanical wheelchair, not a chair with casters.

    FDA · DeviceKi Mobility LlcDomestic: IL, IN, KY, MI, MO, NC, SC, and TN.Official notice →